Transcript Weak D
WEAK D TESTING (DU)
Mr. Mohammed Jaber.
WEAK EXPRESSION OF THE RHD
ANTIGEN (DU)
The
term DU is widely used to describe
cells which have
a
quantitative reduction in the
expression of their RhD antigen.
Or qualitative variation in RhD antigen
expression, these are referred to as
partial D.
Weak D individuals may also be partial D.
DU TEST SHOULD BE DONE IN THE
FOLLOWING SITUATIONS:
When weak or 1+ reactions are found.
Microscopic readings should only be done if
mixed field agglutination* is suspected.
When Rh typing discrepancies are found between
current and previous results.
On Rh negative neonates born to Rh negative
mothers. If the weak D testing is positive, the
neonatal Rh type would be reported as "D
positive" and the mother would be a candidate for
Rh Immune Globulin (RhIG).
PRINCIPLE
To test for a weak expression of the D antigen.
Red cells that react weakly or not at all in direct
agglutination tests with anti-D may react with
anti-D by the indirect antiglobulin test (IAT).
Red cells that fail to react 2+ in direct
agglutination tests with anti-D are incubated
with anti-D at 37° C and examined for
agglutination. The red cells are washed to
remove unbound antibody (IgG anti-D), then
tested with anti-IgG.
SPECIMENS
Clotted or anticoagulated whole blood
Reagents & Equipments
37oC incubator
Wash bottle with normal saline
Coombs serum - either polyspecific or anti-IgG
Coombs control cells (IgG coated control cells ).
All reagents, equipment, and supplies used in the
Rh testing procedure
RECOMMENDED TECHNIQUES
1.
SLIDE TECHNIQUE
1.1 Add to a clean, labelled slide:
• One drop of anti-D IgM & IgG blend.
• One drop of the test red cells suspension.
1.2Mix well by gently and continuously rocking
the slide for approx. 30 seconds and incubate
the test for 5 minutes at room temperature,
with occasional mixing.
1.3 Examine macroscopically for agglutination.
A diffuse light source may aid reading.
2. TUBE TECHNIQUE – IMMEDIATE
SPIN
2.1 Prepare a suspension of test washed red cells 2-3%
or 1.5-2% in LISS.
2.2 Place in a small, labelled test tube:
• 1 volume of anti-D IgM & IgG blend.
• 1 volume of suspended red cells.
2.3 Mix well.
2.4 Centrifuge immediately for 10 seconds at 1000g or
for a suitable alternative force and time.
2.5 Agitate the tube gently to dislodge the cell button
and examine macroscopically for agglutination.
Apparently negative tests which are to be tested for DU
should be further tested by the DU test method.
3. TUBE TECHNIQUE – LISS
3.1 Place in a small, labelled test tube:
• 1 volume of Anti-D blood grouping reagent
• 1 volume of red cells suspended 1.5-2% in LISS.
3.2 Mix well and incubate for 15-20 minutes at
37°C.
3.3 Centrifuge immediately for 10 seconds at 1000g
or for a suitable alternative force and time.
3.4 Agitate the tube gently to dislodge the cell
button and examine macroscopically for
agglutination.
4. DU TEST METHOD
(INDIRECT ANTIGLOBULIN TEST (IAT). )
After
reading the immediate spin results,
re-incubate the test for a further 20
minutes at 37°C before completing the DU
test method described below.
OR
After reading the LISS tube test, complete
the DU test, without further incubation,
following the procedure given below.
DU TEST METHOD
4.1 Wash the test 4 times with a large excess of
PBS (e.g. 4ml of PBS per 12 (or 10) x 75mm glass
tubes).
NOTE: a) allow adequate spin time to sediment
the
red cells
b) make sure that most of the residual saline
is
removed at the end if each wash
to
leave a ‘dry’ cell button.
4.2 Add two drops of anti-human globulin reagent
to each tube.
4.3 Mix thoroughly.
DU TEST METHOD
4.4 Centrifuge at 1000g for 10 seconds or for a
suitable alternative force and time.
4.5 Agitate the tube gently to dislodge the cell
button and examine macroscopically for
agglutination.
4.6 Add 1 drop of IgG-coated control cells to the
tube(s) with negative results. Centrifuge,
resuspend cells, read macroscopically and record
results. Agglutination (2+) shall be present or the
test shall be repeated.
PROCEDURAL NOTES
Tests
should
be
read
immediately
after
centrifugation. Delay may cause bound IgG to
dissociate from red cells and either leave too little IgG
to detect or neutralize AHG reagent causing false
negative results.
Interpretation
A negative result in the immediate spin phase but
agglutination in the D tube following incubation (with
no agglutination in the DC tube) indicates a
positive test for weak D.
Lack of agglutination is a negative test and the
patient is considered truly D negative.
Agglutination in the DC tube invalidates the test.