Learning objectives
Download
Report
Transcript Learning objectives
Image Gently, Pause and Pulse:
Practice of ALARA in
Pediatric Fluoroscopy
Sue C. Kaste, DO1, 2
Marta Hernanz-Schulman, MD3
Ishtiaq H. Bercha, M.Sc. 4
St. Jude Children’s Research Hospital
2 University of Tennessee Health Science Center
3 Monroe Carell Jr. Children’s Hospital at Vanderbilt
4 The Children’s Hospital, Aurora, Colorado.
1
1
ALARA
“As Low As Reasonably Achievable”
General principle guiding radiation exposure
Keep exposure to radiation dose as low as is
possible for each procedure, while obtaining needed
clinical information
= Image Optimization
2
Primary Learning Objective
Review pediatric fluoroscopic procedures
understand the source of radiation
understand methods to reduce radiation
effect on image quality
3
Other Learning Objectives
Fluoroscopy radiation units.
Scope of pediatric fluoroscopic procedures
Methods available for dose reduction
clinical settings to apply dose reduction
4
Fluoroscopy Radiation Units
Basic Radiation Quantities :
Exposure & Exposure Rate
Air Kerma & Air Kerma Rate
5
Fluoroscopy Radiation Units
Radiation Measurement Quantities:
Incident Air Kerma & Rate
Entrance Surface Air Kerma & Rate
6
Fluoroscopy Radiation Units
Risk Related Quantities:
Absorbed dose
Equivalent Dose
Effective dose
7
Basic Radiation Quantities
Exposure – expresses intensity of x-ray energy per
unit mass of air.
Units: Coulomb per kilogram (C/kg).
Commonly used units are Roentgen or milli
Roentgen, expressed as R or mR, respectively.
1 R = 2.58 x 10-4 C/kg
Exposure rate identifies x-ray intensity per unit time.
Commonly used units are R/min or mR/min.
8
Basic Radiation Quantities
Air Kerma (K) – sum of initial kinetic energies of all
charged particles generated by uncharged particles
such as x-ray photons released per unit mass of air.
Unit = Joule per kilogram, Commonly referred to as
Gray/milli Gray (Gy or mGy).
1 Roentgen of exposure 8.7 mGy air kerma
Air Kerma Rate quantifies air kerma per unit time
and is written as, dK/dt, that is, incremental kerma
per unit increment of time.
9
Measurement Quantities
Incident Air Kerma (Ka,i)– is the air kerma from the
incident beam along the central x-ray beam axis at the
skin entrance plane.
Only the primary beam is considered and the effect of
back scattered radiation is excluded.
Unit = Joule per kilogram, Commonly referred to as
Gray/milli Gray (Gy or mGy).
Incident Air Kerma Rate quantifies air kerma per unit
time. It is usually measured as mGy/min.
10
Measurement Quantities
Entrance Surface Air Kerma (Ka,e) –
It is the air kerma from the incident beam along the
central x-ray beam axis at the point where radiation
enters the patient and the effect of back scattered
radiation is included.
Given as Ka,e = Ka,i x B
B = Back Scatter Factor.
Unit = Joule per kilogram, Commonly referred to as
Gray & milli Gray (Gy or mGy).
Incident Air Kerma Rate quantifies air kerma per unit
time.
11
Risk Related Quantities
Absorbed dose – energy deposited per unit mass of a
material, in our case, within tissue.
Initially measured as rads
Current unit based on Systeme Internationale (SI unit)
SI Unit of Absorbed Dose = Gray
1Gray (Gy) = 100 rad
1rad = 10 mGy
12
Risk Related Quantities
Dose Equivalent – accounts for biological effect of type
of radiation
For example, difference in biological effect between
, and radiation
Radiation Weighting factor (wR) – scaling factor used
, Xray wR = 1
(wR) = 20
SI Unit is Sievert
1 Sievert (Sv) = 100 rem
1 rem = 10 mSv
13
Risk Related Quantities
Effective dose – accounts for radio-sensitivity of
specific organs
Includes
A tissue weighting factor (wT) for each sensitive organ
Each tissue included in the clinical examination (HT)
Effective dose = wT x HT, () summed over all
exposed organs.
SI Unit is Sievert
1 Sievert (Sv) = 100 rem
1 rem = 10 mSv
14
Background Radiation Exposure
Non-Medical Radiation
Source
Radiation Dose Estimate Equivalent Amount
Background Radiation
Natural background
radiation
3 mSv
Airline passenger (cross- 0.04 mSv
country)
3mSv/year*
4 days
* = estimate at sea level in US
15
Medical Radiation Exposures
Medical Radiation Source
Radiation Dose
Estimate
Chest x-ray
0.1 mSv
Equivalent Amount
Background
Radiation
10 days
Urinary tract fluoroscopy
(VCUG)
Continuous Mode*
0.45 – 0.59 mSv
2 months
0.05 – 0.07 mSv
1 week
Optimized fluoroscope*
* Ward et al Radiology 2008;249:1002
16
Practical Methods to
Reduce Radiation Dose to
Fluoroscopy Staff &
Patients
17
Staff Protection
18
Reduce Radiation Dose: Staff
Staff dose is due to scattered radiation
Scattered radiation is directly proportional to
Patient Dose
Patient Dose
Staff Dose
19
Staff Protection
Well fitted lead apron (knees)
Leaded glasses (with sides)
Thyroid shield
Lead gloves
20
Staff protection: Hands
Keep hands out of the beam
Collimate
21
Staff protection: Shields
Lead shield on tower
Do not turn your back
to Xray beam if
wearing front apron
only
22
In summary: Have we….
… left our hands in the beam?
… sacrificed personal safety for expediency?
… turned our unshielded backs to the X-ray source?
… unnecessarily prolonged exposure?
… pushed away a protective barrier?
23
Patient Protection
24
Patient Protection
Radiation dose is optimized when we use
Least amount of radiation
That delivers clinically adequate image
quality
Patient Positioning
Proper patient positioning
Make use of Inverse square law!
Maximize distance between x-ray tube & patient
Minimize distance between patient & Image
Intensifier
26
Control Fluoroscopic Exposures
Choose pulsed fluoroscopy
Choose as short a pulse width as possible
Typically 5 – 10 msec pulse width
27
Control Fluoroscopic Exposures
Continuous fluoroscopy
30 pulses per second
33 msec pulse width
Grid-controlled fluoroscopy
e.g. 15 pulses / sec
10 msec pulse width
28
Control Fluoroscopic Exposures
Increase filtration to reduce patient radiation dose
Balanced by need for shorter pulse widths to
freeze motion
Interposition of Aluminum and variable
thickness of Copper
Removes low energy radiation that does not
reach the image intensifier
scattered within the patient
adds radiation dose
does not contribute to image
29
Control Fluoroscopic Exposures
Remove anti-scatter grid whenever possible
Removes scattered radiation
Increased radiation dose
Not necessary in small patients
Avoid unnecessary magnification
30
Control Fluoroscopic Exposures
Collimate to area of interest
No need to radiate tissue that
is not clinically pertinent
31
Control Fluoroscopic Exposures
Use “last image hold”
Whenever you need to inspect the anatomy, and do not need to
observe motion or changes with time
Use Fluoroscopy Store (FS)
this method is ideal to convey and record motion, such as
peristalsis, or show viscus distensibility, as in esophagram
when you need information without excessive detail
Fluoro-grab
Exposure
32
Control Number of Images
Choose appropriate, patient-specific technique
Limit acquisition to what is essential
for diagnosis and documentation
PAUSE– Plan study ahead
PAUSE- think # frames / second
PAUSE – think magnification
PAUSE – think Last Image Hold
PAUSE – think Image Grab
33
Control Fluoroscopic Use
Use fluoroscopic examination when there
is a clear medical benefit.
Use alternative imaging methods whenever
possible
US
MRI
34
Special Pediatric Considerations
Pediatric patient management more critical
Increased radio-sensitivity, small size, longevity.
Pediatric size
Smaller patient leads to less scattered radiation
There is an increased need for magnification
35
Institutional Strategies to
Optimize Radiation
Exposure Fluoroscopy
36
To Start:
An in-house diagnostic medical physicist in
pediatric hospitals is optimal.
The physicist must have proper training and
background in Medical Physics, such as CAMPEP
accredited graduate and residency programs.
Proper training is key
37
To Start:
An Image Management committee, comprised of
radiologists, technologists, administrators and medical
physicists, under the direction of the department Chair,
can be very helpful.
•
Responsible for optimizing radiation procedures.
•
Oversee the departmental QA/QC program.
•
Meet criteria for accreditation, e.g. ACR
38
To Start:
Oversee purchase of capital equipment and periodic
hardware and software upgrades.
Staff training on state of the art technologies.
Technologists, radiologists
Equipment, safety, physics, radiation biology
Compliance with applicable state and federal
regulations.
39
Dosimetry Records
Manage fluoroscopy parameters
e.g., pulsed fluoroscopy, pulse rate, removable
grid
Record information related to patient
radiation dose as displayed by the
equipment:
Cumulative Dose Area Product.
Cumulative Air kerma/Skin Dose.
40
Summary
PAUSE to properly plan and prepare for study
Activate dose saving features of equipment
No image exposures unless necessary
Download image grab instead
PULSE at lowest possible rate
41
References
-Gelfand, D.W., D.J. Ott, and Y.M. Chen, Decreasing numbers of gastrointestinal studies: report of data from 69 radiologic practices.
AJR Am J Roentgenol, 1987. 148(6): p. 1133-6.
-Margulis, A.R., The present status and the future of gastrointestinal radiology. Abdom Imaging, 1994. 19(4): p. 291-2.
-Page, M. and H. Jeffery, The role of gastro-oesophageal reflux in the aetiology of SIDS. Early Hum Dev, 2000. 59(2): p. 127-49.
-Strauss KJ, Kaste SC. The ALARA (as low as reasonably achievable) concept in pediatric interventional and fluoroscopic imaging:
striving to keep radiation doses as low as possible during fluoroscopy of pediatric patients—a white paper executive summary.
Radiology 2006 240(3):621-622.
-Ward VL, Strauss KJ, Barnewolt CE, Zurakowski D, Venkatakrishnan V, Fahey FH, Lebowitz RL, Taylor GA. Pediatric radiation
exposure reduction and effective dose reduction during voiding cystourethrography. Radiology 2008 249:1002-1009.
-Hall, E. and J. Amato, Radiobiology for the Radiologist. 2005: Williams & Wilkins.
-Lederman, H.M., et al., Dose reduction fluoroscopy in pediatrics. Pediatr Radiol, 2002. 32(12): p. 844-8.
-Ward, V., et al., Radiation exposure reduction during voiding cystourethrography in a pediatric porcine model of vesicoureteral reflux.
Radiology, 2005. 235.
-Boland, G.W.L., et al., Dose Reduction in Gastrointestinal and Genitourinary Fluoroscopy: Use of Grid-Controlled Pulsed Fluoroscopy.
Am. J. Roentgenol., 2000. 175(5): p. 1453-1457.
-Brown, P.H., et al., A multihospital survey of radiation exposure and image quality in pediatric fluoroscopy. Pediatr Radiol, 2000. 30(4):
p. 236-42.
-Strauss KJ. Pediatric interventional radiography equipment: safety considerations. Pediatr Radiol (2006) 36 (Suppl 2):126-135.
-Hernanz-Schulman M, Emmons M, Price R. Radiation dose reduction and image quality considerations in pediatric patients.
Radiology RSNA syllabus, November, 2006
42