BME 311: BIOMEDICAL INSTRUMENTATION I Lecturer: Ali Işın

Download Report

Transcript BME 311: BIOMEDICAL INSTRUMENTATION I Lecturer: Ali Işın

FACULTY OF ENGINEERING
DEPARTMENT OF BIOMEDICAL ENGINEERING
BME 311: BIOMEDICAL INSTRUMENTATION I
Lecturer: Ali Işın
Lecture Note 4: ECG Systems
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
1
ECG Systems
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
2
What is an ECG?
The electrocardiogram (ECG) is a representation
of the electrical events of the cardiac cycle.
Each event has a distinctive waveform, the study
of which can lead to greater insight into a
patient’s cardiac pathophysiology.
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
3
The Normal Conduction System
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
4
The Normal Conduction System
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
5
ECG
ECG is a surface measurement of the
electrical potential generated by
electrical activity in cardiac tissue.
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
6
ECG
The human heart can be considered as a
large muscle whose beating is simply
muscular
contraction.
Therefore
contractions of the heart cause a potential
to be developed. The measurement of the
potential produced by cardiac muscle is
called electrocardiology.
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
7
What types of pathology can we identify and
study from ECGs?
•
•
•
•
•
Arrhythmias
Myocardial ischemia and infarction
Pericarditis
Chamber hypertrophy
Electrolyte disturbances (i.e. hyperkalemia,
hypokalemia)
• Drug toxicity (i.e. digoxin and drugs which
prolong the QT interval)
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
8
Waveforms and Intervals
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
9
2. Ventricular
depolarization
3. Ventricular repolarization
1. Atrial
depolarization
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
10
Different Segments of ECG
P wave: the sequential activation (depolarization) of the right and left atria
QRS complex: right and left ventricular depolarization (normally the ventricles
are activated simultaneously)
ST-T wave: ventricular repolarization
U wave: origin for this wave is not clear - but probably represents
"afterdepolarizations" in the ventricles
PR interval: time interval from onset of atrial depolarization (P wave) to onset
of ventricular depolarization (QRS complex)
QRS duration: duration of ventricular muscle depolarization
QT interval: duration of ventricular depolarization and repolarization
RR interval: duration of ventricular cardiac cycle (an indicator of ventricular
rate)
PP interval: duration of atrial cycle (an indicator or atrial rate
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
11
ECG basics
•
•
Amplitude:
Bandwidth:
•
Largest measurement error sources:
– Motion artifacts
– 50/60 Hz powerline interference
•
Typical applications:
– Diagnosis of ischemia
– Arrhythmia
– Conduction defects
1-5 mV
0.05-100 Hz
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
12
ECG Leads
Leads are electrode setups which measure the
difference in electrical potential between either:
1. Two different points on the body (bipolar leads)
2. One point on the body and a virtual reference point
with zero electrical potential, located in the center of
the heart (unipolar leads)
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
13
Cardiac Axis by Different Leads
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
14
Einthoven Triangle:
Note potential difference
for each lead of triangle
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
15
Each lead gives a slightly different
representation of electrical activity of heart
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
16
ECG Leads
The standard ECG has 12 leads:
3 Standard Limb Leads
3 Augmented Limb Leads
6 Precordial Leads
The axis of a particular lead represents the viewpoint from
which it looks at the heart.
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
17
Standard Limb Leads
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
18
Precordial Leads
Adapted from: www.numed.co.uk/electrodepl.html
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
19
12-Lead ECG measurement
•
•
Most widely used ECG measurement setup in clinical environment
Signal is measured non-invasively with 9 electrodes + 1 reference electrode (right
leg)
Einthoven leads: I, II & III
Goldberger augmented leads: VR, VL & VF
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
Precordial leads: V1-V6
20
12 Lead ECG System
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
21
12 Lead ECG System
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
22
Block Diagram of a Basic ECG System
Lead Selector
Isolated Power
Source
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
23
ECG Pre-Amplifier
ECG pre-amplifier is a bioelectric differential amplifier.
It includes;
• High impedance input of bioelectric amplifier
• Lead selector switch
• 1mV calibration pulse source
• Means of protecting amplifier from high voltage
discharge such as a defibrillator discharge used on a
patient
• Pre-Amplifier contains an instrumentation amplifier
as well as an isolation amplifier for patient safety
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
24
Isolation Amplifier
• Needed for safety! Isolates the patient from high voltages
and currents to prevent electric shock by introducing a
specific barrier between passage of current from the power
line to the patient.
• Can be done by using light (photo emitter and photo
detector) or a transformer (set of inductors that are used in a
step up / step down configuration)
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
25
Common Mode Rejection
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
26
ECG Amplifier Circuit With Right Leg
Driver
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
27
Common Mode Voltage (CMV)
• If 2 inputs are hooked together into a differential
amplifier driven by a common source with respect to
ground, then the common mode voltage should be
the same and the ideal output should be zero.
However practically you will see a voltage.
• CMV is composed of 2 parts:
– DC electrode offset potential
– 50Hz AC induced interference caused by magnetic and
electric fields from power lines and transformers
• Capacitively coupled into circuit
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
28
Analysis to reduce noise in ECG
• Right leg driver circuit is used in a feedback
configuration to reduce 50 Hz noise and to drive the
noise on the patient to a lower level.
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
29
“6” ways to reduce Noise in ECG
•
•
•
•
•
•
Common Mode Rejection (differential Amplifier)
Right Leg Drive (feedback loop to decrease noise)
Shielding of wires
Isolation amplifier
Notch filter to reduce 50 Hz noise
Bandpass filter to reduce noise below and above ECG
bandwidth (0.05-100 Hz)
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
30
High Pass Active Filters
Attenuates frequencies lower than the cutoff frequency.
cutoff frequency is 1/(2) =1/ 2RiCi
Rf
Ci
Ri
-
A
Vinput
Vinput
Ci
+
Ri
Ii
0
Voutput
Rf
Voutput
IRf
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
31
Low Pass Active Filters = Integrator
Cf
Attenuates frequencies higher
than the cutoff frequency
Rf
Ri
-
A
Vinput
+
Voutput
Cf
Vinput
Ri
0
Ii
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
Rf
ICf
Voutput
IRf
32
Defibrillator Protection Circuit
• Defibrillator = a high voltage electrical heart
stimulator used to resuscitate heart attack victims
• When the physician uses defibrillator, the high
voltages and currents discharged onto patient can
cause damage to medical equipment(specificly preamplifier circuitry), BUT physician still needs to view
ECG of the patient while defibrillating.
• How do you protect your medical equipment from
excessive voltages and currents?
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
33
Protection Devices in ECGs: Neon Glow Lamps
• Neon Glow Lamps are pair of electrodes mounted in a glass envelope
filled with low pressured neon gas or a mix of other inert gases.
• Normally impedance across the electrodes is very high (so lamps are not
conductive) but when the potential across the electrodes reaches to the
ionization point of the gas, impedance suddenly drops making the
lamps conductive (they light up) creating a short circuit to ground,
transferring excess energy safely to the ground without damaging the
amplifier.
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
34
Protection Devices in ECGs: Zener Diodes
• Zener Diode: is a diode which allows current to flow in the forward
direction in the same manner as an ideal diode, but also permits it to
flow in the reverse direction when the voltage is above a certain
value known as the breakdown voltage.
• In this configuration when a larger potential than the breakdown
voltage is applied to the system (i.e. Defibrillator discharge) the
zener diode allow the current to flow in reverse direction and shunts
it to the ground thus preventing any damage to the amplifier.
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
35
Protection Devices in ECGs: Current-Limiting Diodes
• Current Limiting Diodes are electronic devices that limit current to a
maximum specified value for the device.
• These diodes consists of a n-channel JFET-transistor
• They act as a resistor as long as the current level remains below the
limiting value. If current tends to rise above the limit (as in the case of a
Defibrillator discharge), it will be clamped and limited preventing any
excess current reaching the amplifier thus protecting it.
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
36
Electro-Surgery Unit (ESU) Filtering
• During surgery simultaneous ECG recording is also
required.
• ESU can introduce high frequency signals with
frequencies ranging from 100KHz to 100 MHz and
with magnitudes up to few kVolts into the ECG signal.
This interferance can heavily distort the ECG signal.
• ESU introduces:
– DC offsets
– and obscures the ECG signal
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
37
• Even though the ECG frequency range is between 0.05-100 Hz
and all the higher frequency signals are attenuated by using
low-pass filters at filtering stage (slide 32), ESU signal still
manages to disort ECG.
• Because;
 Internal junction points of the ECG amplifier can rectify high
frequency signals like ESU signal and the parasitic capacitance
between the leg connections of the amplifier further filters
this signal to create a dc ofset potential. (Whenever the ESU
device is triggered during the surgery ECG baseline oscilates
up and down.)
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
38
 And the ESU signal can leak through amplifier and
low pass filter layers obscuring the ECG signal.
• ECG needs to be of diagnostic quality so the ESU
noise should be eliminated
• Common technique is to use a pi-type 3-layered
RC filter
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
39
RC Filters
Vs
Frequency
FH
• Low Pass Filters will pass frequencies lower than cutoff frequency of FH
=1/2RC
Vs
FL
•
High Pass Filters will pass frequencies greater than cutoff frequency of FL
=1/2RC
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
40
3-Layered RC ESU Filter
RC Filter
RA
Defibrillator
Protection
Circuit
ECG
Differential
Amplifier
LA
LC filter design can also be used
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
41
Some Sample ECG Circuits
High-precision analog front end of a portable ECG application. 3 patient
electrodes (only 3 bipolar leads)
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
42
• ECG system with 5-patient electrodes (7-leads)
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
43
Standalone ECG with Compact Design
BME 311 LECTURE NOTE 4 - ALİ IŞIN, 2014
44