Transcript Slide 1
Vital Signs Monitor
UConn BME 4900
Vital Signs Monitor
Purpose
As the population ages, many
people are required by their
doctors to take vital signs on a
daily basis.
Developed for the average
person to use at home.
Vital signs transmitted to a
computer and ultimately to a
doctor.
Measurements
What can it measure?
Electrocardiogram
(ECG)
Three leads (Right
arm, left arm, right
leg)
Respiration
Thermistor
Body Temperature
NTC Thermocouple
What is an ECG?
Records a time waveform of heart electrical
activity.
Used to diagnose heart problems
Arrhythmia
Myocardial infarction
Conduction blocks (bundle branch block)
ECG Trace
Cardiac Electrical Activity
The ECG Waveform
Measuring Electrical Activity
3 leads
2 arm
1 leg
As cardiac muscle depolarizes it creates
a potential on the skin
Potential measured as a vector
Difference between right and left
electrode measured with a differential
amplifier.
Measuring Electrical Activity
Circuit consists of:
3 op amp differential amplifier
Low pass filter (Cutoff: 0.05 Hz)
High pass filter (Cutoff: 160 Hz)
Gain Amplifier
Differential Amplifier
Feedback
Amplifier
Buffer
What does the buffer do?
Provides a high input impedance
What is the purpose of C2 and C3?
Remove DC offset
Why?
Offset would be amplified by 1000x
10 mV at 1000x gain = ~10V
Differential
Amplifier
Differential Amplifier
V1 & V2 are in the
microvolt range
We need it to be in the
milivolt range so Rf/R1 is
set to ~1,000
1st Order LPF
The ECG is known to be a
low frequency signal
A LPF can be used to
remove the high frequency
noise in the signal
High Pass Filter
ECG frequency range ~0.05 Hz – 150 Hz
Why do we need to eliminate frequencies <0.05
Hz?
Avoid distortion of the ST wave
Respiration
Measured with thermocouple placed in nostril
During exhalation warm air passes through nose
During inhalation cooler air is drawn in
Respiration
What is a thermocouple?
Two different metals joined together
Temperature changes induce a voltage
Voltage can be linearized to temperature
Respiration
Non-inverting amplifier
Gain = (1 + R2 / R1)
Multiplies thermocouple
voltage by gain
R2
R1
Body Temperature
Often indicates infection
“Normal” temperature – 98.6ºF (37ºC)
Measured with negative temperature coefficient
(NTC) thermistor
Resistance decreases as temp. increases
Body Temperature
Wheatstone Bridge
Rx = Thermistor
R1, R2, R3 = 4.7k Ohm
Differential amplifier
Body Temperature Waveform
Voltage decreases over
time
Time shown at right ~90
seconds
Voltage at steady-state
can be converted to a
temperature
measurement
Microprocessor
Provides Analog to Digital (A/D) conversion
Waveform y(t) sampled at a fixed rate Δt
Voltage read every Δt and converted to a number
If resolution is 8 bits
8
Gives us 2 = 256 counts over 5V range
~19.5 mV per count
Digital Isolator
Isolates the patient from the computer
Receives data from microprocessor
Uses Giant Magnetoresistance for isolation
Digital pulse induces a magnetic field across an isolated
barrier
RS232 Converter
Input from isolator
Converts 5 V UART signal from microprocessor
to RS-232 standard
RS-232 standard
+10 V – 10 V digital signal
Switching Between Signals
Three signals that can be measured
Only one is recorded at a time
A mode switch is used to select which signal is
processed
When a mode is selected a flag is sent over the serial
port
Flag indicates which mode
Labview reads flag and changes modes
Labview Software
Labview Requirements
Receive serial data from microprocessor
Read mode flag and switch to appropriate screen
Display EKG, and Respiration waveforms
Calculate respiration rate (number of peaks / time)
Read voltage from thermometer and convert to
body temperature