1Introductiona
Download
Report
Transcript 1Introductiona
MRC SEM Workshop
Introduce the FESEM microscope
Familiarize and Train experienced persons
to use the Hitachi S4700 FESEM
Obtain a better understanding on how to
utilize the FESEM
Workshop Outline
Introduction to the FESEM
Sample Contamination
Low Voltage SEM
High Resolution Imaging
X-ray Microanalysis
Sample Preparation
MRC Lab Procedures
Operation of the S4700 FESEM
Hands-on labs and testing for users
Introduction to the FEGSEM
The FEG SEM offers high performance
not just high resolution
This means large probe currents (up to a few
nanoamps), and small diameter electron probes
(from 1 to 3nm), over a wide energy range (130keV)
The FEG SEM package involves both the gun
and the probe forming lens
The Electron Gun
The device which provides the electron beam is
the called the ‘gun’
This is the single most important component of
the SEM because it determines the level of
performance that can be achieved
Electrons can be produced in several different
ways ....
Electron Sources
W hairpin - 50µm
diameter
LaB6 - 5µm
Thermal FEG - 250Å
Cold FEG - 50Å
Nano-FEG - 5Å
Nano tips - atom sized FEG
Nano-tips are field emitters in
which the effective size of the
tip has shrunk to a single
atom.
They can be made by
processing normal tungsten
FE tips or from Pt-Ir, or from
carbon nanotubes
They have exciting properties
and may be part of upcoming
SEMS but now they are still
only a laboratory curiosity
Etched
tungsten
tip
Cut PtIr tip
Comparing emitters
The various types of electron emitters can be
compared by looking at three parameters brightness, source size, energy spread
Other quantities are also important - e.g
vacuum required, lifetime, cost, expected mode
of use of SEM
Source Size
…is the apparent
size of the disc from
which the electrons
come
Small is good - for
high resolution SEM
less demagnification
Big is sometimes
good - e.g. for large
probe sizes and high
beam currents
The physical size of the tip is
not necessarily the same as the
source size
Energy Spread
Electrons leave guns
with an energy
spread that depends
on the cathode type
Lens focus varies
with energy
(chromatic
aberration) so
energy spread spoils
high resolution, and
low energy, images
W hairpin 2.5eV
LaB6
1.0eV
Schottky
0.75eV
Cold FEG 0.35eV
Summary of Electron Guns
The cold FEG offers the best performance parameters in all three
categories for most purposes
FEGs are best for high resolution, and low voltage operation
Thermionic emitters have advantages when very high beam
currents and large spot sizes are required.
Brightness
At a typical imaging
current FEG SEM spot
size is set only by
lens quality
Lower brightness
guns must use bigger
spots to give same
beam - this is
brightness limited
imaging
S4700 Resolution
Performance
The best resolution
is always obtained at
the smallest working
distance (WD)
..but the minimum
WD value varies with
beam energy
At the eucentric/EDS
WD of 12mm high
quality imaging +
analysis is possible
on the 4700
Sharpness, Contrast, Depth of
Field
These are dependent upon the three major
electron-beam parameters:
Electron Probe Size dp
Electron Probe Current
ip
Electron Probe convergence angle
ip
ap
dp
ap
Beam Performance
For the highest resolution beam diameter , dp , must
be as small as possible
For the best image quality and x-ray analysis,
emission current, ip, must be as large as possible
For the best depth of field convergence angle, ap,
must be as small as possible
When dp and ap are made small, ip is also reduced
Imaging modes
On the S4700 the convergence
angle a is set by the operating
mode of the microscope
No manual adjustment to the
condenser aperture strip is
required
Don’t change the aperture!
Gun behavior
The tip must be atomically clean to perform
properly as a field emitter
Even at 10-7 torr a monolayer of gas forms in
just 1 sec so the tip must be cleaned
periodically
It is cleaned by ‘flashing’ - heating the tip to
white heat for a few seconds. This burns off
(desorbs) the gas
Flashing
The flash condition is set up at the factory
Each tip should show a consistent emission
current when it is flashed
Compare the tip current with its own usual
value not with that from other tips
Excessive flashing may blunt the tip
Gas production
The tip gets dirty...
Gas molecules are
desorbed from 1st
anode by electrons
Some of these stick
on the tip making it
less sharp
This causes the
emission current to
fall over time
The life cycle of an FEG tip
Typical characteristics
The tip is usually covered with a mono- layer of gas
after 5-10 minutes of use
The emission then stabilizes for a period of from 2
hours (new machine) to 8 hours (mature machine).
On this S4700 the tip must be re-flashed after 92
hours of operation (the software gives a warning)
On the plateau, or stable, region the total noise + drift
is only a few percent over any period of a few
minutes
The secret of successful
Field Emission Microscopy
Run the tip for at least a few hours every day
even when the microscope is not otherwise
in use
This keeps the first anode - which is the main source
of gas - clean, reducing noise and drift.
Other care
Bake the system often - at least
every six months - and on public
holidays, long weekends
If tip noise is increasing and a
bake is not possible raise the
emission current to 20 or 30µA
with the beam in the ‘freeze’
position for a couple of hours or
more for Field Emission bombard
Drift in microanalysis
For normal EDS analysis drift is not a problem
For quantitative analysis using standards, and
for line scans and X-ray maps which take
significant time to record
Drift can be an issue, this is a reason to sue a
thermonic SEM for quantitative analysis
Hidden benefits of FEGSEM
Reliability and reproducibility - no need to
change tips or break vacuum. Control by
computer ensures reset table values
Ease of use - one button operation, memory
alignment settings
Longevity - with reasonable care time between
tip changes 3-5 years (even with students
operating)
We are on the NINTH year
What isn’t good?
The range of beam
currents available is
limited when
compared to that
from a thermionic
emitter so an FEG is
not ideally suited to
such tasks as WDS
where high currents
are needed