Transcript Slide 1
MedeCal Pulse Oximeter Probe
Team Members: Bogdan Dzyubak, Joe Helfenberger, Jonathan Meyer, Matthew Parlato
Advisor: Dr. John G. Webster, PhD.
Clients: Amit Nimunkar and Jonathan Baran
Introduction
The Problem
Hospitals in the developing world
often lack necessary medical
equipment, such as pulse
oximeters. Donated pulse
oximeters are rare, and often
cannot be repaired or replaced
when they break. The only lowcost probes currently on the
market are disposable.
Design Specifications
A Global Healthcare Perspective
United
States
Money spent on
healthcare per
capita
Money spent on
healthcare as a %
of GDP
Life Expectancy
$6096
Most
Most
Developed Developing
Countries Countries
$1,569
$48.86
15.4%
7.0%
5.57%
77.4
65-81
39-61
The primary goals of this project were as follows:
The probe must be durable, and preferably easy to repair, as many
hospitals can only afford to buy one.
The probe must be versatile, usable with a wide range of age and
ethnic groups.
The final product must cost less than $8.
The probe must interface with the MedeCal.
The probe must be convenient to use, to prevent misuse and disuse.
Oxygen saturation accuracy must be within 5%, preferably 3%.
Figure 1: Comparison of income and life expectancy [1]
Objective
To develop a durable, low-cost pulse oximeter probe to
interface with the MedeCal, for use in developing countries.
The MedeCal
The MedeCal is a low-cost medical computer being developed by our
clients Amit Nimunkar and Jonathan Baran, in conjunction with
Engineering World Health, for use in developing countries. It will analyze
and display the probe’s output, and power its optical elements.
Wires to
MedeCal
In tissues, blood volume regularly fluctuates
with each heart beat. This volume change
affects the amount of light that transmits
through the tissue. Monitoring blood flow
by measuring fluctuations in light
transmittance is called
photoplethysmography (Figure 3).
Rubber end pieces
Oximetry
Oximetry is the measurement of oxygen
saturation, which is the percent of
oxygenated hemoglobin. Since oxygenated
and deoxygenated hemoglobin have unique
light absorption curves, it is possible to
determine oxygen saturation by comparing
light absorption at two wavelengths (red
and infrared). Photoplethysmography is
used to isolate the blood absorbance from
tissue absorbance.
Optical Elements
Inside
The probe we constructed is based on a clothespin. Its two prongs
are connected with a plastic spring. Rubber sleeves protect the
optical elements at the ends of the prongs from mechanical
damage and ambient light. The probe is light enough to hang from
the ear without exerting excessive pressure, thus avoiding
pressure sores, and a decrease in signal quality.
Red
Infrared
Figure 4: Absorption curves of oxyhemoglobin and
deoxyhemoglobin.
It has been tested on a range of subjects
with different ear and finger sizes,
morphologies, and skin colors. A good
signal can be achieved in all of these
cases with gain adjustment.
Cost
Photodiode
Two LEDs (on the same
mount)
$0.72
~ $0.33 x 2
Rubber sleeves
Total
$1.00
$5.88
($15.00 for
prototype)
Plastic clip
$1.50
Cord and serial connector $2.00
Future work will include improving the probe’s mechanical design, and testing its
performance.
Mechanical improvements will include:
Reducing the probe’s spring constant, for long-term monitoring.
Replacing the plastic spring with a metal spring, for improved durability.
Implementing calibration resistors into the probe to correct for deviation in LED
wavelengths. This will require modifications to the clients’ circuit.
Our probe can be used on both the ear and the finger. This greatly
increases the range of patients on which it can be used. The ear
offers a reliable blood supply, even under critical conditions, while
the finger is more convenient.
Figure 3: Photoplethysmograph [4]
The probe operates in plethysmography
mode yielding a clear waveform on both
the finger and the ear, with both red and
infrared.
Item
Future Work
The MedeCal will be a central processor for multiple medical devices in
addition to the pulse oximeter, such as low-cost digital thermometers,
electrocardiograms (ECGs), and digital spirometers.
Theory
Plethysmography
The designed probe costs under $8, thus
satisfying our goal.
The probe has been worn for an hour without significant discomfort.
Probe Design
Modified
Clothespin
Results
The optical elements consist of two LEDs (red and IR) and a
photodiode. The photodiode amplifies the signal internally and
returns a voltage signal to the MedeCal.
Performance tests will include:
The accuracy of oxygen saturation measurements will be compared with other
pulse oximeters on the market. This will be possible when the clients’ circuit is
completed.
The probe’s durability will be tested by subjecting the probe to cyclic
deformation until failure.
A larger range of ear and finger sizes will be tested, including:
• Infants through adults with thicker ears
• Different ear morphologies
More tests on people with different skin colors will be conducted.
The effects of different cardiovascular conditions will be investigated:
• Decreased perfusion due to cold
• Oxygenation below 70%, achieved by breathing an air-nitrogen
mixture
Acknowledgements
Dr. John Webster (mentorship), Chris Esser (circuit design), Eamon Bernardoni
(mechanics consultation), Dr. Rod Lakes (mechanics consultation), Amit Nimunkar
and Jonathan Baran (encouragement and advice).
References
[1] United Nations. (2007/2008). Human Development Report 2007/2008. Retrieved March 12, 2009, from Table 6: Commitment to
Health: Resources, Access, and Service: http://hdr.undp.org/en/media/HDR_20072008_EN_Complete.pdf
[2] Engineering World Health. (n.d.). Engineering World Health Website. Retrieved March 12, 2009, from Engineering World Health
Website: http://www.ewh.org
[3] Fortney, L. (2009, January 31). Dr. (D. Team, Interviewer)
[4] “Figure 1.” (2008) Accessed April 30. 2009, at http://bobjunior.com/wp-content/uploads/2008/01/figure1.JPG