Transcript Slide 1

The bird IPPB machine
• The Bird is pneumatically powered &
controlled.
• it can be triggered [1] by negative
pressure [2] by time or [3] manually.
• It is pressure limited, changing the
breath as pressures rise
• It is pressure or time-cycled off
Ambient
chamber
Sensitivity
control
Gas inlet
From 50 psig
Pressure
chamber
Pressure
control
Nebulizer
drive line
Main flow
drive line
•
•
•
•
•
A flow rate control
B ceramic switch
C restricted orifice for entrainment
D nebulizer drive line to in-line SVN
E exhalation valve [mushroom valve]
left
This is a simplified picture of the ceramic switch and what makes
it move to start or end a breath.
On the left side of the screen is the AMBIENT CHAMBER [see the
hole toward the outside?]
On the right side of the screen see the PRESSURE CHAMBER,
[see the outlet going to the patient circuit?]
In A you see that the diaphragm is mid-line, the ceramic switch is
closed. No air moves
In both the AMBIENT CHAMBER and in the PRESSURE
CHAMBER, there is no extra pressure, both chambers are at
zero [ 1 atmosphere]. There is no gas flow and the ceramic switch
is closed.
This negative pressure pulls the clutch plate toward the magnet in
the PRESSURE CHAMBER & when it gets close enough, the
magnet pulls the clutch plate closer.
The diaphragm moves toward the patient, to open the ceramic
switch.
Gas flows into the PRESSURE CHAMBER, into the circuit and into
the patient’s lungs.
In C, the diaphragm stays toward the magnet
As long as the pressure in the PRESSURE CHAMBER is lower
Than the magnet force in the AMBIENT CHAMBER …the breath
continues
We can control how deep a breath the patient takes by altering
the positive pressure needed to overcome the magnet that will
stop the breath
As the patient’s lung fills with gas, the pressure builds up in the
lung, then in the circuit and finally in the PRESSURE CHAMBER
Once high enough, the positive pressure in the PRESSURE
CHAMBER pushed the diaphragm toward the AMBIENT
CHAMBER and flow stops--- the breath has stopped.
On inspiration
ceramic switch is
open, gas flows to
the patient whose lung
fill
On exhalation: the ceramic
switch is closed, gas stops
flowing and the patient exhales
into the exhalation valve
On inspiration, the gas flows from the
top into the Pressure chamber, into
the circuit.
At the same time, the exhalation
valve is closed by the same pressure
As long as there is pressure building
up in the pressure chamber, the
exhalation valve is closed.
Gas can move only into the lung
On exhalation phase: the
pressure drops in the pressure
chamber and the exhalation
valve is open to the atmosphere
so that the patient exhales
into the atmosphere
The ceramic switch is closed
and there Is no gas flow into
any chamber.
Air mix: entrainment devices
During inspiration, 02 from the
ceramic switch moves into the
line that powers the entrainment
device.
The flow goes through
the restricted orifice decreasing lateral
wall pressure in the ambient chamber
so that it pulls air from the
ambient chamber into the gas flow
going to the pressure chamber.
What do you suppose happens to
the Fi02 and to the total flow down
the main flow drive line as the pressure
rises inside the pressure chamber?