Cleaning of Hydrocarbon Buildup on Metallic-oxide Thin

Download Report

Transcript Cleaning of Hydrocarbon Buildup on Metallic-oxide Thin

Comparison of Cleaning
Methods for Thin Film Surfaces
Lena Johnson, Mitch Challis, Ross Robinson,
Richard Sandberg
Group 4
English 316 Oral Report
June 16, 2003
Applications for EUV Light Research
Multilayer Mirrors
EUV Astronomy
www.lbl.gov/Science-Articles/Archive/xray-inside-cells.html
EUV Lithography
EUV Microscopes
Images from www.schott.com/magazine/english/info99/
How small is that?

Want to reflect EUV (extreme
ultraviolet) light (50-5 nm or 10-7 cm)
Image from http://imagers.gsfc.nasa.gov/ems/waves3.html
The Problem
 Organic
contaminants collect on the
surface (Junk).
– Interferes with measuring optical
constants
 Example
with visible light
Question:
How do we effectively remove contaminants?
Outline
1.
2.
3.
Measurement
methods
Cleaning methods
-Opticlean®
-Plasma Etch
-UV lamp
Conclusions
Video
 Measurement
method
– Ellipsometry
 Cleaning
methods tested
– Opticlean
– UV Lamp
– Plasma Etch
Short Exposure Contamination Experiment (10 sec)
Additional Thickness (Ǻ)
450
400
350
300
250
450
200
150
100
143
18
50
16
0
1
Spit on
Touched with bare fingers (10s)
Dipped in Deionized Water
Touched and rubbed with latex glove (10s)
Opticlean® Process leaves
hydrocarbon residue
• Opticlean® significantly
removes contaminants,
but leaves a residue
http://www.dantronix.com
• Ellipsometric Results
Opticlean® residue
thickness on two runs:
1) 17 Ångstroms
2) 22 Ångstroms
•
•
Opticlean® made by Dantronix Research
and Technologies, LLC
Ellipsometer Type: J.A. Woollam Co., Inc
Multi-Wavelength Ellipsometer (EC110)
http://www.jawoollam.com/
Does the Opticlean®
process damage
thin films?


Scanning Electron
Microscope showed no thin
film damage, nor trace of
materials used in thin films
on pulled of Opticlean®
films (U, Sc, Va).
X-Ray Photoelectron
Spectroscopy found no
trace of materials used in
thin films on pulled of
Opticlean® .
XPS revealed components of Opticlean®,
but not heavier metals used in thin films.
Prominent thin-film lines: U-380 eV, V-515 eV, Sc-400 eV
684 eV (F 1s)
527 eV (O 1s)
Photons (Counts)
281 eV (C 1s)
97 eV (Si 2p)
148 eV (Si 2s)
Photon Energy (eV)
Cleaning residue with “Matrix”
Plasma Etch



RF Plasma Etching with
O2 Plasma
– 0.120 Torr Pressure
– 250W RF (max 350)
– 0.75 SCCM O2 flow
– No extra heat applied
Good for removing
polymers, but not bulk
contaminants (i.e. Dust)
Oxygen plasma plus
mechanical sputtering
removes surface layers
RF
Plasma Results
Matrix System
Change in thickness (Å)
2
0
-2 0
2
4
6
8
10
12
-4
-6
-8
-10
-12
Minutes In Plasma
14
16
18
UV Lamp
UV Results
Change in Thickness (Ǻ)
UV Lamp
2
0
-2
0
5
10
15
-4
-6
-8
Time Under UV Lamp (s)
20
25
Conclusion
Method
Effectiveness
Cleaning Time
Ease of Use
Notes
Opticlean®
Left residue
Must wait for
polymer to cure
Can be difficult to
peel off.
Good for dust
etc..
Oxygen Plasma
Effective. Cl not
completely removed
Setup takes a few
minutes. Clean
under a minute
Equipment in clean
room. Complex to
setup.
Builds up
silicon dioxide
Eximer UV Lamp
Effective
1 to 5 minutes
Very easy
Less silicon
dioxide
buildup then
plasma.
Opticlean® +
Oxygen Plasma
Effective Possible 1
Å residue or oxide
Long, plasma
setup and polymer
cure time.
Complex clean
room equipment
and skill needed to
peel

Recommended Procedure:
– UV lamp for 5 minutes
Conclusion
Method
Effectiveness
Cleaning Time
Ease of Use
Notes
Opticlean®
Left residue
Must wait for
polymer to cure
Can be difficult to
peel off.
Good for dust
etc..
Oxygen Plasma
Effective. Cl not
completely removed
Setup takes a few
minutes. Clean
under a minute
Equipment in clean
room. Complex to
setup.
Builds up
silicon
dioxide
Eximer UV Lamp
Effective
1 to 5 minutes
Very easy
Less silicon
dioxide
buildup then
plasma.
Opticlean® +
Oxygen Plasma
Effective Possible 1
Å residue or oxide
Long, plasma
setup and polymer
cure time.
Complex clean
room equipment
and skill needed to
peel

Recommended Procedure:
– UV lamp for 5 minutes