Poster - Indico

Download Report

Transcript Poster - Indico

A Range Verification Method for Proton Therapy
using a Photon Counting Detector
Jin Sung Kim1, Su Jung An2,3, Yong Hyun Chung2,3*
1Department
of Radiation Oncology, Samsung Medical Center, Seoul, 135-710, Korea
2Department of Radiological Science, College of Health Science, Yonsei University, Wonju, 220-710, Korea
3Institute of Health Science, Yonsei University, Wonju, 220-710, Korea
C. Simulation setup for in-beam PET imaging
INTRODUCTION
•
In the past few decades, proton radiotherapy has evolved into a widely practiced modality for cancer treatment.
• The main advantage of proton beam is its capability to irradiate a highly conformal dose distribution in a given
target volume, due to its unique interaction and energy deposition.
• The deposited dose reaches a maximum called the Bragg peak beyond which it sharply falls to zero.
• Owing to the typically steep dose gradient at the distal edge of each individual proton beam, however,
uncertainties in treatment planning and treatment delivery can have a significant influence on the delivered 3D
•The detector system consists of two opposing parallel-planes of CZT detectors to measure proton range
verification.(figure1(c))
•To verify proper detection of annihilation gamma , several thickness of CZT crystals were used. (Table 1)
•The distance between the detector heads was fixed at 20 cm.
•Using phase space file as input, e+ sources were emitted from PMMA phantom using particle filter in
GATE 6 actor toolkit.
•The 2D image of positron emitting radionuclide in the phantom was acquired by plotting the coincidence
counts per opposing detector pixel pair.
dose distribution.
• This study presents the results of the vary first multi-modality imaging system for X-ray and gamma-ray
coincidence imaging using a single CdZnTe detector to measure proton range verification.
RESULTS
• The detector system consists of two parallel planes of detectors and an X-ray generator.
• An X-ray image is acquired using one detector for the verification of 2-dimensional anatomical structure of the
A. Generation of positron emitting radio-nuclides
patient.
• 110 and 140 MeV proton beams, cylindrical tissue phantom, and two rectangular CdZnTe detectors were
modeled, and the imaging performance of this system was evaluated using GATE simulation
• The results showed the potential benefits of a X-ray/gamma-ray imaging with a single detector for range
Dose
Relative counts of total nuclei
verification in proton therapy
MATERIALS & METHODS
Relative counts
a. 110MeV
Dose
Relative counts of total nuclei
b. 140MeV
1.0
1.0
1.0
1.0
0.8
0.8
0.8
0.8
0.6
0.6
0.6
0.6
0.4
0.4
0.4
0.4
0.2
0.2
0.2
0.2
0.0
0
20
40
60
80
100
120
140
0.0
160
0.0
0
20
Depth position (mm)
40
60
80
100
120
140
Normalized dose
• Image registration is intrinsic because the X-ray and gamma ray images are acquired in the same geometry.
Relative counts
proton penetration.
•Figure 2 shows the simulated Bragg curve and spatial distribution of the positron emitting nuclei along
the depth direction for 110 and 140 MeV proton beams.
•This distribution shows a nearly flat plateau, dropping to zero in close proximity to the Bragg peak.
•The systematic distance between the 50% level of the total activity and dose maximum in PMMA was
observed as about 6 mm.
•This results were subsequently used as the source distribution on the PET simulation.
Normalized dose
• The paired gamma rays from the annihilation are imaged with two modules to determine the maximum range of
0.0
160
Depth position (mm)
Figure 2.The total isotope distribution from GATE6 simulation
B. X-ray imaging with CZT photon counting detector
a. PMMA phantom
c. Coincidence imaging
b. X-ray imaging
•Figure 3 shows the spectral x-ray image of the PMMA phantom with CZT photon
counting detector.
•1 mm-thick CZT detector is usable for x-ray imaging to see anatomical structure of
the object.
Figure 1. Schematic diagram of a simulation setup
Table 1. Thickness of CZT detector used in simulation
Figure 3. Simulated X-ray image using CZT photon counting detector
Thickness of CZT detector
C. In-beam PET imaging with parallel-planes of CZT detector
DETECTOR A
1 mm CZT + 5 mm CZT
DETECTOR B
1 mm CZT + 10 mm CZT
DETECTOR C
1 mm CZT + 15 mm CZT
•The simulated 2D images of positron emitting radionuclides in the phantom were illustrated in Figure 4
for 110 and 140 MeV proton beams.
•The results demonstrate the better image quality from a thicker detector.
•Simulation has proven that the proton therapy dose distribution could be imaged by CZT pair detectors.
A. Generation of positron emitting radio-nuclides
Thickness of CZT
detector
•
PET image
PET + radiography image
140 MeV
PET image
PET + radiography image
GATE(Version 6.0) was used to simulate the generation of positron emitting radio-nuclides by proton
irradiation in PMMA cylindrical phantom with a height of 20 cm and a diameter of 15 cm.
•
110 MeV
Cylindrical phantom contains 5 balls (Aluminum, Glass, Spinebone, PVC, Water) that were arranged as shown
DETECTOR A
(1 mm CZT + 5 mm CZT)
in figure 1(a).
•
Two monoenergetic pencil-like proton beam were used.
(energy: 110 and 140 MeV; FWHM: ~10 mm; intensity : 1 x 108 p/s)
•
The passage of protons through the phantom was simulated through physical interactions and the distribution
of generated positron emitting isotopes was calculated.
•
DETECTOR B
(1 mm CZT + 10 mm CZT)
GATE V6 allows the user to record so-called phase-space files containing the essential properties of the
simulated particles at a given geometric level.
•
The properties of C11, O15, C10 particles in the phantom are stored in a phase-space file.
•
This output file can then be used as an input file in subsequent PET simulations.
B. Simulation setup for X-ray imaging
•
The detector system consists of an X-ray generator and one CZT detector module for 2-D X-ray imaging.
(figure1 (b))
•
Date acquisition was performed using an 1 mm-thick CZT detector and a PMMA phantom same as in section
A.
•
X-ray system has a source to detector distance of 68.5 cm, maximum fan angle of the x-ray beam of 17°, and
magnification factor of 1.17.
•
The spectral X-ray system using a photon counting CZT detector consists of 0.5 x 0.5 mm2 pixel and 400 x
400 array of pixels.
•
X-ray spectrum of 120 kVp tube voltage was generated in SRS-78 program.
•
The x-ray beam quality was 7 mm Al equivalent half value layer(HVL).
DETECTOR C
(1 mm CZT + 15 mm CZT)
Figure 4. in-beam PET image using CZT detector
DISCUSSION & CONCLUSION
• The proton range verification with x-ray imaging using single CZT detector, as described in this paper could be a
simple tool for accurate in vivo range verifications in proton therapy.
• According to our simulations, it is achievable to use two CZT detectors system, which have different thickness
for positron emitting radionuclides.
• Although 1 mm thickness CZT system was enough for x-ray imaging to see anatomical anatomy of phantom,
additional 15 mm thickness CZT detector was required to measure proton distribution.
• Based on this work, we will proceed with investigation of this method for our clinical practice and the next step
will be finding a suitable detector and design.
• This study also presented the potential feasibility about 3D volume measurement for proton dose distribution
using PET/CBCT system with single detector system.