QC in a Digital World - Diagnostic Accreditation Program
Download
Report
Transcript QC in a Digital World - Diagnostic Accreditation Program
QC in a Digital World
John Aldrich PhD FCCPM
Department of Radiology
Vancouver Coastal Health
University of British Columbia
Digital Imaging
Any sufficiently advanced
technology is indistinguishable
from magic… Arthur C Clarke 1961
Overview
New paradigms
Standards
Image acquisition systems
Radiography DR, CR
Fluoroscopy, Angiography DF
CT
US
PACS
New Paradigm
In electronic imaging the
functional parts of
conventional radiology
have been separated:
Image Capture
Image Storage
Image Display
Imaging QC Principles
Proactive QC rather than Reactive QC
Test tool/phantom
Standard imaging
parameters/conditions
Scheduled testing (Daily/Weekly)
Defined and objective
acceptance/rejection
criteria
Patient replaces the
phantom
Non-standard imaging
parameters/conditions
Frequent testing (every patient)
Ill-defined and subjective
acceptance/rejection
criteria
Quality Control (QC)
Acceptance testing
First constancy testing
New equipment
Conformance to
manufacture’s specs/criteria
Baseline value determination
Clinical use period
Routine performance
evaluations
Data evaluation
Specific tests performed at
regular intervals
Consistency checks
Next constancy testing
Evaluate malfunctioning or
out-of-spec equipment
Within the
PASS
established criteria
FAIL
Remedy
Digital System QC
Film
Developed
And
Fixed
Detector
Reading
Digital
Processing
Stored
PACS
Viewed
Display
QC of the digital systems is an additional requirement
– in addition to the usual x-ray performance tests
Health Canada - Quality Control
Safety Code 20A (19812000)
Recommended safety
procedures for the
installation, use and
control of x-ray
equipment. Mainly
concerned with the x-ray
output parameters of the
equipment
Only film processor QC
defined
Safety Code XX (due
2008)
Recommended safety
procedures for the
installation, use and
control of x-ray
equipment. Mainly
concerned with the x-ray
output parameters of the
equipment
25% of the Code is
concerned with QC of the
digital imaging detector
systems
Digital X-ray Systems
Direct Radiography DR
Formation of image without a secondary read-out device
Computed Radiography CR
Use of storage phosphor plate usually in a cassette-based
system
Digital Fluoroscopy/Angiography DF
Image intensifier/video system replaced by digital plate.
Computed Tomography CT
Ultrasound US
DR, CR and DF – Extra QC
Routine QC interval will depend on system
– not less than annually
Dose Calibration
Spatial Resolution
Low Contrast
Uniformity
Artifacts
Spatial Linearity
Dose Calibration
Each system should be calibrated
according to the manufacturers protocol,
as they are all slightly different
General set-up
Arrange for defined dose at surface of cassette at
80 kVp
Expose and read image
Record Exposure Index
The image can also be used to check for
uniformity, linearity and artifacts
Image Quality
All CR and some DR/DF manufacturers have
special Image Quality phantoms and
automatic software to analyze image quality
Resolution and Contrast
Any high contrast resolution
phantom can be used to provide
comparative information
Low contrast resolution is one of
the most difficult parameters to
measure
There are several phantoms and
measurement is subjective, so
consistent technique is essential
Digital Radiography QC
Many DR systems require more frequent
calibration of the uniformity eg every month
Flat field measurement (uniform copper plate)
Uniformity correction
Noise
Artifacts
Contrast-detail and resolution phantom
Special Requirements for CR QC
In film screen systems the film is changed for every
image
With CR the IP is read up to 10,000 times
Almost all plates suffer from wear artifacts
If you are suspicious about an artifact take an image
using the same plate and no patient
Make sure there is a QC program to detect wear
before you detect it clinically
Hammerstrom et al
J Digital Imaging 2006 19:226
Observations
Observations
Observations
Sharp particulates
embedded in the felt
lining under a plastic
clip etched phosphor
surface to create density
on radiograph
Not enough pressure
beside plastic clip to
cause 2nd wear mark to
effect radiograph
Observations
Yellowing of
phosphor
Virox
Observations
Dust
Scratches
CR QC Recommendations
Quality Control (QC) - perform monthly
Inspection – cassette and IP
Visual
Radiographic
CR Cassette cleaning
CR IP cleaning
Benefits
Fewer image artifacts and repeated exposures
Increased life cycle of cassettes, IPs, and readers
Compliance with vendor warranties
Consistency Checks
Weekly/daily
Simple phantom to test reproducibility
To use if there seems to be a problem
Vancouver Phantom
This phantom we have
developed for routine
constancy QC of
digital systems
Field collimation
Standard operating
conditions
Resolution
Contrast
Low contrast
circles
High contrast
mesh
Orthopaedic Measurements
QC in CT - Daily
In-air calibration of scanner every 24 hours
Adjusts sensitivity of all detectors
Important to do this – build into schedule.
QC Frequency
Function
Mechanical
Test
Scan plane congruence
Couch movement
Collimated beam width
Weekly
Annually
X
X
X
Image Quality
Noise and uniformity
CT number
Spatial resolution
Imaged slice thickness
Image linearity
X
X
X
X
X
X
X
Dose
CTDI100 in air
X
QC Phantoms
ACR CT Accreditation Phantom
(RMI)
Alignment, noise, uniformity, CT
number, resolution, MTF, low contrast,
image slice width
Scanner QC phantoms
GE: noise, uniformity, resolution, MTF,
low contrast
Siemens: noise, uniformity, MTF
Ultrasound QC - Phantom
US Probe Test Report
Cracked/Dead
Elements
Ultrasound QC – Clinical
6 dead elements – right image
Slight shadowing in the middle of the image
Discernable loss of signal amplitude
Optimization of Displays
Clean the surface of the display
With the display OFF look at
reflections on the surface of the
display such as lamps,
windows, white coats and name
tags. Reduce these artifacts as
much as possible
Display the SMPTE test pattern
Ensure you can see the 5% and
95% grey scales
Radiology Workstation Contrast
350.00
Darkened Room
Lighting
300.00
Bright Room Lighting
250.00
Contrast Ratio
AAPM
200.00
150.00
100.00
50.00
0.00
1
2
3
4
5
6
7
8
9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34
Unit
Aldrich JE et al. J Digital Imaging 2005;18:287-295
Calibration of Displays
Software generates grayscale levels
Photometer measures the luminance output
at each level and adjusts video card output
to obtain a perceptually linear gradation
between grayscale levels
Calibrates display to DICOM standard
181
Primary PACS Displays
Primary reporting
workstations should be
used in custom-built
reporting areas with low
reflecting surfaces,
ergonomically-designed
chairs, recessed pot lighting
with dimmer controls and
climate control.
Our primary reporting
stations are calibrated for
luminance and contrast
ratio every three months.
Secondary PACS Displays
In contrast, the secondary displays
are used under a range of conditions,
often with the possibility of
distracting reflections and high
ambient lighting.
The secondary displays are checked
normally only on installation
Calibration factors can often be
changed by the user.
Location:
Operating Rooms
Emergency Rooms
3D Processing workstations
(Offices, wards, home)
The Imaging Chain
Detector
Reading
Digital
Processing
Stored
PACS
Viewed
Display
Image are used to follow disease processes so
it important that the whole digital chain is
linear
Linearity should be checked after changes to
software/hardware in any component
The Future
The only perfect science
is hindsight