IMAGE QUALITY
Download
Report
Transcript IMAGE QUALITY
IMAGE QUALITY
NOISE
LINEARITY
CROSS-FIELD UNIFORMITY
IMAGE ARTIFACTS
NOISE IN CT
FLUCTUATION OF
CT# BETWEEN
POINTS IN THE
IMAGE FOR A SCAN
OF UNIFORM
MATERIAL SUCH
WATER.
NOISE IN CT
NOISE LEVEL
PERCENTAGE OF CONTRAST OR IN CT
NUMBERS
NOISE LEVEL %= STANDARD DEVIATION /CT # RANGE
NOISE CAN BE MEASURED BY
SCANNING
A WATER PHANTOM
AND COMPUTING THE MEAN
AND STANDARD DEVIATION
FOR FOR A REGION OF
INTEREST. (ROI)
THE NOISE LEVELS FOR CT
SCANNERS VARY AND DEPEND
ON:
MAS
SCAN TIME
KVP
SLICE THICKNESS
OBJECT SIZE
ALGORITHM
MAS
KVP
SCAN TIME
SLICE T.
NOISE
OBJECT SIZE
NOISE
ALGORITHM vs NOISE
DETAIL
EDGE-ENHANCEMENT
STANDARD
SMOOTHING
NOISE
NOISE IN CT IS MAINLY RELATED
TO
PHOTON FLUX ( # OF DETECTED
PHOTONS)
MATRIX SIZE (PIXEL SIZE)
SLICE THICKNESS
ALGORITHM
ELECTRONIC NOISE ( DETECTOR
ELECTRONICS)
SCATTER RADIATION
NOISE
WIDTH OF PIXEL
SLICE THICKNESS
DOSE
NOISE
SPATIAL R.
NOISE
SPATIAL R.
NOISE
LINEARITY
RELATIONSHIP OF CT# TO THE LINEAR
ATTENUATION COEFFICIENT
CROSS-FIELD UNIFORMITY
THE UNIFORMITY OF CT # THROUGHOUT
THE ENTIRE SCAN FIELD OF VIEW IS
ONE INDICATION THAT CT SCANNER IS
PERFORMANCE IS ACCEPTABLE.
IT REFERS TO THE VALUES OF THE
PIXELS IN THE RECONSTRUCTED IMAGE
THEY SHOULD BE CONSTANT AT ANY POINT IN THE IMAGE OF PHANTOM
CROSS-FIELD UNIFORMITY
CROSS-FIELD UNIFORMITY
IMAGE ARTIFACTS
A DISTORTION OR ERROR IN AN IMAGE
THAT IS UNRELATED TO THE SUBJECT
BEING STUDIED
IMAGE ARTIFACTS
STREAKS
RING
CT# DISTORTION
SOURCES OF ARTIFACTS
PATIENT
IMAGING PROCESS
EQUIPMENT
TYPES AND CAUSES OF ARTIFACTS
STREAKS
IMPROPER
SAMPLING OF DATA
PARTIAL VOLUME
AVERAGING
PATIENT MOTION
METAL & BEAM
HARDENING
NOISE
SPIRAL SCANNING
MECHANICAL
FAILURE
TYPES AND CAUSES OF ARTIFACTS
SHADING
PART. VOLUME AVG.
BEAM HARDENING
SPIRAL SCANNING
SCATTER
OFF-FOCUS
RADIATION
INCOMPLETE
PROJECTION
TYPES AND CAUSES OF ARTIFACTS
RINGS AND BANDS
BAD DETECTOR IN III
GENERATION CT
MOTION ARTIFACT- VOLUNTARY &
INVOLUNTARY
CORRECTION:
IMMOBILIZATION
PROPER EXPLANATION OF
PROCEDURE
SHORT SCAN TIME
MOTION REDUCTION
SOFTWARE (SHIMADZU –MAC)
I
MOTION
METAL ARTIFACTS
INCOMPLETE PROJECTION PROFILES
STAR SHAPED ARTIFACTS
METAL ARTIFACTS
METAL ARTIFACT
MAR
BEAM HARDENING ARTIFACTCUPPING ARTIFACT
INCREASE IN THE MEAN ENERGY OF
THE X-RAY BEAM AS IT PASSES
THROUGH THE PATIENT
IT CAN OCCUR WHEN RADIATION BEAMS
HAVE DIFFERENT PATH LENGTHS
BEAM HARDENING
BEAM HARDENING ARTIFACT
STREAKS OR BROAD DARK BANDS
REDUCTION OF BEAM
HARDENING
BOW TIE FILTER UTILIZATION
SPECIAL CORRECTION SOFTWARE
CORRECTED
PARTIAL VOLUME
ARTIFACT
AVERAGE CT NUMBERS IN SLICE
CORRECTING PARTIAL VOLUME
ARTIFACT
THIN SLICES
NOISE – PHOTON STARVATION
OUT OF FIELD ARTIFACT
EQUIPMENT INDUCED ARTIFACTS
STREAK ARTIFACTS
MECHANICAL FAILURE
POOR GANTRY RIGIDITY
MECHANICAL ASSIGNMENT
TUBE ROTOR WOBBLE
POOR SAMPLING OF THE DETECTOR
BAD DETECTOR-RING ARTIFACTS ( III GEN.)
TUBE ARCING
RING ARTIFACTS- III GENERATION
CORRECTION OF RING ARTIFACT
DETECTOR RECALIBRATION
BALANCING ALGORITHM UTILIZATION
ALIASING ARTIFACT
NOT ENOUGH SAMPLES
CORRECTION OF ALIASING
ARTIFACT
INCREASE NUMBER OF VIEWS OR RAY
SAMPLES PER VIEW.
CONVOLUTION FILTER (SOFTWARE
METHOD)
ALIASING ARTIFACT
TUBE ARCING
SPIRAL ARTIFACTS
WINDMILL
MOTION
STAIRCASE
SHADING
WINDMILL
MOTION
MOTION
STAIRCASE