IAEA Training Material on Radiation Protection in Radiotherapy
Download
Report
Transcript IAEA Training Material on Radiation Protection in Radiotherapy
IAEA Training Material on Radiation Protection in Radiotherapy
Radiation Protection in
Radiotherapy
Part 10
Good Practice including Radiation
Protection in EBT
Lecture 4: Treatment verification and reporting
The Problems:
The correct dose of radiation shall be
delivered just to the target.
The dose to surrounding structures shall be
as low as possible.
Must be achieved on many occasions
(typically >30 treatment fractions)
It must be verifiable
Must be documented in a way that allows
others to understand all important factors of
the treatment performed
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
2
Objectives
To understand the different frames of reference used
in radiotherapy
To be familiar with techniques which allow to verify
that the treatment is delivered to the appropriate
location
To appreciate the need for reporting dose AND
volume in prescription and treatment reporting in
radiotherapy
To be aware of the reports of the ICRU regarding
reporting of radiotherapy treatments
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
3
Contents of lecture 4 in part
10
1. Sources of uncertainty
2. Methods to verify dose delivery
portal films
in vivo dosimetry
3. Prescription and reporting
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
4
1. Sources of uncertainty
Patient localization
Organ motion
Imaging (resolution, distortions,…)
Definition of anatomy (outlines,…)
Beam geometry
Dose calculation
Dose display and plan evaluation
Plan implementation
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
5
Patient localization
The patient should be
positioned identically
during diagnostics (CT),
in simulation and 30+
times during treatment
Sources of uncertainty:
motion
reliability of marks on
the skin
couch sag
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
6
Organ motion
Affects most organs - particularly noticeable
for lung cancer, liver, prostate and other pelvic
malignancies
Shown here is the difference in CT scan
between inhale and exhale position
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
7
Imaging issues
Partial volume effects
Distortion (MR)
Limited spatial resolution (PET)
More discussion on these issues in the
companion course on diagnostic
radiology
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
8
Target definition, outlining of
organs
Radiation Protection in Radiotherapy
Decide where the
organ is and what
extend it has.
Are the seminal
vesicles really
where they are
drawn here?
Part 10, lecture 4: Treatment verification
9
Beam geometry and positioning
Is the block exactly
where it should be?
Is there gantry sag?
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
10
Dose calculation
Radiation Protection in Radiotherapy
There are many
different dose
calculation algorithms
All have limitations (and
be it the long time
required to calculate the
dose)
Must know what to trust
Part 10, lecture 4: Treatment verification
11
Plan display and evaluation
Comparison of competing plans...
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
12
Implementation of the plan
Transfer of data between
units
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
13
2. Verification of dose delivery
The plan looks great...
However, one must ensure that during
treatment everything is matching the
treatment plan
ADAC
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
14
In practice there are many
systems...
Patient
Diagnostic tools
Treatment
planning
Radiation Protection in Radiotherapy
Treatment unit
Part 10, lecture 4: Treatment verification
15
Ensure different co-ordinate
systems match...
Patient
Diagnostic tools
Treatment
planning
Radiation Protection in Radiotherapy
Treatment unit
Part 10, lecture 4: Treatment verification
16
Good practice
Verify data transfer
Use critical approach
Verify treatment by comparing an image
taking during treatment with a reference
image taken either during simulation or
taken from the treatment planning
system
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
17
H u m an errors in d ata tran sfer d u rin g th e p rep aration
an d d elivery of rad iation treatm en t affectin g th e fin al
resu lt: " garb age in , garb age ou t"
L eu n en s, G ; V erstra ete, J ; V a n d en B og a ert, W ; V a n D a m , J ; D u treix, A ; va n d er S ch u eren , E
D ep artm en t o f R ad io th erapy , U n iv ersity H o sp ital, S t. R afaël, L euv en , B elg iu m
A b stract
D u e to th e larg e n u m b er o f step s an d th e n u m b er o f p erso n s in v o lv ed in th e p rep aratio n o f a rad iatio n
treatm en t, th e tran sfer o f in fo rm atio n fro m o n e step to th e n ex t is a v ery critical po in t. E rro rs du e to
in ad eq u ate tran sfer o f in form atio n w ill b e reflected in ev ery n ex t step and can seriou sly affect th e fin al
resu lt o f th e treatm en t. W e stu d ied th e freq u en cy and th e so u rces o f th e tran sfer erro rs. A to tal n u m b er o f
4 6 4 n ew treatm en ts h as b een check ed o v er a p erio d of 9 m o n th s (Janu ary to O cto b er 1 9 90 ). E rron eo u s d ata
tran sfer h as b een d etected in 1 3 9 /2 4 ,12 8 (less th an 1 % ) o f th e tran sferred p aram eters; th ey affected 26 %
(1 1 9 /46 4 ) o f th e ch eck ed treatm en ts. T w en ty-fiv e o f th ese d ev iation s co u ld h av e led to larg e g eog raph ical
m iss o r im p o rtan t o v er- o r u n d erd o sag e (m u ch m o re th an 5 % ) o f th e o rg an s in th e irrad iated vo lu m e, th u s
in creasin g th e com p licatio n s o r d ecreasin g th e tu m ou r con tro l pro b ab ility , if no t co rrected . S u ch m ajo r
d ev iatio n s, o n ly o ccurrin g in 0 .1 % o f th e tran sferred p aram eters, affected 5 % (25 /4 6 4) o f th e n ew
treatm en ts. T h e so u rces of th ese larg e d ev iatio n s w ere n early alw ay s hu m an m istak es, w h ereas a
co n sid erab le nu m b er o f th e sm aller d ev iatio n s w ere, in fact, co n scio u sly tak en d ecisio n s to d ev iate from th e
in ten d ed treatm en t. N early h alf o f th e m ajor d ev iation s w ere in trod u ced d u ring in pu t o f th e d ata in th e
ch eck -and -con firm sy stem , d em o n stratin g th at a sy stem aim ed to p rev en t accid en tal erro rs, can lead to a
co n sid erab le nu m b er o f sy stem atic erro rs if u sed as an u n con tro lled set-u p sy stem . T h e resu lts o f th is stu d y
sh o w th at h u m an m istak es can serio u sly affect th e ou tco m e o f p atien t treatm en ts.(A B S T R A C T
T R U N C A T E D A T 2 50 W O R D S ) [Jou rn al A rticle; In E n g lish ; N eth erland s]
Green Journal 1992: > 50 occasions of data transfer
from one point to another for each patient!
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
18
Most important comparison
Reference from
planning
• Simulator film
• DRR
Radiation Protection in Radiotherapy
Check film during
treatment
• Port film
• EPID
Part 10, lecture 4: Treatment verification
19
Portal films
Taken during (or
directly before)
treatment with:
beam from the
treatment unit
patient in treatment
position
shielding in place
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
20
Port films
Usually taken before
or after treatment
If the field itself does
not show enough
anatomy, a double
exposure technique
can be used:
Expose treatment field
Open collimators and expose the same film again
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
21
In radiotherapy practice
The risk from the additional exposure in
dual exposure technique portal images
is negligible compared to the risk not
treating the correct area in the patient...
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
22
Electronic Portal Images
A film less way to
verify field location
Mounted on the
linac
Different systems:
ion chamber
fluoroscopic screen
semiconductor
arrays
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
23
Electronic Portal Imaging
Devices - EPIDs
Offered by all major
manufacturers
Has several
advantages:
Radiation Protection in Radiotherapy
Easy use and
positioning of the
system
Allows on line
verification
Multiple images (‘cine’)
can be taken during one
treatment
Images available in
digital format
Part 10, lecture 4: Treatment verification
24
Verification films/images
Two different aims:
Verify correct shielding
Verify patient location
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
25
Verify block shape and position
should be done for every treatment field
at least once per course of radiotherapy
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
26
Verify patient location
Should be done using two images which give
adequate information on the location of the
target in respect to the treatment beam
geometry.
These do not necessarily need to be
treatment fields (e.g. in case of IMRT or arcs)
This verification should be repeated every
week during treatment for radical treatments
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
27
Verify dose delivered to the
patient
May be done in
customized
anthropomorphic
phantoms
More likely using in
vivo dosimetry
Radiation Protection in Radiotherapy
Breasts modeled on a particular
patient to verify skin dose using
TLDs
Part 10, lecture 4: Treatment verification
28
IN VIVO DOSIMETRY
ICRU report 24 (1976):
“An ultimate check of the actual
treatment given can only be made by
using in vivo dosimetry.”
Why do in vivo dosimetry
Quality Assurance – Treatment Verification
Measure because we don’t know
Limitations of dose planning
Patient movement
Verify dose for the record
Critical organs
Legal aspects
Clinical trials
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
30
Challenges for in vivo dosimetry
Typically low dose where the detectors
can be located
Variation of detector response with
Temperature
Radiation Quality
Direction of radiation
Need to not interfere with therapeutic
objective
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
31
Potential detectors
Needed: High sensitivity, tissue
equivalence, high spatial resolution
TLD
Semiconductors (diodes, MOSFETs)
Radiochromic film
Others (alanine, gel dosimetry, …)
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
32
In vivo exit dosimetry
Radiation Protection in Radiotherapy
calculate exit
fluence
determine what the
portfilm/EPID should
look like
verify dose by
projecting back
Part 10, lecture 4: Treatment verification
33
3. Prescription and reporting
Prescription may be at the discretion of
individual clinicians, depending on equipment
available, experience and training
Reporting must be uniform - any adequately
educated person must be able to understand
what happened to the patient in case of:
need for a different clinician to continue treatment
re-treatment of the patient
clinical trials
potential litigation
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
34
Recommendations by the ICRU
International
Commission on
Radiation Units and
Measurements
ICRU reports provide
guidance on
prescribing, recording
and reporting
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
35
The required dose information
Should describe the treatment
concisely and unambiguously
Gets more and more complex as
treatment approaches become more
complex
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
36
The required dose information
1950: Patient X: x
Gy in n fractions to y
tumour
1993: ICRU 50,
appendix I
1999: ICRU 62 …plus
the same for OAR...
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
37
Important
Information includes dose AND volume
Information is provided for target AND
normal structure
Information includes dose AND
fractionation
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
38
Dose Volume Histograms
Dose and
Volume
information
Example
from ICRU
report 62:
prostate
cancer
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
39
Normal tissue information
rectum
Radiation Protection in Radiotherapy
bladder
Part 10, lecture 4: Treatment verification
femoral heads
40
Normal tissue information
rectum
bladder
femoral heads
Quick Question: Why is there a ‘kink’ in the DVH for
the femoral heads at about half of the dose to the
prostate?
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
41
Answer
1
3
60 Gy
2
4
In a ‘four field box’ treatment two of the
four fields go through the femoral
heads...
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
42
Record keeping required...
BSS.appendix II.31 “… in radiation
therapy, a description of the planning
target volume, the dose to the centre of
the planning target volume and the
maximum and minimum doses
delivered to the planning target volume,
the doses to other relevant organs, the
dose fractionation, and the overall
treatment time; …”
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
43
A final point
The best localization is
void if it is the wrong
patient!!!
Always check patient
identification prior to
treatment
In practice this is a surprisingly
common problem - there are
many patients called ‘Mr Smith’
and patients may be confused
when presenting for treatment.
A photo in the treatment sheet
may also be useful
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
44
Summary
Optimization of radiotherapy includes the
optimization of radiation beam design
Treatment verification using portal imaging is
essential
In vivo dosimetry is a useful complementation of
portal imaging
A large amount of information - including dose AND
volume - is required to describe a radiotherapy
treatment concisely
Radiotherapy treatments should be reported
following international conventions
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
45
Where to Get More Information
International Commission on Radiation Units and
Measurements. ICRU report 50: Prescribing,
recording, and reporting photon beam therapy.
Bethesda. 1993.
International Commission on Radiation Units and
Measurements. ICRU report 62: Prescribing,
recording, and reporting photon beam therapy
(Supplement to ICRU report 50). Bethesda 2000.
Radiation Protection in Radiotherapy
Part 10, lecture 4: Treatment verification
46
Any questions?
Question:
Please discuss the advantages (and
disadvantages) of electronic portal imaging as
compared to portal films for verification of patient
positioning during radiotherapy.
The answer should include:
Advantages
No additional dose
required for most images
Easy positioning of the
system
Allows on line verification
Multiple images (‘cine’)
can be taken during one
treatment
Images available in
digital format
Radiation Protection in Radiotherapy
Disadvantages
High
initial investment
costs
Dual exposures may be
more difficult
Image quality of many
systems inferior to film
Part 10, lecture 4: Treatment verification
49