Multi-slice CT
Download
Report
Transcript Multi-slice CT
CT
Multi-Slice CT
Third Generation CT
Single or Multislice
Z-axis orientation
perpendicular to page
Patient
Single Slice Thickness Determined by Collimation
Z
Detector
Single-Slice Detectors
Many detectors rotate around
patient
Single row in z-direction
Slice thickness determined by
collimation
Z-Axis
Single Slice CT: Slice Thickness
Collimated Beam Thickness
Collimated Beam Thickness
Thin
slice
Thick
slice
Z-Axis
Z-Axis
Multi-slice CT
Tube
Post-collimator
Detectors
What’s Different for Multislice CT?
Multislice CT
Multiple rows of
detectors
Open collimators
in “Z” direction
34
2
1
http://www.veterinaryimaging.com/images/MSS_CT.gif
Multi-slice CT
Developed in late
1990’s
Detector array
segmented in zdirection
Simultaneous
acquisition of
multiple slices
http://www.ctisus.com/gallery/images/m
ultidetector/multislice_ct.jpg
Single Slice vs. Multislice Detector
Collimated Beam Thickness
Z-Axis
Single slice
detector
Multislice
detector
Multi-Slice Detectors
Many detectors going around patient
Many detector rows in z-direction
Slice thickness determined by
Collimation
electronic detector selection
“Z” Direction
Single
Multi
Multi-slice CT
Size & distribution of detectors in non-axial
direction similar to previous CT’s
Similar spatial & contrast resolution
Distribution of detectors in axial direction
varies with manufacturer
All detectors
same width
“Z”
Direction
Variable widths
Multi-slice CT
Uniform Detector Thickness
Multiple detectors in axial
“z” direction
direction
Size must accommodate
thinnest slice
1 2 3 4
Detector signals can be used
Individually
In groups
Four thin slices
1
2
3
Four thicker slices
4
Detectors vs. Channels
# Physical Detectors not necessarily equal to # of
possible Slices
Maximum # slices limited by Digital Acquisition
System (DAS) channels
Electronic counters
Imaging speed bottleneck
How fast data can be received from
detector arrays
Detectors vs. Channels Example
16 detectors
4 channels
Multi-Slice Detector Example
16 Detector Rows, 4 Channels
Detectors vs. Channels
4 X 1.25 mm
Beam collimated to 4
detector rows
1 detector row per DAS
channel
Effective
Detector
Detectors vs. Channels
4 X 2.5 mm
Beam collimated to 8
detector rows
2 detector rows per DAS
channel
Effective
Detector
Detectors vs. Channels
4 X 3.75 mm
Beam collimated to 12
detector rows
3 detector rows per DAS
channel
Effective
Detector
Detectors vs. Channels
4 X 5 mm
Beam collimated to 16
detector rows
4 detector rows per DAS
channel
Effective
Detector
Capture Efficiency
Fraction of detector area
that is active detector
Equal-width Detectors
Disadvantage
Many gaps
Gaps are dead space
Reduce capture
efficiency
Multi-slice CT
“Adaptive Array Detectors”
“z” direction
Some scanners use
detectors of various
widths
Post-collimators
used to partially
block wider
elements for
thinner slices
1 2 3
Three thicker slices
Postcollimators
1 2 3 4
Four thinner slices
Variable Width Detectors
Center detectors
thinner
Thicker
detectors can
function as
thinner ones
using
collimation
Thinner
detectors can
function a
thicker one by
combining
signals
Single Slice Pitch Definition
table motion during one rotation
Slice Pitch = --------------------------------------slice thickness
Beam Pitch
Defined only for Multi-slice scanners
table motion during one rotation
Beam Pitch = --------------------------------------Beam thickness
Beam thickness
Beam Pitch
Defined only for Multi-slice scanners
Beam Pitch > 1
Beam Pitch = 1
CT Beam Pitch
Example
5
table motion during one rotation
Beam Pitch = --------------------------------------Beam thickness
mm slices
4 simultaneous
slices
Beam pitch = 1
1 revolution / sec.
Beam thickness?
Table speed?
Beam Thickness
5
table motion during one rotation
Beam Pitch = --------------------------------------Beam thickness
mm slices
4 simultaneous
slices
Beam pitch = 1
1 revolution / sec.
Beam Thickness = 5 X 4 = 20 mm
Table Speed
5
table motion during one rotation
Beam Pitch = --------------------------------------Beam thickness
mm slices
4 simultaneous
slices
Beam pitch = 1
1 revolution / sec.
20 mm beam
thickness
Table speed = 20 mm rotation (1 sec) = 20 mm / sec
Slice Thickness Defined at
Rotational Center
Rotational
Center
Tube
Detector Field must be Larger than Slice
Thickness at Rotational Center
Rotational
Center
Beam Divergence More of a Problem for
Multi-Slice
Rays
diverge
No longer
essentially parallel
Leads to Cone Angle
Artifact
Significant for 16, 32,
64 … data channels
Requires use of
special reconstruction
algorithms to
compensate
Multislice CT Doses
Can be 10-30% higher than for single slice units (ICRP
#47)
Cause
Divergent beam
Other considerations
Tendency to cover more volume (anatomy)
Better availability of equipment
Other Reasons for High CT Doses
Repeat Exams
No adjustment of technique factors for different size
patients
No adjustment for different areas of body
Multislice CT Advantage?
Speed!
Single slice / Multislice Images about the
same!
Speed = Power
Speed enables new applications
How do we spend our new speed?
Multi-slice CT Imaging
Clinical Advantages
1.
2.
3.
Thinner slices for improved z-direction resolution
Same acquisition in shorter time
Scan larger volumes in same time
Multi-slice CT Imaging
Clinical Advantages
Thinner slices
Improvement in CTA of neck, aorta, renal vessels
Better reconstructions
Sagittal, coronal, oblique
3-D
Fundamental Trade-off
“z” axis resolution vs. image noise
Multi-slice CT Imaging
Clinical Advantages
Improved x-ray tube utilization
Reduced x-ray tube loading
4 slices acquired with same tube loading previously used
for 1
Less need to pause of tube cooling
Reduced wear & tear
Other anticipated benefits
CT endoscopy
Diagnosis of pulmonary embolism
Multi-slice CT Imaging
Clinical Advantage: Angiography
Simplifies contrast bolus timing
Continuous observation of target vessel
Can reduce amount of contrast required
Coverage from aorta to lower extremities
Runoff
Continuous CT Imaging
Interventional Procedures
Biopsy & drainage
Neuro
Chest
Abdominal
Spine
Catheter and tube placement
Helps operator avoid critical structures near path of
biopsy needle
Better visualizing of moving structures
Respiration
Functional CT
Brain perfusion
Multi-Slice Compared to
Single-slice helical
Much Faster
No significant image quality differences
Equivalent Patient Dose
Ref:
Willi Kalender, Ph.D
Institute of Medical Physics
University of Erlanger, Germany
Organ Coverage Time Depends On
1.
2.
3.
Beam Pitch
Rotation Time
Detector
Acquisition Length
64 Slice CT Typical Coverage Times
Heart & coronary arteries / brain / lungs
5 seconds
Whole body coverage for blood clot search
30 seconds
Philips
64-Slice Commercial Cardiac CT
IGE Philips
Siemens
(1 tube)
Siemens
(2 tube)
Toshiba
Min. Rotation Time (s)
.35
.53
.33
.33
.4
Detector length (mm)
40
40
19.2
19.2
32
Time to cover heart
(120 mm) (s)
5.3
6.3
10.3
5.1
7.5
www.impactscan.org
What’s Next?
Slice Wars
Philips
256 Slicer
Toshiba
320 Slicer
Implications of 256+ Slices
Full organ coverage in single rotation
12-16 cm coverage
Improved temporal resolution
Reduced artifacts
Whole-organ function (perfusion) studies
Functional data perfectly registered to anatomic data
The Future
More slices
Flat panel area detectors
???
Multi-slice challenges: More Slices
Computer issues
More archival capacity
Requires faster computer systems
Requires faster communications for remote viewing
Radiologist responsible for all images
Acknowledgement
Many drawings obtained from www.impactscan.org
website