Medical Laboratory Instrumentation 2010

Download Report

Transcript Medical Laboratory Instrumentation 2010

Spectrophotometer –Continue
Dr Fadhl Alakwaa
2013-2014
Third Year
Biomedical Engineering Department
www.Fadhl-alakwa.weebly.com
AGENDA
•
•
•
•
•
Atomic Absorption Spectrometry
Photometry: Atomic Emission Spectroscopy
Fluorometer
Configuration of the spectroscopy systems
spectrophotometer components:
1- Radiation source
Continuum sources
Line source
AGENDA- Continue
2-Wavelength Selection Devices(Resolution Power)
Filters
Colored glass
Interference filter
Monochromator
Prisms.
Diffraction Gratings.
3- Detectors
Photovoltaic cell
Phototube
Photomultiplier tube (PMT)
Atomic Absorption Spectrometry
• AAS is an elemental analysis technique capable of
providing quantitative information on 70 elements in
almost any type of sample.
• AAS are that no information is obtained on the
chemical form of the analyte (no “speciation”) and that
often only one element can be determined at a time.
• This last disadvantage makes AAS of very limited use
for qualitative analysis.
• AAS is used almost exclusively for quantitative analysis
of elements, hence the use of the term “spectrometry”
in the name of the technique instead of
“spectroscopy”.
Atomic Emission Spectroscopy
• Atomic emission spectroscopy has relied in the past on
flames and electrical discharges as excitation sources, but
these sources have been overtaken by plasma sources,
such as the inductively coupled plasma (ICP) source.
• Atomic emission spectroscopy is a multielement technique
with the ability to determine metals, metalloids, and some
nonmetal elements simultaneously.
• The major difference between the various types of atomic
emission spectroscopy techniques lies in the source of
excitation and the amount of energy imparted to the atoms
or ions (i.e., the excitation efficiency of the source).
Photometry: Flame atomic emission
spectroscopy
• Flame atomic emission spectrometry is
particularly useful for the determination of
the elements in the first two groups of the
periodic table, including sodium, potassium,
lithium, calcium, magnesium, strontium, and
barium.
• The determination of these elements is often
called for in medicine, agriculture, and animal
science.
Photometry Application
• Flame photometry is used for the quantitative
determination of alkaline metals and alkalineearth metals in blood, serum, and urine in
clinical laboratories. It provides much simpler
spectra than those found in other types of
atomic emission spectrometry, but its
sensitivity is much reduced.
• sodium, potassium, magnesium and calcium
in blood
MOLECULAR EMISSION
SPECTROMETRY
Fluorometer
Fluorometer Application
• Fluorometry is used in the analysis of clinical
samples, pharmaceuticals, natural products,
and environmental samples.
• There are fluorescence methods for steroids,
lipids, proteins, amino acids, enzymes, drugs,
inorganic electrolytes, chlorophylls, natural
and synthetic pigments, vitamins, and many
other types of analytes.
Spectroscopy components
•
•
•
•
•
•
Many optical instruments share similar design
(1) stable radiation source
(2) transparent sample holder
(3) wavelength selector (dispersive)
(4) radiation detector
(5) signal processor and readout
Light Sources
UV Spectrophotometer
1.
Hydrogen Gas Lamp
2.
Mercury Lamp
Visible Spectrophotometer
1.
Tungsten Lamp
InfraRed (IR) Spectrophotometer
1.
Carborundum (SIC)
Cells
•UV Spectrophotometer
Quartz (crystalline silica)
• Visible Spectrophotometer
Glass
• IR Spectrophotometer
NaCl
Configuration of the spectroscopy systems
Radiation Source
• An ideal radiation source for spectroscopy should have the
following characteristics:
1. The source must emit radiation over the entire wavelength range to
be studied.
2. The intensity of radiation over the entire wavelength range must be
high enough
so that extensive amplification of the signal from the detector can be
avoided.
3. The intensity of the source should not vary significantly at different
wavelengths.
4. The intensity of the source should not fluctuate over long time
intervals.
5. The intensity of the source should not fluctuate over short time
intervals. Short time fluctuation in source intensity is called “flicker”.
• Most sources will have their intensities change
exponentially with changes in voltage, so in all
cases a reliable, steady power supply to the
radiation source is required.
• Voltage regulators (also called line
conditioners) are available to compensate for
variations in incoming voltage.
Radiation
Source
And
Detectors
Fig 7.3
Continuum sources
• Continuum sources emit radiation over a wide range of
wavelengths and the intensity of emission varies slowly as a
function of wavelength.
• Typical continuum sources include :
• the tungsten filament lamp which produces visible radiation (white
light),
• the deuteriumlamp for theUVregion,
• high pressure mercury or xenon arc lamps for the UV region,
• and heated solid ceramics or heated wires for the IR region of the
spectrum.
• Xenon arc lamps are also used for the visible region.
• Continuum sources are used for most molecular absorption and
fluorescence spectrometric instruments.
Line sources
• Emit only a few discrete wavelengths of light, and the
intensity is a strong function of the wavelength.
• Typical line sources include:
• hollow cathode lamps and
• electrodeless discharge lamps, used in the UV and
visible regions for AAS and atomic fluorescence
spectrometry,
• sodium or mercury vapor lamps (similar to the lamps
now used in street lamps) for lines in the UV and visible
regions, and lasers.
• They are used as sources in Raman spectroscopy,
molecular and atomic fluorescence spectroscopy.
Wavelength Selection Devices
• Filters
– absorption filters Colored glass
– Interference filter
• Monochromator
– The entrance slit
– Prisms.
– Diffraction Gratings.
Wavelength
selector
Fig 7.2
Colored glass
• stable, simple, and cheap,
• blue glass transmits blue wavelengths of the
visible spectrum but absorbs red and yellow
wavelengths.
• The range of wavelengths transmitted is broad
compared with prisms and gratings which are
also devices used to select a narrow wavelength
range from a broad band polychromatic source.
The transmission range may be 50–300 nm for
typical absorption filters.
Interference Filter
• two thin sheets of metal
sandwiched between glass
plates, separated by
transparent material.
• Interference filters can be
constructed for transmission
of light in the IR, visible, and
UV regions of the spectrum.
• The wavelength ranges
transmitted are much smaller
than for absorption filters,
generally 1–10 nm, and the
amount of light transmitted is
generally higher than for
absorption filters.
Interference Filter
• The filter operates on the
principle of constructive
interference to transmit
selected wavelength ranges.
The wavelengths
transmitted are controlled
by the thickness and
refractive index of the
center layer of material.
• Interference for transmitted
wave through 1st layer and
reflected from 2nd layer
Prisms
Prisms are used to disperse IR, visible, and UV radiation. The most common prisms are
constructed of quartz for the UV region, silicate glass for the visible and near-IR region, and
NaCl or KBr for the IR region.
Diffraction Grating (most modern
instruments)
• UV, visible, and IR radiation
can be dispersed by a
diffraction grating.
• A diffraction grating consists
of a series of closely spaced
parallel grooves cut (or ruled)
into a hard glass, metallic, or
ceramic surface.
• A grating for use in the UV and
visible regions will contain
between 500 and 5000
grooves/mm,
• while a grating for the IR
region will have between 50
and 200 grooves/mm.
d, the distance between grooves
n, the order of diffraction.
Resolution Power
• Resolution Required to Separate Two Lines of
Different Wavelength.
• Ex: in order to observe an absorption band at
599.9 nm without interference from an
absorption band at 600.1 nm, we must be able to
resolve, or separate, the two bands.
• The resolving power R of a monochromator is
equal to λ/ dλ, where λ is the average of the
wavelengths of the two lines to be resolved and
dλ is the difference in wavelength between these
lines.
Prism Resolution Power
• refractive index of the prism material
• t is the thickness of the base of the prism
• glass prisms disperse visible light better than
quartz prisms.
Resolution of a Grating.
• where n is the order and N is the total number
of grooves in the grating.
• Ex: Suppose that we can obtain a grating with
500 lines/cm. How long a grating would be
required to separate the sodium D lines at
589.5 and 589.0 nm in first order?
•
•
•
•
R=1179=nN
For first order n=1 N=1179 lines
(1179/500) cm long, or 2.358 cm.
Ex2: how many lines per centimeter must be
cut on a grating 3.00 cm long to resolve the
same sodium D lines?
• nN =N =1179
• 1179 l 3:00 cm= 393 lines/cm
Detectors
• There are a number of different types of photon
detectors, including the photomultiplier tube, the
silicon photodiode, the photovoltaic cell, and a
class of multichannel detectors called charge
transfer devices.
• Charge transfer detectors include photodiode
arrays, charge-coupled devices (CCDs), and
charge-injection devices (CIDs).
• These detectors are used in the UV/VIS and IR
regions for both atomic and molecular
spectroscopy.
Photomultiplier tube
PMT
PAGE 106 EXAMPLE 2
• A photomultiplier with 12 dynodes each of
multiplication factors 4. its output feeds an
operation of amplifier with a feedback resistor of
10 mega ohm find:
• The minimum rate of incident photons
• The minimum rate of quantum energy of UV
radiation of wavelength 250 nm incident on the
photomultiplier cathode to produce a potential
difference of 2 volt across this feedback resistor?
Radiation
Source
And
Detectors
Fig 7.3
INSERT LECTURE IS THE MIDTERM EXAM.