Sterilization Device for Liquid Chromatography Solvents

Download Report

Transcript Sterilization Device for Liquid Chromatography Solvents

Sterilization Device for Liquid
Chromatography Solvents
Design Team
Nick Roulleau, Michael Vose
Michael Racette, Michael McKay
Advisor
Professor Mohammad Taslim
Introduction
• Background
• Problem Statement
• Past Art
• Design Requirements
• Design Concepts
• Prototype Design
• Component Analysis
• Recommendations
What is Liquid Chromatography?
A substance comprised of components A and B is dissolved
in a solvent and enters the analytical column, where it is
separated
Basic Components of an HPLC System
From http://www.waters.com/WatersDivision
Problem
From http://www.waters.com/WatersDivision
Design Goal
To mitigate the risk of blockage at the
inlet frit due to bacterial contamination
and extend the useful life of the UPLC
column.
Existing Solutions
• In-line filters
• Guard columns and cartridges
• Pre-filtration of samples and
mobile phase liquids
Product Requirements
• Mandatory:
» Must be adaptable for use worldwide
» Must extend the useful life of columns
» Must meet safety standards (ISO, UL and CE)
» Must operate for 1 year w/o user intervention
• Desirable:
» Should be able to filter two bottles simultaneously
» Should meet customer acceptance criteria
– Low-maintenance
– Easy to use
– Cost
Constraints
• Cannot change the chemical composition
» Of the solvent
» Of the sample
• Cannot create risk of causing pump cavitation
• Cannot hinder bottle accessibility
• Cannot negatively impact system resolution
Initial Design Concepts
UV Probe
Pump/filter--Cap enclosure
Pump/filter--External enclosure
Preliminary Design – UV Probe
• Inexpensive
• Simple Design
Why Not Use Ultraviolet Radiation as a
Primary Solution?
• Degradation of organic solvent modifiers
(Low Risk)
• Degradation of aqueous additives
(Low Risk)
• User safety from UV-C exposure
(Medium Risk)
• UV can inactivate but not remove
bacteria
Filter Sizing
• How many bacteria could be generated per year?
• Logarithmic growth:
» Assuming worst case
– 100% replicating
– Short generation time
– Neglecting lag phases and cell death
• Filter capacity = 107 CFU/cm2
Filter Sizing
With logarithmic bacterial growth, filter area
becomes exceptionally large in a short period
Current Design
External Filter Enclosure with UV
Dual-head
brushless DC
pump
UV lamp
with multiple
sterilization
lines
Pall AcroPak 200 filters
Filter Selection
• Membrane with material
compatibility
• Sufficient capacity to contain 1
year of inactivated bacteria
Pump Selection
Micro-diaphragm pump
• Dual pump heads
• Ability to run dry
• DC brushless motor for
long life
Pump Pressure Requirements
• Pump must deliver sufficient differential
pressure (Δp) to move fluid through filter
• Darcy’s equation for porous media:
kA( p1  p2 )
Q
 ( L)
Q = flow rate
k = permeability constant for filter
A = effective filter area (EFA)
µ = viscosity
L = membrane thickness
p1 = pump-side pressure
p2 = outlet pressure
UV Block Design-Initial Concept
UV Block Design
• 99.99% inactivation requires a UV dose of at least 40
mJ/cm2 for nearly all species of bacteria
• Dose is a function of the irradiance (mW/cm2) and time of
exposure (in seconds)
Dose = Irradiance x time
Fd 1 2
1
L
L  X  2H
1

tan
 
tan 1
H
H 2  1   H XY
X  (1  H )2  L2
Y  (1  H ) 2  L2
h
H
r
l
L
r
X ( H  1) 1
H 1
1
 tan

Y ( H  1) H
H 1
A
dA
UV Block Design
UV Block Design
13 loops necessary with an 18W UV bulb and thin wall FEP tubing
Test Planning
• Verification Test
»
Does the Device Meet
Design Requirement?
»
Pump Particle-Laden
Water from Bottles With
and Without Device
Device
Pump
Sensor
Column
»
Compare Backpressure
and/or Flow Rate
PRESSURE
80
70
60
50
40
30
20
10
0
0
100
200
300
Test Results
Backpressure Test
12
10
Backpressure (PSI)
8
6
No Device
Device (Norm)
4
2
0
1
599
1197 1795 2393 2991 3589 4187 4785 5383 5981 6579 7177 7775 8373 8971 9569 10167 10765 11363 11961
Elapsed Time (s)
Backpressure was reduced by 28% when our device was used
Cost Analysis
• Developed target costs by estimating:
» Annual costs without the assistance of our device
(excluding operational costs)
» Savings in material costs by implementing our device
• Potential savings for high-end users = $44,000
• Minimum estimated annual savings = $600
• Target production cost = $500
• Target prototyping cost = <$1500
Recommendations for Further
Development
• Improve manufacturability of the design
» Simplify tubing system
» Smaller pump
» Custom filter size
• Analyze effectiveness of UV with
microbiological testing
Summary
• Introduction to liquid chromatography
• The problem and its source
• Requirements of a good solution
• Design considerations
• Prototype design and analysis
• Recommendations
Questions???