Transcript Batteries

Batteries
User Brianiac (2003), Batteries [photograph]. Retrieved from https://commons.wikimedia.org/wiki/File:Batteries.jpg
Clinical Use

Primary:

Single use


Alkaline, mercury, zinc air etc.
Secondary:

Rechargeable




Lead acid/gel cell
Nickel Cadmium (Ni-Cd)
Li-ion
NiMH
Common Problems:

All rechargeable batteries have a life span;



Typically # of charge cycles
The specification varies with technology
Heat & discharge cycles limit lifespan
Caring for Wet Lead Acid:


Minimum cell voltage is 2V
Hydrometer is the best test



1.23Kg/L density
Rechargeable batteries lose capacity when
stored-lead acid are no exception
Trickle charge regularly when stored
Charging Rechargeables:

Rule of thumb;



Non-current limiting charger-set voltage to
nominal +20%
If an ammeter is not available-check
repeatedly for excessive heat
Current limiting charger-set volts +20%, set
current to max charge limit, check for
excessive heat
By Andrevruas (Own work) [CC BY 3.0 (http://creativecommons.org/licenses/by/3.0) or GFDL (http://www.gnu.org/copyleft/fdl.html)], via Wikimedia Commons
Rejuvenating Batteries:


Procedure; Only if cell does not respond to normal
charging! Never leave cells unattended!

Set current limiting supply to nominal +20%V

Set current limit to 20% of Ahr rating

Increase voltage until max current flows

Check cell heat-warm is acceptable, hot is not

Reduce voltage while current flow is
maintained
Note: This is a temporary repair, new batteries will be
req'd.
Substituting Chargers:



Important Specs;

Line voltage

O/P voltage

Max current o/p

Physical dimensions/mounting

Connector
Some wall units are AC o/p

Bridge rectifier can change to DC
Any customized charger should be permanently
attached to the equipment it is used with
Battery Substitution
• One can replace rechargeable cells for one use
batteries if:
- The equipment does NOT have an internal
charger!
• If the charger delivers excessive charge, then the
cells and device may be damaged
Substitution Example
• You make a pack of 12, 600 mAh, 1.2 V, NiCd
cells connected 4x3:
– 3 sets of 4 batteries in series
1.2 V x 4 = 4.8 V
– 3 sets in parallel
600 mAh x 3 = 1800 mAh
• You have created one, 4.8 V, 1800 mAh battery!
Substitution Example
• To charge your pack, use Trickle charge:
1800/100 = 18 mA for a few days Max charge:
for NiCd=C=1800 mA (1.8 A)!
Minimum Testing:






Check battery volts;

Nominal typically needs no adjustment
Corrosion; clean if present
Capacity testing; use equipment it supplies
Check cables
Check insulation
Monitor voltage during charge from built in
charger – fast rise typically Ok
Acid Batteries
• Hydrometers are
available in auto parts
stores for acid batteries
• If the electrolyte level
appears low or the acid
level is too high, add
distilled water
By Palagiri (Own work) [CC BY-SA 3.0 (http://creativecommons.org/licenses/by-sa/3.0)], via Wikimedia Commons
Safety
• Acid batteries - Hydrogen gas may build up if
proper venting has failed: EXPLOSION
• 12.7 volt lead-acid battery can destroy a
finger if a ring shorts the terminals