Transcript PR570TRN

PR570 POWERITE II TRAINING
Dedicated to Success in Electronic
Servicing for 50 Years
“POWERITE II”
6 WAY SAFETY
ANALYZER
1. AC Receptacle Tester & Voltage Monitor
2. AC Variable Power Supply
3. Isolation Transformer
4. Output Current/Power Monitor
5. Output Circuit Breaker
6. Safety Leakage Tester
Equipment
Familiarization
PR570
Displays AC Receptacle
& Isolated Output
AC Voltages
Adjusts ACV at
Isolated Output
Measures Current or
Power Delivered to
Isolated Output
Displays Current, Power,
& Voltage at
Isolated Output
VARIABLE ISOLATION TRANSFORMER &
SAFETY ANALYZER
POWERITE II
AC VOLTS
AC VOLTS
HOT TO
SAFTEY GND
HOT TO
NEUTRAL
CURRENT/POWER
ISOLATED OUTPUT
VOLTAGE
CURRENT/POWER
TRIP RESET
TRIP SET
HOT TO
EXTERNAL
EARTH GND
PR570
Amps
POWER
Amps
ISOLATED
OUTPUT
Watts
HOT
CHASSIS
Valid only when
connected to
External Earth Gnd
TRIP SET
CURRENT
LEAKAGE
uAmps
uAmps
MIN
AUTO-TOGGLE
MAX
MIN
AC Line Safety Gnd
Both Must Light
ISOLATED OUTPUT
POWER
OFF
ON
100 uAMP LEAKAGE
CAL CHECK
TO EXTERNAL
EARTH GND
SAFETY PROBE
INPUT
WARNING:
Set Output to 120VAC &
Touch Safety Probe Here.
Refer To
Pull Chart
Isolated Output
On When Lit
For Equipment
Under Test
Only!
!
3 Amps At 117VAC Continous, 4 Amps Intermittent Use
Measures Input ACV to
Safety Ground
Resets
Isolated Output
Trip Current
Measures Variable ACV at
Isolated Output
Measures AC
Leakage
Select to View
Trip Current
Indicates
Isolated Output
is Active
Adjusts
Isolated Output
Trip Current
Level
Isolated Variable
AC Output
AC Electrical Outlet
AC outlet
120VAC
Hot
Neutral
Earth
Ground
AC Electric Box
Neutral
Hot
(120VAC)
Hot (120V)
Black
Electric
Box
White
Green
Earth
Ground
Rod
Hot To External Earth Ground
VARIABLE ISOLATION TRANSFORMER &
SAFETY ANALYZER
PR570
POWERITE II
AC VOLTS
AC VOLTS
HOT TO
SAFTEY GND
HOT TO
NEUTRAL
CURRENT/POWER
ISOLATED OUTPUT
VOLTAGE
CURRENT/POWER
TRIP RESET
TRIP SET
HOT TO
EXTERNAL
EARTH GND
Watts
HOT
CHASSIS
Valid only when
connected to
External Earth Gnd
Amps
POWER
Amps
ISOLATED
OUTPUT
TRIP SET
CURRENT
LEAKAGE
uAmps
uAmps
MIN
AUTO-TOGGLE
MAX
MIN
AC Line Safety Gnd
Both Must Light
ISOLATED OUTPUT
POWER
OFF
ON
TO EXTERNAL
EARTH GND
100 uAMP LEAKAGE
CAL CHECK
SAFETY PROBE
INPUT
WARNING:
Refer To
Pull Chart
Set Output to 120VAC &
Touch Safety Probe Here.
Isolated Output
On When Lit
For Equipment
Under Test
Only!
!
3 Amps At 117VAC Continous, 4 Amps Intermittent Use
Earth Ground
AC Receptacle Tests
VARIABLE ISOLATION TRANSFORMER &
SAFETY ANALYZER
PR570
POWERITE II
AC VOLTS
AC VOLTS
HOT TO
SAFTEY GND
HOT TO
NEUTRAL
CURRENT/POWER
ISOLATED OUTPUT
VOLTAGE
CURRENT/POWER
TRIP RESET
TRIP SET
HOT TO
EXTERNAL
EARTH GND
Watts
HOT
CHASSIS
Valid only when
connected to
External Earth Gnd
Amps
POWER
Amps
ISOLATED
OUTPUT
TRIP SET
CURRENT
LEAKAGE
uAmps
uAmps
MIN
AUTO-TOGGLE
MAX
MIN
AC Line Safety Gnd
Both Must Light
ISOLATED OUTPUT
POWER
OFF
ON
TO EXTERNAL
EARTH GND
100 uAMP LEAKAGE
CAL CHECK
SAFETY PROBE
INPUT
WARNING:
Refer To
Pull Chart
Set Output to 120VAC &
Touch Safety Probe Here.
Isolated Output
On When Lit
!
3 Amps At 117VAC Continous, 4 Amps Intermittent Use
Hot
Neutral
For Equipment
Under Test
Only!
Earth
Ground
A Variable AC Transformer
Variac or Autotransformer
A variac or autotransformer provides an AC output
voltage variable from 0 to approximately 140 VAC.
A variac does not provide isolation.
Variable AC Power Supply
0-140 VAC
VARIABLE ISOLATION TRANSFORMER &
SAFETY ANALYZER
PR570
POWERITE II
AC VOLTS
AC VOLTS
HOT TO
SAFTEY GND
HOT TO
NEUTRAL
CURRENT/POWER
ISOLATED OUTPUT
VOLTAGE
CURRENT/POWER
TRIP RESET
TRIP SET
HOT TO
EXTERNAL
EARTH GND
Watts
HOT
CHASSIS
Valid only when
connected to
External Earth Gnd
Amps
POWER
Amps
ISOLATED
OUTPUT
TRIP SET
CURRENT
LEAKAGE
uAmps
uAmps
MIN
AUTO-TOGGLE
MAX
MIN
AC Line Safety Gnd
Both Must Light
ISOLATED OUTPUT
POWER
OFF
ON
TO EXTERNAL
EARTH GND
100 uAMP LEAKAGE
CAL CHECK
SAFETY PROBE
INPUT
WARNING:
Refer To
Pull Chart
Set Output to 120VAC &
Touch Safety Probe Here.
Isolated Output
On When Lit
For Equipment
Under Test
Only!
!
3 Amps At 117VAC Continous, 4 Amps Intermittent Use
Hot Ground
Direct off-line DC supply with bridge rectifier results in a “HOT” chassis
“HOT” ground has direct path to electrical hot side of electrical outlet
Approximately 67 VDC between “HOT” ground and Earth
ground
Connecting a ground from a test instrument to a “HOT” ground causes
high current through the diode
Test
Equipment
V
 67 VDC
• An isolation transformer electrically
isolates the “HOT” side of the AC plug,
eliminating the voltage difference and
current that may result between ground
references
Isolation
Transformer
Test
Equipment
V
Defeating Isolation
V
120VAC
Between
Chassis
Grounds
The isolation transformer can be defeated if
twohot chassis are connected at the same time.
Isolated Output On/Off Control
VARIABLE ISOLATION TRANSFORMER &
SAFETY ANALYZER
PR570
POWERITE II
AC VOLTS
AC VOLTS
HOT TO
SAFTEY GND
HOT TO
NEUTRAL
CURRENT/POWER
ISOLATED OUTPUT
VOLTAGE
CURRENT/POWER
TRIP RESET
TRIP SET
HOT TO
EXTERNAL
EARTH GND
Watts
HOT
CHASSIS
Valid only when
connected to
External Earth Gnd
Amps
POWER
Amps
ISOLATED
OUTPUT
TRIP SET
CURRENT
LEAKAGE
uAmps
uAmps
MIN
AUTO-TOGGLE
MAX
MIN
AC Line Safety Gnd
Both Must Light
ISOLATED OUTPUT
POWER
OFF
ON
TO EXTERNAL
EARTH GND
100 uAMP LEAKAGE
CAL CHECK
SAFETY PROBE
INPUT
WARNING:
Set Output to 120VAC &
Touch Safety Probe Here.
Refer To
Pull Chart
Isolated Output
On When Lit
For Equipment
Under Test
Only!
!
3 Amps At 117VAC Continous, 4 Amps Intermittent Use
Output Off - Output On
Switch(Switches Isolated
Output Voltage On or Off)
Isolated Output On
Indicator (On when Lit.)
Isolated Output Current & Power Monitoring
VARIABLE ISOLATION TRANSFORMER &
SAFETY ANALYZER
PR570
POWERITE II
AC VOLTS
AC VOLTS
HOT TO
SAFTEY GND
HOT TO
NEUTRAL
CURRENT/POWER
ISOLATED OUTPUT
VOLTAGE
CURRENT/POWER
TRIP RESET
TRIP SET
HOT TO
EXTERNAL
EARTH GND
Watts
HOT
CHASSIS
Valid only when
connected to
External Earth Gnd
Amps
POWER
Amps
ISOLATED
OUTPUT
TRIP SET
CURRENT
LEAKAGE
uAmps
uAmps
MIN
AUTO-TOGGLE
MAX
MIN
AC Line Safety Gnd
Both Must Light
ISOLATED OUTPUT
POWER
OFF
ON
TO EXTERNAL
EARTH GND
100 uAMP LEAKAGE
CAL CHECK
SAFETY PROBE
INPUT
WARNING:
Refer To
Pull Chart
Set Output to 120VAC &
Touch Safety Probe Here.
For Equipment
Under Test
Only!
!
3 Amps At 117VAC Continous, 4 Amps Intermittent Use
Isolated Output
Current & Watts
Meter
Isolated Output
On When Lit
Isolated Output Circuit Breaker
VARIABLE ISOLATION TRANSFORMER &
SAFETY ANALYZER
PR570
POWERITE II
AC VOLTS
AC VOLTS
HOT TO
SAFTEY GND
CURRENT/POWER
ISOLATED OUTPUT
VOLTAGE
HOT TO
NEUTRAL
CURRENT/POWER
TRIP RESET
TRIP SET
HOT TO
EXTERNAL
EARTH GND
Watts
HOT
CHASSIS
Valid only when
connected to
External Earth Gnd
Amps
POWER
Amps
ISOLATED
OUTPUT
TRIP SET
CURRENT
LEAKAGE
uAmps
uAmps
MIN
AUTO-TOGGLE
MAX
MIN
AC Line Safety Gnd
Both Must Light
ISOLATED OUTPUT
POWER
OFF
ON
TO EXTERNAL
EARTH GND
100 uAMP LEAKAGE
CAL CHECK
SAFETY PROBE
INPUT
WARNING:
Refer To
Pull Chart
Set Output to 120VAC &
Touch Safety Probe Here.
Isolated Output
On When Lit
For Equipment
Under Test
Only!
!
3 Amps At 117VAC Continous, 4 Amps Intermittent Use
Position To Meter
Circuit Breaker
Level 0-4Amps
Reset if
Circuit
Breaker
Trips
Control
Sets
Circuit
Breaker
0-4A
•
If the “HOT” ground of the power supply becomes electronically
connected to exterior parts of the cabinet housing, a shock
hazard exists.
•
Protect yourself, your business and your customer. ALWAYS
perform a leakage current check after completing a repair!
Causes of Excess AC Leakage Current
Defective AC Line or Switching Supply
Isolation Transformers
Leaky AC Bypass Filter Capacitors
Improper Part Installation - Safety Critical
Isolation Parts
Foreign Objects - Coins, screws, paper clips
SMPS Safety Leakage Paths
MIC 1
MIC 3
Switching
Transformer
MIC 2
Startup
Switching
Transistor
Drive
Generator/
Control
Feedback/Control
MIC 4
Switch Mode Transformer Primary to Secondary
AC Input Filter Capacitors
Isolation Filter Capacitors Between Hot & Isolated
Grounds
Effects of AC Leakage
10A
Severe burns, but not fatal
unless vital organs burned
5
4
1A
Heart stops during shock May Restart if current is
removed before death
500
250
100mA
50
Heart fibrillation in 1.4
seconds - Usually fatal
Breathing stops
Often fatal
Highest
Danger
30
10mA
Cannot let go
Painful sensation
5
1mA
Mild sensation
UL limit for consumer
products
500
Imperceptible
100A
UL limit for hospital
equipment not
connected directly to
the body
Safety Leakage Testing
A safety leakage test ensures that the
monitor doesn’t have AC line leakage
to exposed metal parts
Virtually all manufacturer’s service
literature requires a safety leakage test
be performed at the completion of any
repair
Safety leakage testing applies to all
consumer electronic devices powered
from the AC line, whether or not they
have a 3-wire power cord
Underwriter’s Laboratory (UL)
standard is 500a
Other standards may apply
International Electrotechnical
Commission (IEC)
standard is 700 a
Manufacturer Safety Leakage Test
•
Underwriter’s Laboratory (UL) standard is 500a
•
International Electrotechnical Commission (IEC)
standard is 700 a
0.15 F
TO
INSTRUMENT'S
EXPOSED
METAL PARTS
1500  , 10W
WATER
PIPE
(EARTH
GROUND)
1 Do not use an isolation transformer for this test. Plug the completely reassembled
unit directly into the ac outlet.
2 Connect a 1.5K, 10W resistor paralleled by an .015uF capacitor between each
exposed metal cabinet part and a good earth ground such as a water pipe, as
shown above.
3 Use an ac voltmeter with at least 5000 ohms/volt sensitivity to measure the
potential across the resistor.
4 The potential at any point should not exceed 0.75 volts. A leakage current tester
may be used to make this test; leakage current must not exceed 0.5 milliamps. If
a measurement is outside of the specified limits, there is a possibility of shock
hazard. The chassis should be repaired and rechecked before returning it to the
customer.
5 Repeat the above procedure with the ac plug reversed. (Note: An ac adapter is
necessary when a polarized plug is used. Do not defeat the polarizing feature of
the plug.)
Safety Leakage Testing
VARIABLE ISOLATION TRANSFORMER &
SAFETY ANALYZER
PR570
POWERITE II
AC VOLTS
AC VOLTS
HOT TO
SAFTEY GND
HOT TO
NEUTRAL
CURRENT/POWER
ISOLATED OUTPUT
VOLTAGE
CURRENT/POWER
TRIP RESET
TRIP SET
HOT TO
EXTERNAL
EARTH GND
Watts
HOT
CHASSIS
Valid only when
connected to
External Earth Gnd
Amps
POWER
Amps
ISOLATED
OUTPUT
TRIP SET
CURRENT
LEAKAGE
uAmps
uAmps
MIN
AUTO-TOGGLE
MAX
MIN
AC Line Safety Gnd
Both Must Light
ISOLATED OUTPUT
POWER
OFF
ON
TO EXTERNAL
EARTH GND
100 uAMP LEAKAGE
CAL CHECK
SAFETY PROBE
INPUT
WARNING:
Refer To
Pull Chart
Set Output to 120VAC &
Touch Safety Probe Here.
Isolated Output
On When Lit
For Equipment
Under Test
Only!
!
3 Amps At 117VAC Continous, 4 Amps Intermittent Use
Checking Leakage Test Calibration
VARIABLE ISOLATION TRANSFORMER &
SAFETY ANALYZER
PR570
POWERITE II
AC VOLTS
AC VOLTS
HOT TO
SAFTEY GND
HOT TO
NEUTRAL
CURRENT/POWER
ISOLATED OUTPUT
VOLTAGE
CURRENT/POWER
TRIP RESET
TRIP SET
HOT TO
EXTERNAL
EARTH GND
Watts
HOT
CHASSIS
Valid only when
connected to
External Earth Gnd
Amps
POWER
Amps
ISOLATED
OUTPUT
TRIP SET
CURRENT
LEAKAGE
uAmps
uAmps
MIN
AUTO-TOGGLE
MAX
MIN
AC Line Safety Gnd
Both Must Light
ISOLATED OUTPUT
POWER
OFF
ON
TO EXTERNAL
EARTH GND
100 uAMP LEAKAGE
CAL CHECK
SAFETY PROBE
INPUT
WARNING:
Set Output to 120VAC &
Touch Safety Probe Here.
Refer To
Pull Chart
Isolated Output
On When Lit
For Equipment
Under Test
Only!
!
3 Amps At 117VAC Continous, 4 Amps Intermittent Use
Set Isolated Output to
120VAC. Leakage Test
Should Read 100uA +/2 Counts
PR570 Function Review
Displays AC Receptacle
& Isolated Output
AC Voltages
Adjusts ACV at
Isolated Output
Measures Current or
Power Delivered to
Isolated Output
Displays Current, Power,
& Voltage at
Isolated Output
VARIABLE ISOLATION TRANSFORMER &
SAFETY ANALYZER
POWERITE II
AC VOLTS
AC VOLTS
HOT TO
SAFTEY GND
HOT TO
NEUTRAL
CURRENT/POWER
ISOLATED OUTPUT
VOLTAGE
CURRENT/POWER
TRIP RESET
TRIP SET
HOT TO
EXTERNAL
EARTH GND
PR570
Amps
POWER
Amps
ISOLATED
OUTPUT
Watts
HOT
CHASSIS
Valid only when
connected to
External Earth Gnd
TRIP SET
CURRENT
LEAKAGE
uAmps
uAmps
MIN
AUTO-TOGGLE
MAX
MIN
AC Line Safety Gnd
Both Must Light
ISOLATED OUTPUT
POWER
OFF
ON
100 uAMP LEAKAGE
CAL CHECK
TO EXTERNAL
EARTH GND
SAFETY PROBE
INPUT
WARNING:
Set Output to 120VAC &
Touch Safety Probe Here.
Refer To
Pull Chart
For Equipment
Under Test
Only!
Isolated Output
On When Lit
!
3 Amps At 117VAC Continous, 4 Amps Intermittent Use
Measures Input ACV to
Safety Ground
Resets
Isolated Output
Trip Current
Measures Variable ACV at
Isolated Output
Measures AC
Leakage
Select to View
Trip Current
Indicates
Isolated Output
is Active
Adjusts
Isolated Output
Trip Current
Level
Isolated Variable
AC Output
GETTING STARTED:
1. Turn the PR570 power switch “ON.” Confirm proper AC supply voltage to the PR570 by reading
the “HOT TO SAFETY GROUND” and “HOT TO NEUTRAL” voltages (expect 115VAC to
125VAC in the left LCD readout).
NOTE: An open safety ground results in a low “HOT TO SAFETY GROUND” measurement.
2. Turn the AC VOLTS switch to the “ISOLATED OUTPUT” position. Adjust the Isolated Output
Voltage (viewed in the left LCD readout) to 120 VAC using the ISOLATED OUTPUT control.
3. Set the Isolated Output “TRIP SET” to 2 AMPS by placing the CURRENT/POWER switch in
the “TRIP SET” position and adjusting the TRIP SET control until 2.00 is displayed in the
right LCD readout.
Congratulations! You are now
an experienced PR570 User