Biokit Elisa System
Download
Report
Transcript Biokit Elisa System
BEST 2000 Characteristics
BEST 2000 (Biokit Elisa System) is a bench top
analyser with capacity for 4 micro plates. Multiple
assays can be performed per plate. BEST 2000
offers the user an easy to use sample-in /result-out
processor.
The instrument automates all steps of the ELISA
process:
•Positive sample and plate barcode identification
•Sample, control and calibrator dilution and
dispensing (disposable tip)
•Reagent dispensing
•Incubation (with or without shaking)
•Washing
•Reading
•Data reduction
Specifications BEST 2000 Dimensions and Weight
Depth <91 cm (35.8 in)
Width <106 cm (41.8)
Height <80 cm (31.5 in)
Weight <110 Kg (243 lbs)
Footprint 106 cm x 91 cm (41.8 in x 35.8 in)
Capacities
Samples 96
Reagents 24
Controls and/or Standards 33
Sample Pipetting Tips 432
Reagent Pipetting Tips 41
Wash Buffer Bottles 4 bottles, 2 liters each
Waste Container 8 liters
Operation
Ambient Drawer Module
Incubation Temperatures Ambient plus 5 °C
Incubator Module
Incubation Temperatures Ambient plus 7 °C to 50
°C
Assays
Blanking Air
Individual, paired or average wells
Whole plate or last plate, Row or column
Each well on the plate
Wavelength Modes Single (10 sec) or dual (20 sec)
Standard Curves Linear, quadratic, cubic, quartic,
spline, polygon, sigmoid or Akima
Additional Data Analysis Threshold, ratio, QC
equations
Flexible Template Up to eight different well types.
Power Requirements
Voltage Power Frequency
Main Unit 100 - 240 V 800 VA 50/60 Hz
Line Voltage Variation ± 10%
Line Frequency Variation ± 3 Hz
Environment
Operating Range 15° C to 30° C
15% to 85% relative humidity (non-condensing)
Altitude 2000 Meters
Chemwell Fully Automated Chemistry
Analyzer
Stat Fax 303 plus EIA Strip Reader
Stat Fax 2100 EIA Plate Reader
Stat Fax 2600 EIA Plate Washer
Stat Fax 2200 Plate Shaker/Incubator
Stat Fax 3100 Manual Microwell Washer
ChemWell® is a fully automated analyzer
capable of running both EIAs and general
chemistries in standard micro wells.
Awareness Technology, Inc. is proud to
announce the release of the updated Rev 6.0
Software. Currently it is available for
biochemistries ONLY, but the EIA capable
version is nearing completion as well. (for
ChemWells with microplate washers, please
find out about Rev 5.11. This software allows
you to run the EIA's). Some of the features of
the new Rev 6.0 software are:
•Continous Loading - Instead of collecting a
batch of specimens for a run, tests and samples
can now be run on demand, even while other tests
are running.
•Calibrators and Controls can be run at any time
and independently of tests, allowing you to update
stored curves and validate tests as required. If
tests associated with the newly loaded calibrator
were already running, the sample concentrations
for that test will be automatically adjusted for the
new curve.
•Results for individual samples can be accepted or
re-run for verification, simply by clicking a button.
You will no longer have to setup a new job or
reposition sample cups when re-testing.
•Sample and Reagent positions can be moved
around using the Windows® "drag and drop"
feature.
•Assay programming has been greatly simplified.
•Reports format easier to modify.
•Intuitive new interface.
ChemWell® is a completely open system that
is easy to program. Set up assays, routine jobs,
quality control, panels, and even index
calculations to suit your laboratory. With a
completely open system you can program
additional wash volume, increase the number of
washes, or even direct the probe to pick up and
dispense a probe cleaning solution after each
specimen. And it is entirely password protected.
Precise pipetting of low volume specimens,
elimination of carryover, and careful control of
temperature are all essential for reliable
results. ChemWell® can make pre-dilutions,
however, this is often not necessary since a high
level of precision is achieved even with a 2uL
sample. ChemWell® can process EIA at ambient
temperature or 37C. An incubator is provided for
biochemistry wells to provide precise kinetic
reactions. The probe is also temperature
controlled to deliver 37C reagents.
RCA - Reagent Cooling Accessory <RCA.htm>
now available for ChemWell®. Protects reagents
from ambient heat. The removable racks can be
refrigerated, pre-loaded, and ready for the next
use.
ChemWell® is designed
to maximize
dependability, versatility,
and economy.
ChemWell® uses
standard, off-the-shelf
microwells for all
reactions. Performing
biochemistry assays in
microwells offers many
advantages such as
increased throughput and
decreased cost per test.
Probe washes inside and out
Washes 8 wells at once
Uncoated microwells make low volume sample cups
for biochemistry. They can also be washed for reuse. ChemWell® reads vertically using 4 optical
channels and 8 wavelengths.
Handles up to 27 reagents, 96 samples.
Here are more features that you can expect
with ChemWell®:
· 200 tests per hour
· No carry-over
· Liquid sensing probe tip
· 8-well wash head
· Level-sensing wash, rinse, and waste
bottles
· LIS, STAT
· QC tracking
· Self-monitoring
· Reflex testing
· Reaction Volume <250uL
· No custom disposables
· IAD filters
· Creates automatic service reports
· Edit standard curves
· self-monitoring mechanics and optics
All of this and more in one instrument, the
ChemWell®.
ChemWell Analyzer
Wins State of Florida
Recognition
Awareness Technology, Inc. of Palm City, Florida
has been selected to receive the 2005 Governor's
New Product Award for their innovative ChemWell
2910 automatic laboratory analyzer.
This award, co-sponsored by the Florida Engineering
Society, in cooperation with the Governor's Office,
recognizes companies with the foresight to bring new
products to the marketplace. The purpose of the
Governor's New Product Award is to promote the full
spectrum of benefits that come from the research and
engineering of new products and improve quality of
life. This is the 19th year of the program and the first
time the award will go to the clinical diagnostic
industry.
Awareness Technology, a 23 year old Florida
designer and manufacturer of lab instruments,
competed with its new product, ChemWell 2910,
which combines two fully automatic instruments into
one economical package: a continuous loading
biochemistry analyzer and a user-programmable
enzyme immunoassay (EIA) system. This bench-top
instrument performs all test processes including
dispensing and diluting reagents, incubating,
washing, optical reading, calculating, and reporting
results.
Mary Freeman, president of Awareness Technology,
will accept the award on April 6 in the Florida capital
where it will be presented by Mark Mongeau, PE,
President of the Florida Engineering Society. State
winners are eligible to compete for the national New
Product award.
•Automatic strip
transportation.
•3-strip loading capacity or
break-apart, single, and
double strips. The carriers
come in two styles, 3x12 for
break-apart microwell strips,
and an optional 3x8 for nonbreak-apart microwell strips.
•Bichromatic optics.
Standard model includes
four filters, six filters are
available for a nominal
charge.
•Easy to use with step-by-step prompting and onboard graphics printer.
•log-logit mode and other programming
enhancements.
•On-board software calculates single & multi-point
calibration curves,
•regressions, & a variety of cutoff options.
•Large non-volatile memory stores more than 50
user-entered tests.
•NRTL and CE certified.
•Fully automatic.
•Bichromatic optics.
Standard model includes
four filters, six filters are
available for a nominal
charge.
•On board data reduction
calculations include
regressions, curve fitting
and cutoff absorbance
calculations with step-bystep user prompting.
•Large, non-volatile memory
stores at least 50 user tests.
•Includes parallel and serial outputs.
•NRTL and CE certified.
•Fully automatic.
•Washes flat, round, and Vbottom plates and strips.
•No user adjustments
required. Automatic
calibration, alignment, and last
row detection.
•Large, non-volatile memory
stores at least 50 user wash
protocols.
•Precision operation with
constant monitoring of
vacuum and pressure. Pump
cycles only when necessary
•for quiet operation.
•Programmable automatic rinse cycle.
•Ready to run: bottles, aerosol shield, 8-way manifold
included.
•12 and 16 position wash heads available as
accessories.
•NRTL and CE certified.
•Holds two 96 well
microplates.
•Digitally controlled 8speed shaker with orbital
mixing from 575 to 1500
rpm.
•Temperature adjustable
from ambient to 40°C with
resolution of 0.1°C.
•Digital timer controls
shaking or can be operated
independently as a timer.
•Compact and easy to
operate.
•Outer tinted cover provides further insulation and
protection from light.
•NRTL and CE certified
•No more squirt bottle
washing. Reduce exposure
to the potential dangers of
splashing biohazards with
Stat Wash®, an economical
instrument designed to
wash immunoassays in
microwell strips and plates
with professional precision.
Washing 8 wells
simultaneously is faster
too.
•Aspirating and dispensing are controlled by push
buttons. Easy to use and maintain, Stat Wash®
includes an 8-probe manifold, a 1L wash bottle, and a
1L safe plastic bottle with a full sensor. Stat Wash™ is
CE certified.
•If your lab doesn't need an expensive automated
instrument, simply switch to Stat Wash® for clean,
fast, and precise results.
Biokit, Spain
· Bioelisa HCV
· Bioelisa HBsAg
· Bioelisa HIV
· Bioelisa Syphlis
Standard Diagnostic,
Korea
· HCV
· HBsAg
· HIV
· Malaria
· Syphilis
DiaSys, Germany
· ATP Hexokinase
Solid phase: Guinea pig polyclonal
antibodies anti-HBs subtype ad+ay
Conjugate: Goat polyclonal
antibodies anti HBs subtype ad+ay
Polyclonal antibodies assure the
detection of possible virus mutations .
Qualitative assay:
COV:Negative+0,040
Sensitivity:
- 0,034 PEI/ml ad antigen
- 0,024 PEI/ml ay antigen
- 0,054 UI/WHO standard
- 0.2 ng/ml
Specificity: 99,9%
1 hour 37°C
Cod: 3000-1100 96 Tests
Cod. 3000-1180 480 Tests
TMB
30 min 37°C 30 min RT
Solid phase: Recombinant antigens from the
Core, NS3 NS4 and NS5 regions
Conjugate: Anti-human IgG labelled with
peroxidase
Sample dilution 1/20 in plate (IPD)
Qualitative assay:
- COV: Negative + 300
Sensitivity:
99,9%
Specificity: 99.87
%
TMB
Cod: 3000-1117 96 Tests
Cod. 3000-1119 480 Tests
1 hour 37°C
30 min 37°C 30 min RT
• Solid phase: Recombinant antigens:
HIV-1: p24, gp41, gp120
HIV-2: gp36, gp125
Breakable wells
• Conjugate: Anti-human IgG and IgM labelled
with peroxidase
• Sample dilution in plate 1/20 (IPD)
• Subtype O Detection
• Qualitative assay: COV:
Positive control x 0.32
• Sensitivity: 100%
• Specificity: >99.5%
OPD
Cod: 3000-1143 96 Tests
Cod. 3000-1144 480 Tests
Spectrafuge 6C
Speed Control: Digital
– Variable (100 rpm inc.)
Max. Speed: 6500 rpm
Max g-force 4000xg
Capacity: 6x10ml
Timer: 1-30 min or
continuous
Rotor angle: 30 degree
Dimentions (WxDxH) 8.25x9.5x7 in
(21x24x18cm)
Weight: 10lbs / 4.5 kg
Electrical: 230V, 50-60 Hz