phys586-lec20-ion3
Download
Report
Transcript phys586-lec20-ion3
Geiger-Mueller Tube
Introduced in 1928 by Geiger and
Mueller but still find application today
Used in experiments that identified the He
nucleus as being the same as the alpha
particle
1
Geiger-Mueller Tube
Operation
Increasing the high voltage in a proportional tube
will increase the gain
The avalanches increase not only the number of
electrons and ions but also the number of excited gas
molecules
These (large number of) photons can initiate
secondary avalanches some distance away from
the initial avalanche by photoelectric absorption in
the gas or cathode
Eventually these secondary avalanches envelop
the entire length of the anode wire
Space charge buildup from the slow moving ions
reduce the effective electric field around the
anode and eventually terminate the chain reaction
2
Geiger-Mueller Tube
3
Geiger-Mueller Tube
Gas
The main component is often argon or
neon
However when the large number of these
noble ions arrive at the cathode and are
neutralized, the released energy can cause
additional free electrons to be liberated
from the cathode
This gives rise to multiple pulsing
(avalanches) in the G-M tube
4
Geiger-Mueller Tube
Gas
Multiple pulsing can be quenched by the addition
of a small amount of ethyl alcohol, chlorine (Cl2),
or bromine (Br2) (the quench gas)
As mentioned in an earlier lecture, collisions
between ions and different species of gas
molecules tend to transfer the charge to the one
with the lowest ionization potential
When the halogen ions are neutralized at the
cathode, disassociation can occur rather than
extraction of a free electron
5
Geiger-Mueller Tube
Use
Geiger tubes are often used as survey
meters to detect or monitor radiation
They are rarely used as dosimeters but there are
some applications
Survey meters generally have units of CPM
or mR/hr but beware/check the calibration
information
If calibrated, the survey meter is calibrated
to some fixed gamma ray energy
For other gamma ray energies one must account
for differences in efficiency
6
Geiger-Mueller Tube
7
Geiger Tube
How is 900V generated from 2x1.5V
batteries?
Diodes are nonlinear circuit elements that
only conduct current in one direction
8
Geiger Tube
Voltage doubler
9
Geiger Tube
On one half-cycle, D1 conducts and charges C1
to V
On the other half-cycle D2 conducts and
charges C2 to 2V
A long string of half-wave doublers is known as
a Cockcroft-Walton multiplier
10
Geiger Tube
This can be extended to an n multiplier
11
Cockcroft-Walton Accelerator
12
Proportional Counters
Many different types of gas detectors have
evolved from the proportional counter
13
Proportional Counters
Most of these variants were developed to
improve position resolution, rate capability,
and/or cost
MWPC (multi-wire proportional tube)
CSC (cathode strip chamber)
Drift chamber (e.g. MDT)
Micromegas (micromesh gaseous detector)
RPC (resistive plate chamber)
Nearly every application has made some
attempt to transfer to medical applications
14
Momentum Measurement
Let v, p be perpendicular to B
qvB
mv2
pT GeV 0.3B T m
L
sin
2
2 2
0.3LB
pT
2 0.3L2 B
s 1 cos
2
8
8 pT
15
Momentum Resolution
The sagitta s can be determined by at least 3
position measurements
This is where the position resolution of the
proportional chambers comes in
x1 x3
s x2
2
3
s x
2
pT s
pT
s
3
x 8 p
2
2
0.3BL
16
Magnets
Solenoid
Large homogeneous
field
Weak return field in
return yoke
Dead material in
beam
Toroid
Field always
perpendicular to p
(ideal)
Large volume
Non-uniform field
Complex
17
Magnets
ATLAS
CMS
18
Magnets
19
Momentum Resolution
ATLAS muon momentum resolution
20
Multiwire Proportional Chambers (MWPC’s)
Nobel prize to Charpak in 1992
Simple idea to extend the proportional tube
Effectively spawned the era of precision high energy
physics experiments
21
MWPC’s
You might expect that because of the large C
between the wires, a signal induced on one
wire would be propagated to its neighbors
Charpak observed that a positive signal would
be induced on all surrounding electrodes
including the neighbor wires (from the positive
ions moving away)
22
MWPC’s
Typical parameters
Anode spacing – 1-2 mm
Anode – cathode spacing – 8 mm
Anode diameter – 25 mm
Anode material – gold plated tungsten
Cathode material – Aluminized mylar or
Cu-Be wire
Typical gain - 105
23
Cathode Strip Chambers (CSC)
The negative charge induced on the anode
induces positive charge on the cathodes
This provides a second detectable signal
If the surface charge density is sampled by
separate cathode electrodes then the location of
the avalanche can be determined
If the cathode pulse heights are well measured
the position resolution can be precisely
determined (~100μm vs 600μm for 2mm/√12)
24
Cathode Signal
Consider the geometry
The cathode charge distribution is given by
Where λ = x/d and Ki are geometry dependent
constants
25
Cathode Signal
The shape is quasiLorentzian with a
FWHM ~ 1.5 d,
where d is the
anode-cathode
spacing
26
Cathode Signal
In order to reduce
the number of
readout channels
one can use
capacitive coupling
between strips
Strip pitch is onehalf or one-third
Readout pitch
stays the same
27
ATLAS Muon System
28
ATLAS Muon System - Barrel
29
ATLAS CSC’s
30
ATLAS CSC’s
31
ATLAS CSC’s
Some numbers
16 four-layer CSC’s per side
Both r (precision) and f (transverse) position is
measured for each layer
Each CSC has 4 x 192 precision strips
Each CSC has 4 x 48 transverse strips
32,000 channels total
32
ATLAS CSC’s
33
ATLAS CSC’s
34
ATLAS CSC’s
35
Drift Chambers
Another variation on the MWPC is the drift
chamber
36
Drift Chambers
Advantages
Better position resolution
Smaller number of channels
Disadvantages
More difficult to construct
Need time measurement
The position resolution of drift chambers is
limited by diffusion, primary ionization
statistics, path fluctuations, and electronics
Many different geometries are possible
37
Drift Chambers
Planar chambers
38
Drift Chambers
CDF central tracker
39
ATLAS MDT’s
40
ATLAS MDT’s
41
ATLAS MDT’s
42
ATLAS MDT’s
Some numbers
~1200 drift chambers with ~400000 drift
tubes
Covers ~5500 m2
Optical monitoring of relative chamber
positions to ~ 30mm
Ar:CO2 (93:7) pressurized to 3 bar
Track position resolution ~ 40mm
43
Micromegas Detector
44
Micromegas
Principle of operation
Bulk micromegas use photolithographic techniques
to produce narrow anodes and precise micromesh –
anode spacing
45
Micromegas
46
Micromegas
47
Resistive Plate Chambers (RPC’s)
Principle of operation
Very high electric field (few kV/mm) induces
avalanches or streamers in the gap
High resistivity material localizes the avalanche
Signal is induced on the readout electrodes
48
RPC’s
Avalanche mode
Like a proportional
chamber
Streamer mode
Small “spark”
Excellent time
resolution
1-2 ns
r 0.1 cm 2
In both cases charge
must recover to reestablish E field after
avalanche or streamer
+++++++++++++++
___________
Before
+++
___
After
+++++
____
49
RPC’s
50
ATLAS RPC’s
HV
X readout
strips
Y readout
strips
Bakelite
Plates
Gas
Foam
PET spacers
2mm gas gap
8.9kV operating voltage
Grounded
planes
Graphite
electrodes
51
ATLAS RPC’s
A few notes on linseed oil
The linseed oil lowers the current draw
through the gas and the singles rate by a
factor of 5-10
It makes a smooth inner surface which gives a uniform
electric field
It absorbs UV photons produced in the avalanche
Babar RPC’s had problems associated with
linseed oil
52
Radiation Units
Exposure
Defined for x-ray and gamma rays < 3 MeV
Measures the amount of ionization (charge Q) in a
volume of air at STP with mass m
X == Q/m
Basically a measure of the photon fluence (F = N/A)
integrated over time
Assumes that the small test volume is embedded in a
sufficiently large volume of irradiation that the number of
secondary electrons entering the volume equals the
number leave (CPE)
Units are C/kg or R (roentgen)
1 R (roentgen) == 2.58 x 10-4 C/kg
Somewhat historical unit (R) now but sometimes still
found on radiation monitoring instruments
X-ray machine might be given as 5mR/mAs at 70 kVp at
100 cm
53
Radiation Units
Absorbed dose
Energy imparted by ionizing radiation in a volume
element of material divided by the mass of the
volume
D=E/m
Related to biological effects in matter
Units are grays (Gy) or rads (R)
1 Gy = 1 J / kg = 6.24 x 1012 MeV/kg
1 Gy = 100 rad
1 Gy is a relatively large dose
Radiotherapy doses > 1 Gy
Diagnostic radiology doses < 0.001 Gy
Typical background radiation ~ 0.004 Gy
54
Geiger Tube
Notes
Survey meters generally have units of CPM or mR/hr
Generally the Geiger tube is not used to determine
the absorbed dose
The G-M tube scale is in mR/hr – what is the
absorbed dose?
Dairabsorbed
XW dose in air is
The
Dair
Dair
C / kg
J
2.58 10
33.97
R
C
2 Gy
X 0.876 10
R
4
55
Geiger Tube
56
Relations
Absorbed dose and kerma
D K col K 1 g
g is the radiative fraction
g depends on the electron kinetic energy as well as
the material under considerat ion
The above relation assumes CPE
In theory, one can thus use exposure X to
determine the absorbed dose
Assumes CPE
Limited to photon energies below 3 MeV
57