SMART Infusion Pump
Administration of Intravenous Therapy &
Medications at North York General Hospital
January 2013
1
Purpose of Using SMART Pumps
•
SMART pumps are a Patient Safety Initiative.
•
“SMART medication delivery systems have the potential to dramatically
improve patient medication safety”. (OHTAC, July 2009)
•
Traditional pumps are volumetric – SMART pumps are dose specific.
•
Types of alerts are directly related to Drug Library and Dose Error
Reduction Systems (DERS).
2
SMART Pump Technology
SMART pumps are infusion pumps that allow intravenous (IV) medications to
be delivered at precise rates or in specific amounts. SMART pump technology
offers:
•
Separate Clinical Care Areas (CCA)
•
Customized drug libraries
•
Listing of medications used in particular CCA
•
TALL-man lettering for look–alike medications
•
Soft and/or hard drug limits on each medication
•
Priority listing of medications that are most frequently used
•
Standardized IV concentrations specific to NYGH
3
Policy
All regulated health care providers (RHCP) may operate infusion pumps (other
than syringe pumps) according to their scope of practice once the following
competency requirements are met:
a)
Self-Directed Infusion Pump e-Learning Module on My Learning Edge
(MLE)
b)
Passing score of 80% on the Infusion Pump quiz
c)
Submit the completed/signed Infusion Pump Competency Checklist to the
Clinical Nurse Educator and/or the charge nurse from Medical Imaging
d)
Annual recertification on Infusion Pump education on MLE
4
Infusion Standards
An infusion pump and tubing should be used for all patients receiving IV medications, solutions and
blood/blood products. In emergency situations, IV infusions may be administered without an infusion
pump in order to prevent delay of treatments.
BD Nexiva closed catheter system – SINGLE lumen must be used to deliver IV infusions.
Exceptions: IV infusions in Day Surgery, Operating Room, Post Anaesthesia Care Unit, Endoscopy,
Child and Teen Unit, and NICU
All RHCPs should select the appropriate CCA on an infusion pump which should correspond to the
nursing area that they are working in.
Infusion pumps MUST BE used on:
•
High alert medications/infusions (Medication Administration Policy II-226).
•
Medications which specifically state “infusion pumps required” in the IV manuals posted on the
Intranet.
•
Paediatric/neonatal patients who require IV infusions and medications.
Exceptions: An infusion pump will not be used in Day Surgery, Operating Room, Post Anaesthesia
Care Unit, and Endoscopy. IV infusions will be administered via gravity using a buretrol in these areas.
5
Hospira Symbiq Infusion Device
6
Symbology
!
•
On/Off button
• Primary line icon
•
Silence button
• Bolus icon
•
Emergency stop button
• Piggyback icon
•
Load/Eject button
• Lock/Unlock button – password
activated
7
Symbology
OVERRIDE SYMBOLS indicate that the programmed dose or rate is
outside the predefined parameters
•
Upper Soft Limit
•
Lower Soft Limit
•
Upper Hard Limit
•
Lower Hard Limit
•
No Rule Set icon is displayed when pump is being operated
without rule sets.
8
Symbology
•
Battery symbol is displayed bottom right. Symbol
will begin to turn black as battery is depleted.
•
Wireless Connection Available icon is displayed
when the device is connected to a wireless
network.
•
Soft dose limits indicate that dosage is outside the
recommended range but can be overridden.
•
Hard dose limits indicate that dosage is unsafe
and cannot be overridden.
9
Priming an Administration Set
Note: All administration sets are LATEX FREE.
Drip Chamber
1.
Close all clamps and the Flow Stop.
2.
Spike the IV bag, squeeze the drip chamber to fill to
score mark.
3.
Invert the cassette with the purple collar facing down.
4.
Open the Flow Stop and any other clamps and allow
fluid to flow freely.
5.
Tap the cassette to release any residual air.
6.
Close Flow Stop.
Purple Collar
Cassette
(Flow Stop within cassette)
Roller Clamp
10
Loading Infuser Cassette
Note: The IV pump must be powered on to load or unload the cassette.
1.
Press the On/Off button for 2-3 seconds to power
2.
Press the Load/Eject button.
3.
Grasp the tubing above the cassette with one hand and
below the cassette with the other hand. Insert the cassette
in the carriage with the purple collar facing upwards and the
white Flow Stop facing inwards toward the infuser. Ensure
the cassette is seated at the bottom of the carriage and the
tubing is aligned vertically.
4.
Press the Load/Eject button or wait for the carriage to
close automatically (approximately 10 seconds).
5.
Secure administration set in upper tubing guide located below
the display screen.
Purple Collar
on the infuser.
Flow Stop
11
Removing Infuser Cassette
1. Confirm the infuser is in STOP mode, and close the roller clamp prior to removing the
cassette from the infuser.
2. Press Load/Eject to open the cassette carriage. Ensure the cassette carriage is fully
opened.
CAUTION: Do not remove the administration set until
the movement of the cassette carriage has stopped.
3. Grasp the tubing above and below the cassette. Slide the cassette upwards out of the
cassette carriage.
4. The cassette carriage will close automatically in approximately 10 seconds or press
Load/Eject to close the cassette carriage sooner.
5. To discontinue use of the device, press and hold the On/Off button for 2-3 seconds to
power down. Discard the administration set and IV bag as per hospital guidelines.
12
Basic Infusion
Powering the pump on
1. Press On/Off button for 2-3 seconds to power on.
Power
Button
2. The New Patient dialog box defaults to Yes only if
pump has been off for less than 5 hours.
If you don’t see this screen, please move to next page.
The pump settings are cleared when pump
has been off for 5 hours or more.
3. Press Continue to clear all previous program info.
To retain previous program parameters and settings:
• Press the New Patient field and select No.
• Press Continue to display the Patient Information screen.
13
Basic Infusion
Selecting CCA (Clinical Care Area)
1. Press Select CCA field.
2. Scroll down by continually pressing
the down arrow, select appropriate CCA.
3. Press Done.
14
Basic Infusion
Single channel infuser automatically moves to program screen. On a double
infuser, touch anywhere within the “A” screen to select channel A.
Note tabs across top of Channel A:
•
Bolus
•
Basic (same as primary infusion
and is already selected)
•
Piggyback
Press Select Infusion field.
15
Basic Infusion
•
To scroll through Drug List a full screen at a time (6 medications), use dark
orange arrows.
•
To scroll one medication at a time, use inner grey arrows.
16
Basic Infusion
1.
Select IV fluid.
2.
Press Rate field, program rate and press Enter.
Maximum rate is 1000 ml/hr
3.
Press VTBI (Volume To Be Infused) field, program volume and
press Enter.
4.
Time will automatically be calculated; however, a change
to time will automatically recalculate VTBI.
DO NOT CHANGE THE TIME FIELD after initial
programming. This alters the original value that was
entered in VTBI. Therefore, only change the rate if needed.
3.
Press Next A.
4.
Review confirmation screen for accuracy and press Start Basic A.
5.
Confirm the green channel LED light is lit, and the primary line icon is visible.
Touch the screen to display split channels and the Larger Font Size (“Far View Screen”), or it will
automatically default to Far View Screen in 20 seconds.
17
Piggyback Infusion
1.
Do NOT stop Basic Infusion.
2.
Tap screen to move to near view screen.
3.
Press Piggyback button.
4.
Press Select Infusion field – select drug and concentration.
Note: If required, press Weight field and enter the patient’s weight.
5.
Press VTBI field, program volume and press Enter.
6.
Press Time field, program time and press Enter.
7.
Rate will automatically be calculated; however, a change to time will alter the VTBI.
DO NOT CHANGE THE TIME FIELD after initial programming. This alters the original value that was
entered in VTBI and medication delivery. Therefore, only change the rate if needed.
8.
Press Next A and press OK to confirm Clinical Note.
9.
Press Start Piggyback A.
10. Confirm the green channel LED light is lit, and the piggyback icon is visible.
18
Piggyback Infusion
•
“Piggyback” and “Secondary” are interchangeable terms.
•
Piggyback option is only available if a primary infusion has been programmed and
confirmed.
•
The maximum rate a piggyback infusion can be programmed is 500 ml/hr.
•
A pop-up screen will appear as a reminder before the start
of every piggyback infusion.
•
When a piggyback infusion is completed, the infuser automatically switches back to
the primary infusion at the pre-programmed rate.
•
Rate is automatically calculated when VTBI and time are entered; however, a change
to time will alter VTBI. DO NOT change the time field after initially programming. Only
change the rate if needed.
19
Outside Limits
If the dose or rate entered is higher or lower than
the predefined parameters, an Outside Limit Message
will appear. The RHCP must first ensure that it is not a key
stroke error.
Note:
•
Doses/rates within the green zone are safe.
Only doses/rates within the soft limits can be overridden.
Hard limits cannot be overridden.
An icon will be displayed indicating the delivery is outside the predefined parameters.
20
Soft/Hard Limits
Soft Limit Alert:
•
Review the medication order and IV Manuals to determine whether an override of the
limit is warranted.
•
Consult the most responsible physician if required. Pharmacy is available as a resource.
•
Press Edit to re-enter a value or press Override to accept the dose and begin infusion.
•
Rationale must be documented as per unit standards.
Hard Limit Alert:
•
Review the medication order with the physician. Pharmacy is available as a resource.
•
Upon physician approval, deliver the medication using the “Other Drug” option.
•
Independent double check on the MAR/eMAR must be performed before delivering the
medication as “Other Drug”.
•
Rationale must be documented as per unit standards.
21
“Other Drug” Option
Caution: The use of “Other Drug” option does not provide any drug dosing limits as a safety feature.
Therefore, “Other Drug” option should only be used in the following 3 situations:
1.
“Hard Limit Alert” with physician approval.
2.
Medications that are NOT listed in both the IV manuals or the current CCA’s drug library.
3.
Medications that are listed in the IV manuals but NOT listed in current CCA’s drug library.
Every time the “Other Drug” option is used, the following steps must be done:
•
Review IV manuals and physician orders.
•
Consult the most responsible physician for approval before proceeding.
•
Pharmacy is available as resource.
•
An independent double check must be performed.
•
The rationale for using this option must be documented as per unit standards.
•
Pharmacy must be notified by paper/electronic reporting tool.
22
“Other Drug” Option
1.
Select Other Drug.
2.
Select [ _ _/_mL] concentration.
3.
Press Select and press the appropriate medication unit.
4.
Press Total Drug Amount in Container field, program drug amount and press Enter.
5.
Press Total Volume in Container field, program volume and press Enter.
6.
Press Enter again.
7.
Press Select, choose unit from list, program dose and press Enter.
8.
Press VTBI field, program volume and press Enter.
9.
Press Time field, program time and press Enter.
10. Rate will be calculated automatically; however, a change to time will alter VTBI.
DO NOT CHANGE THE TIME FIELD after initial programming.
This alters the original value that was entered in VTBI and medication
delivery. Therefore, only change the rate if needed.
11. Press Next A and press OK to confirm clinical note.
12. Press Start Piggyback A.
23
Bolus Infusion
1. Press Bolus.
2. In the Deliver From field, select Primary
Container as the bolus source.
3. In the When Bolus Completes field, when there
is a basic infusion running, select Resume Primary
Infusion or Stop Infusion upon completion of
the bolus.
4. Press Next A.
24
Bolus Infusion
5.
Enter weight if applicable.
6.
Select Rate field, VTBI field, and Time field to enter bolus parameters.
7.
Rate will automatically be calculated when VTBI and Time are entered;
however, a change to time will alter VTBI.
DO NOT CHANGE THE TIME FIELD after initial programming.
This alters the original value that was entered in VTBI and
medication delivery. Therefore, only change the rate if needed.
8.
Press Next A to advance to the Confirm Bolus screen.
9.
Verify that the displayed values are correct.
10. Press Start Bolus to begin Bolus infusion and confirm the Bolus icon is visible.
11. Verify flow from the appropriate fluid container and confirm the green channel LED indicator is
illuminated.
25
Titrate/Change Rate, Dose or VTBI
without Stopping Infusion
1. From the far view screen, press either the Dose,
Rate or VTBI button to titrate.
2. Program the new value in the appropriate field
and press Enter.
Titration
Buttons
3. Press Next A to display the confirmation screen.
4. Verify that the displayed values are correct.
5. Press Start Titration A to begin infusion.
26
Stopping Programs
1. Tap the far view screen to bring up the near view screen.
2. Press Stop button to stop current infusion.
Clearing Programs
1. Bring the near view screen up and select current
infusion type (e.g. Bolus, Basic, Piggyback).
2. Press Stop button to stop current infusion.
3. Select the current infusion type.
4. For Basic and Piggyback, press Clear to delete all entries.
For Bolus, press Clear All to delete all entries.
5. The Clear Program pop-up is displayed. Press Continue
to clear all programmed values.
27
End of Infusion Alarm
•
Alarm activates when the programmed VTBI completes.
1.
Depending on the medication, at the end of infusion alarm, the pump
will do one of the following two things:
2.
•
Continue at same rate for all high alert infusions OR
•
Revert to the preset rate on primary line for all other medications
For maintenance infusion, at the end of infusion alarm, the pump will
default to the designated KVO rates to prevent loss of IV access.
KVO Rates for Basic Infusion
NICU
Paediatrics
Adults
1-2 ml/hr
0-10kg
5 ml/hr
10-30kg
10 ml/hr
30-50kg
10 ml/hr
20 ml/hr
28
End of Infusion Alarm
Steps for resolving End of Infusion Alarm:
1.
Press the Silence button on the top of pump.
2.
Press VTBI and the keypad will pop up.
3.
Program volume for new VTBI and press Enter.
4.
Press Next A.
5.
Confirm that the program is correct.
6.
Press Start Titration A.
7.
Touch screen one more time so that it reverts to “far view screen”.
29
Logs Icon
1. Press Logs icon at bottom of screen.
2. Press Shift Totals.
Shift Total Screen
3. Shift totals are broken down by Channel A,
Channel B, and Infuser Shift Totals (All
Programs).
4. Press Clear A, Clear B, or Clear Both.
5. Press Yes to confirm.
6. Press Exit then Exit.
30
Standby Icon
•
•
•
Standby is a feature that enables the clinician to put the infusion in a “sleep mode”.
It allows the clinician to suspend the infusion without losing the current settings or
patient information.
When the pump is in the Standby mode, the pump is powered on and the alarms are
disabled.
Examples for using the Standby mode:
• Pre-program pump settings prior to admission
• Bed bath/shower, patient transfer or repositioning
• Saline lock between medication delivery
• Patient leaving unit/hospital on a pass
• MRI (pump cannot go into room)
To Activate Standby:
1. From any infuser screen, locate and press Standby.
2. Select the channel to be placed in Standby.
31
Patient Transfer
Upon patient transfer to a different unit, the receiving RHCP must:
•
reprogram the pump with the appropriate CCA.
•
verify physician’s IV orders as soon as possible.
At the beginning of each shift, the RHCP must verify that:
•
the infusion against pump settings.
•
the appropriate drug library is being used.
Two ways to reprogram infusion pumps:
•
clear settings from previous CCA and reprogram infusions under current CCA, or
•
change settings from previous CCA to multi-CCA
32
Patient Transfer – Multi CCA
1. Press CCA icon at top of screen.
2. Press current CCA Field (e.g. ICU) and select
desired CCA.
3. Press Yes when Change CCA? screen appears.
4. Press OK when Multiple CCAs screen appears.
5. Press Done.
your
Important Note:
• All previous infusions will continue to run under initial CCA until VTBI is completed
OR program is cleared. Your general CCA label will say Multiple CCA.
•
When the infusion is completed, silence the alarm and stop the infusion. Press
desired program tabs (e.g. Basic) to enter more volume. Press OK when Program is
being CLEARED screen appears. Pump will automatically switch to new CCA.
Channel will now operate under the new CCA drug library.
33
Locking the Keypad
Steps to lock the keypad:
1. Press the Lock button to display
the Enter Code window.
2. Enter the code 4 – 3 – 2 – 1.
3. Press Enter.
4. Screen is locked and the Lock button
is now red.
34
Troubleshooting
The Channel LED and Alarm Tab appear in red indicating High Urgency Alarm.
Air-In-Line
Remove air from administration set. Press Start to clear alarm.
Distal Occlusion
A closed clamp, clogged filter, or other occlusion prevent fluid flow between infuser and
patient. Correct the problem that causes the occlusion and press Start to clear alarm.
Proximal Occlusion
A closed clamp or other occlusion prevents fluid flow between the fluid container and
infuser. Correct the problem that causes the occlusion and press Start to clear alarm.
35
Troubleshooting
White Screen and Touchscreen Drifting
White screen and touchscreen drifting are both glitches in the software. When a
white screen is displayed or drifting occurs, the IV tubing set must be removed
by using the Emergency Release on the bottom of pump.
Note: The pump is now inoperable and
must be sent to bio-med for service.
36
Pump Cleaning
•
As with all shared equipment, pumps must be cleaned between patients
and when visibly soiled.
•
The exterior of the pump may be wiped with CaviWipesTM.
DO NOT use ViroxTM wipes!!!
•
Cleaning
Lock
If the pump is still infusing, activate the
Cleaning Lock button located on the back of
the pump before cleaning.
37
Pump Cleaning
•
To activate the Cleaning Lock, press and hold the button for one second,
then release it. The Cleaning Lock Active message displays.
•
To deactivate the Cleaning Lock, press the button for one second, then
release it.
•
The Cleaning Lock can also be deactivated by pressing the ON/Off button
or the Emergency Stop button.
38
Drug Library Update
Periodic updates to the drug libraries will be done by Pharmacy.
1. If the SMART pump is infusing when the new drug library is
uploaded, a yellow alarm will display.
• Press the yellow tab to minimize the alarm.
• You can continue with the present infusion.
2. The next time the pump is turned off, the LIBRARY TRANSFER
IN PROGRESS window will display.
3. Silence the alarm when the LIBRARY TRANSFER COMPLETE window appears, or
the device will automatically turn off in 30 seconds.
4. The next time the pump is turned on, the NEW LIBRARY ACTIVATED window will
display. Press OK to continue programming.
39
References
Institution for Safe Medication Practices Canada. (2009). Proceedings from the
ISMP Summit on the Use of SMART Infusion Pumps: Guidelines for
Safe Implementation and Use. Retrieved from
http://www.ismp.org/tools/guidelines/smartpumps/printerVersio n.pdf
North York General Hospital. (2012). Infusion Pump Policy.
North York General Hospital. (2012). Medication Administration Policy II-226.
Ontario Health Technology Advisory Committee. (2009). OHTAC
Recommendation: Implementation and Use of Smart Medication
Delivery Systems. Retrieved from
http://www.health.gov.on.ca/english/providers/program/ohtac/tech/
recommend/rec_smd_20090701.pdf
York Central Hospital. (2011). Hospira Symbiq Infusion System: Self-Paced
Training Module.
40