Operation instruction
Download
Report
Transcript Operation instruction
July 2007
P48610 Rev.2
1
SmartInfuser PainPump™
P49220, P49224N
Operation Instructions
Smart Flow Control Systems
אפליקציות טיפוליות נוספות
2
System Contents
Regulating Set
(SmartInfuser Set ™)
Carry pouch
- Carry Pouch
- Patient Guidelines and
Instructions
3
Catheter Set (for P49220 only)
- SmartCatheter
- Peal away introducer
- Catheter Instructions
Syringe
Compression Unit
- Compression Unit loaded
over empty 500 ml bag
- Pump labels
- Medication labels
Wound Dressing
Physician Guidelines and
Instructions
Notes
• Intended for patients under the care of physician or licensed health care provider.
• The SmartInfuser PainPump™ is for single patient use only. Do not re-sterilize.
• Medications used with this system should be administrated in accordance with the
instructions provided by the drug manufacturer.
• Patients should be advised to contact their physician or licensed health care
provider if any of the system connections become disconnected or the catheter
exits the surgical site.
• Avoid dropping the pump. In the event of damage or leakage, advise patient to
contact their physician or licensed health care provider.
•In case of emergency, advise patients to close the clamp on the administration set
in order to stop the flow of medication immediately.
• Do not suture catheter.
4
Considerations
•SmartInfuser PainPump™ is not for intravenous,
intra-arterial, or epidural.
• SmartInfuser PainPump™ is not intended for
delivery of blood, blood products or lipids .
5
(1) Prepare the Administration Set
SmartInfuser Set ™
Catheter Set
The SmartCatheter and SmartInfuser Set ™ comprise the Administration Set.
Remove the SmartInfuser Set ™ and the SmartCatheter from the sterile packaging.
6
(2) Connecting the Administration Set
(a)
Connect the
SmartCatheter to the
Regulating Set using
the luer connectors.
(b)
Tear off the bag port
cover and insert the
spike.
(c)
Verify the spike is
positioned all the
way in.
7
(3) Fill the Pump
C
(b)
(a)
Position the SmartReg® at
C (Closed position).
Warning:
Use aseptic technique to fill the pump.
After use, handle and discard in
accordance with standard medical
protocol.
8
I.
Fill the syringe with medication.
II.
Drain any remaining air, and connect it to
the set Check Valve.
III. While the Check Valve is supported by the
table, push down the syringe plunger until it
is empty.
IV. Repeat the filling process until the desired
fill volume is reached.
Maximum volume: 500ml
(4) Prime the System
R
P
(b)
(a)
Position the
SmartReg® at P
(Priming position).
I.
IMPORTANT: Hold BOTH the Compression
Unit AND the Mutli-Bolus with tubes
positioned vertically/upward .
II.
The SmartCatheter is flushed during priming.
III. Set the SmartReg® at R (Regulating Position)
once all air bubbles have drained from the set
tube and the catheter been flushed.
9
(5) Set the Regulating Set
Push to Lock
10
I.
Set the system flow rate by dialing the
Adjusting Ring: the digit aligned
represents the particular ml/hr.
II.
Lock the selected flow rate by pushing
the Adjusting Ring toward the regulator.
It will “click” when locked in place.
(6) Set the Multi-Bolus
(a)
Set the bolus volume by
dialing the Push Button
using the Key: the digit
represents the particular ml
selected for the bolus.
11
(b)
Lock the selected volume
by pushing in the Safe
Lock using the Key.
(7) Placing the Introducer
a
b
• While holding the T-handle, remove the protector from the introducer needle.
• Insert the introducer needle through the skin to the wound site at a distance of
approximately 1.5” to 2” (see photo a).
• While holding the T-handle, pull out the introducer needle (see photo b).
12
(8) Placing the Catheter
c
d
• Cut the catheter infusion segment to the desire length (see photo c).
• Insert the catheter through the opening of the introducer’s T-handle to
the wound site (see photo d).
13
(9) Placing the Catheter (continue)
e
f
• While holding the edge of the catheter, pull the introducer’s T-handle all the
way off the skin (see photo e).
• Peel off the introducer by pulling out the handles; each handle to other
direction (see photo f).
Warning
Care should be taken during catheter placement to assure that occlusion will not occur during use and
that catheter removal will not be impeded.
Ensure that the catheter tip has not been placed in a vein or artery!
14
(10) Fix the Catheter
Filter
g
h
End of infusion
tubing
• Make certain the catheter’s infusion tubing is entirely positioned at the
wound site. The end of the infusion tubing is marked by a black band that
should be positioned just below the skin surface (see photo g).
• Secure the catheter by coiling it close to the insertion site and attach it to the
patient skin using the wound dressing (see photo h).
Caution
15
Do not cover the filter of SmartInfuser Set™
(11) Organizing the System
Place the pump in the Carry Pouch: verify the regulator laying
loosely within the the body of the pouch to avoid any kinks.
16
(12) Using the Bolus
17
(a)
(b)
Place the bolus over the wrist
and secure it by tighten the
strip.
Push the bolus button for an
extra dose of medication when
suffering exceptional pain.
THANK YOU!
18