Pacemaker Anatomy
Download
Report
Transcript Pacemaker Anatomy
Pacemakers and Internal Cardiac
Defibrillators
Mark Wahba
Resident Rounds
September 11, 2003
See:
Brady’s, Blocks, & Pacers
Moritz Haager
1 hr rounds
July 17, 2002
Brief History
First described in 1952
Introduced into clinical practice in 1960
First endocardial defibrillators in 1980
1991 in USA 1 million people had
permanent pacemakers
Outline
Indications
Basics, Pacemaker Components and Code
Complications of Implantation
Pacemaker Malfunction
Management
Disposition
ICD
Guest
Basically:
Device that provides electrical stimulation to
cause cardiac contraction when intrinsic
cardiac electrical activity is slow or absent
http://www.emedicine.com/emerg/topic805.htm
Pacemaker Functions
1.
2.
3.
4.
Stimulate cardiac depolarization
Sense intrinsic cardiac function
Respond to increased metabolic
demand by providing rate responsive
pacing
Provide diagnostic information stored
by the pacemaker
Indications for Pacer
30 AVB and any of:
–
–
–
–
–
Symptomatic bradycardia
Asystole >3 sec or vent escape <40bpm
Post-AVN ablation
Post-op and not expected to improve
Neuromuscular disease
20 AVB + symptomatic bradycardia
Chronic bi-/trifasicular block w/ intermittent 30 AVB or 20 AVB
Type II
Post-MI and any of:
– Persistent 20 AVB or 30 AVB
– Transient 20 AVB or 30 AVB and BBB
SAN dysfunction + symptomatic brady’s (e.g. SSS)
Recurrent syncope due to carotid sinus stimulation
Pacemaker Components and
Anatomy
Pacemaker Components
Pulse Generator
Electronic
Circuitry
Lead System
Pulse Generator
Subcutaneous or submuscular
Lithium battery
4-10 years lifespan
long life and gradual decrease in power
sudden pulse generator failure is an
unlikely cause of pacemaker
malfunction
Electronic Circuitry
Sensing circuit
Timing circuit
Output circuit
Lead System
Bipolar
Lead has both negative,
(Cathode) distal and
positive, (Anode)
proximal electrodes
Separated by 1 cm
Larger diameter: more
prone to fracture
Compatible with ICD
Unipolar
Negative (Cathode)
electrode in contact
with heart
Positive (Anode)
electrode: metal
casing of pulse
generator
Prone to oversensing
Not compatible with
ICD
Difference on an ECG? Bipolar
current travels only
a short distance
between electrodes
small pacing spike:
<5mm
+
Anode
Cathode
Difference on an ECG? Unipolar
current travels a
longer distance
between electrodes
larger pacing spike: +
>20mm
Anode
Cathode
Pacemaker Code
I
Chamber
Paced
II
Chamber
Sensed
III
Response
to Sensing
IV
Programmable
Functions/Rate
Modulation
V: Ventricle
V: Ventricle
T: Triggered P: Simple
programmable
A: Atrium
A: Atrium
I: Inhibited
M: Multiprogrammable
D: Dual (A+V) D: Dual (A+V) D: Dual (T+I) C: Communicating
O: None
O: None
S: Single
S: Single
(A or V)
(A or V)
O: None
V
Antitachy
Function(s)
P: Pace
S: Shock
D: Dual (P+S)
R: Rate modulating O: None
O: None
Common Pacemakers
VVI
– Ventricular Pacing : Ventricular sensing; intrinsic
QRS Inhibits pacer discharge
VVIR
– As above + has biosensor to provide Rateresponsiveness
DDD
– Paces + Senses both atrium + ventricle, intrinsic
cardiac activity inhibits pacer d/c, no activity:
trigger d/c
DDDR
– As above but adds rate responsiveness to allow
for exercise
Rate Responsive Pacing
When the need for oxygenated blood increases,
the pacemaker ensures that the heart rate
increases to provide additional cardiac output
Adjusting Heart Rate to Activity
Normal Heart Rate
Rate Responsive Pacing
Fixed-Rate Pacing
Daily Activities
Determining type of pacemaker
Wallet card: 5
letter code
CXR: code visible
Single lead in
ventricle: VVI
Separate leads
DDD or DVI
Single Chamber
VVI - lead lies in
right ventricle
Independent of atrial
activity
Use in AV
conduction disease
Paced Rhythm Recognition
VVI / 60
Dual Chamber
Typically in pts with
nonfibrillating atria and
intact AV conduction
Native P, paced P,
native QRS, paced
QRS
ECG may be
interpreted as
malfunction when none
is present
May have fusion beats
Four “Faces” of Dual Chamber
Pacing
Atrial Pace, Ventricular Pace (AP/VP)
AV
AP
V-A
VP
Rate = 60 bpm / 1000 ms
A-A = 1000 ms
AV
AP
VP
V-A
Four “Faces” of Dual Chamber
Pacing
Atrial Pace, Ventricular Sense (AP/VS)
AV
AP
V-A
VS
Rate = 60 ppm / 1000 ms
A-A = 1000 ms
AV
AP
VS
V-A
Four “Faces” of Dual Chamber
Pacing
Atrial Sense, Ventricular Pace (AS/ VP)
AV
AS
V-A
VP
V-A
AV
AS
Rate (sinus driven) = 70 bpm / 857 ms
A-A = 857 ms
VP
Four “Faces” of Dual Chamber
Pacing
Atrial Sense, Ventricular Sense (AS/VS)
AV
AS
V-A
VS
Rate (sinus driven) = 70 bpm / 857 ms
Spontaneous conduction at 150 ms
A-A = 857 ms
AV
AS
V-A
VS
Pacemaker Interventions
Magnet application
– No universal function of magnet
– Model-specific magnet, some activate reed
switch asynchronous pacing at pre-set
rate
Interrogation / Programming
– Model-specific pacemaker programmer
can non-invasively obtain data on function
and reset parameters
Magnet Application
Complications of Pacemaker
Implantation
Complications of Pacemaker
Implantation
Venous access
Infection
Thrombophelbitis
Pacemaker Syndrome
Venous Access
Bleeding
Pneumo / hemothorax
Air embolism
Infection
2% for wound and ‘pocket’ infection
1% for bacteremia with sepsis
S. aureus and S. epidermidis
If bacteremic: start Vancomycin, remove
system, TV pacemaker and IV abx for
4-6 weeks, new system
Thrombophlebitis
Incidence 30-50%
1/3 have complete venous obstruction
b/c of collateralization only 0.5-3.5%
devp symptoms
Swelling, pain, venous engorgement
Heparin, lifetime warfarin
Pacemaker Syndrome
Presents w/ worsening of original Sx postimplant of single chamber pacer
AV asynchrony retrograde VA conduction
atrial contraction against closed MV + TV
jugular venous distention + atrial dilation
sx of CHF and reflex vasodepressor effects
Dx of exclusion
Tx w/ dual chamber pacer
Pacemaker Malfunction
4 broad categories
1.
2.
3.
4.
Failure to Output
Failure to Capture
Inappropriate sensing: under or over
Inappropriate pacemaker rate
Failure to Output
absence of pacemaker spikes despite indication
to pace
dead battery
fracture of pacemaker lead
disconnection of lead from pulse generator
unit
Oversensing
Cross-talk: atrial output sensed by vent lead
No Output
Pacemaker artifacts do not appear on
the ECG; rate is less than the lower rate
Pacing output delivered; no
evidence of pacing spike is seen
Failure to capture
spikes not followed by a stimulus-induced
complex
change in endocardium: ischemia,
infarction, hyperkalemia, class III
antiarrhythmics (amiodarone,
bertylium)
A: failure to capture atria in DDD
Inappropriate sensing:
Undersensing
Pacemaker incorrectly misses an intrinsic
deoplarization paces despite intrinsic
activity
Appearance of pacemaker spikes occurring
earlier than the programmed rate:
“overpacing”
may or may not be followed by paced
complex: depends on timing with respect to
refractory period
AMI, progressive fibrosis, lead displacement,
fracture, poor contact with endocardium
Undersensing
Pacemaker does not “see” the intrinsic
beat, and therefore does not respond
appropriately
Intrinsic beat
not sensed
Scheduled pace
delivered
VVI / 60
Undersensing
An intrinsic depolarization that is
present, yet not seen or sensed by the
pacemaker
P-wave
not sensed
Atrial Undersensing
Inappropriate sensing:
Oversensing
Detection of electrical activity not of cardiac
origin inhibition of pacing activity
“underpacing”
pectoralis major: myopotentials
oversensed
Electrocautery
MRI: alters pacemaker circuitry and results
in fixed-rate or asynchronous pacing
Cellular phone: pacemaker inhibition,
asynchronous pacing
Oversensing
Marker channel
shows intrinsic
activity...
...though no
activity is present
VVI / 60
An electrical signal other than the
intended P or R wave is detected
Inappropriate Pacemaker Rate
Rare reentrant tachycardia seen w/ dual
chamber pacers
Premature atrial or vent contraction
sensed by atrial lead triggers vent
contraction retrograde VA conduction
sensed by atrial lead triggers vent
contraction etc etc etc
Tx: Magnet application: fixed rate, terminates
tachyarrthymia,
reprogram to decrease atrial sensing
Causes of Pacemaker
Malfunction
Circuitry or power source of pulse
generator
Pacemaker leads
Interface between pacing electrode
and myocardium
Environmental factors interfering with
normal function
Pulse Generator
Loose connections
– Similar to lead fracture
– Intermittent failure to sense or pace
Migration
– Dissects along pectoral fascial plane
– Failure to pace
Twiddlers syndrome
– Manipulation lead dislodgement
Twiddler’s Syndrome
Twiddler’s Syndrome
Leads
Dislodgement or fracture (anytime)
– Incidence 2-3%
– Failure to sense or pace
– Dx w/ CXR, lead impedance
Insulation breaks
– Current leaks failure to capture
– Dx w/ measuring lead impedance (low)
Cardiac Perforation
Early or late
Usually well tolerated
– Asymptomatic inc’d pacing threshold,
hiccups
– Dx: P/E (hiccups, pericardial friction rub),
CXR, Echo
Environmental Factors
Interfering with Sensing
• MRI
• Electrocautery
• Arc welding
• Lithotripsy
• Cell phones
• Microwaves
• Mypotentials from muscle
Management
Management: History
Most complications and malfunctions
occur within first few weeks or months
pacemaker identification card
Syncope, near syncope, orthostatic
dizziness, lightheaded, dyspnea,
palpitations
Pacemaker syndrome: diagnosis of
exclusion
Management: Physical Exam
Fever: think pacemaker infection
Cannon “a” waves: AV asynchrony
Bibasilar crackles if CHF
Pericardial friction rub if perforation of
RV
Management: adjuncts
CXR: determine tip position
ECG
Potential Problems Identifiable on an
ECG Can Generally Be Assigned to
Five Categories:
Failure to output
Failure to capture
Undersensing
Oversensing
Pseudomalfunction
Pseudomalfunction: Hysteresis
Allows a lower rate between sensed
events to occur; paced rate is higher
Lower Rate 70 ppm
Hysteresis Rate 50 ppm
Management: ACLS
Drug and Defibrillate as per ACLS
guidelines
However keep paddles >10cm from
pulse generator
May transcutaneously pace
Transvenous pacing may be inhibited
by venous thrombosis: may need
flouroscopic guidance
AMI + Pacers
Difficult Dx; most sensitive indicator is
ST-T wave changes on serial ECG
If clinical presentation strongly
suggestive then should treat as AMI
Coarse VF may inhibit pacer
(oversensing)
Successful resuscitation may lead to
failure to capture (catecholamines,
ischemia)
Disposition
Disposition
Admit
– Pacemaker infections /unexplained fever or WBC
– Myocardial perforation
– Lead # or dislodgement
– Wound dehiscence / extrusion or erosion
– Failure to pace, sense, or capture
– Ipsilateral venous thrombosis
– Unexplained syncope
– Twiddlers syndrome
Disposition
Potentially fixable in ED w/ help
– Pacemaker syndrome
– Pacemaker-mediated tachycardia
– Cross-talk
– Oversensing
– Diaphragmatic pacing
– Myopotential inhibitors
Internal Cardiac Defibrillators
Internal Cardiac Defibrillators
Device to treat tachydysrhythmias
If ICD senses a vent rate > programmed cutoff rate of the ICD device performs
cardioversion/defibrillation
All ICDs are also vent pacemakers
Required shock is approximately <15 Joules
Similar problems with implantation as
pacemakers
Indications for ICD
Cardiac arrest from VF or VT not due to
reversible etiology
Spontaneous sustained VT
Syncope NYD + inducible symptomatic VF or
VT in setting of poor drug tolerance or
efficacy
Non-sustained VF or VT + CAD, prior MI, LV
dysfunction and inducible VF or VT not
responding to Class I antiarrhythmic Tx
ICD Malfunction
Inappropriate Cardioversion
Ineffective Cardioversion
Failure to Deliver Cardioversion
Inappropriate Cardioversion
Most frequently associated problem
Sensing malfunction: SVT sensed as VT
Shocks for nonsustained VT
T waves detected as QRS complex and
interpreted as HR
h/r Could be incidence of VT, VF (hypoK,
hypoMg, ischemia +/- infarction)
Ineffective Cardioversion
Inadequate energy output
Rise in defibrillation threshold
antiarrhythmics
MI at lead site
Lead fracture
Dislodgement of leads
Failure to Deliver Cardioversion
Failure to sense
Lead fracture
Electromagnetic interference
Inadvertent deactivation
ACLS Interventions
ICD may not prevent sudden cardiac
death
Same approach as with pacemakers
Person performing CPR may feel a mild
shock if ICD discharges during
compressions
Can deactivate device with magnet
during resuscitation efforts
Disposition
“in almost all instances, admission to a
monitored setting with extended
telemetric observation will be
necessary”
Rosen’s
Thanks to:
Calgary Health Region Pacemaker nurses
Karen and Sandra
References
Brady et al. 1998. EM Clinics NA. 16(2): 361-388
Xie et al. 1998. Em Clinics NA. 16(2): 419-462
Shah et al. 1998 EM Clinics NA. 16(2): 463-487
Harrigan and Brady. 2000. EMR 21(19): 205-216
Rosen
American College of Cardiology ECG of the Month Feb 2001:
http://www.acc.org/education/online/ecg_month/0201/Feb01_02.htm
Pacemaker and Automatic Internal Cardiac Defibrillator, Weinberger et.
al http://www.emedicine.com/emerg/topic805.htm
CorePace presentation 99912 by Medtronic Inc. 2000 available from
Pacmaker Nurses at Foothills Hospital, www.medtronic.com
Pacemaker and Defibrillator Clinics
-Open Mon.-Fri. 0800-1600 hrs.
-Electrophysiology for ICD’s 41248
-Pacemaker Clinic 41188
-On call pager #0569
ECG analysis
Building on routine ECG
interpretation skills
Low Rate
– Pacemakers not programmed below 50
BPM
– ICD’s often low rate of 40 BPM
Assess atrial rhythm
P waves and rate
Atrial sensing
Any atrial pacing
Atrial capture
Assess Ventricular Rhythm
QRS rate and morphology
Relation to atrial rhythm
Ventricular pacing
Ventricular capture
Upper Rate
Pacemakers do not prevent intrinsic heart rate from going too
fast
Pacing therapy and Drug therapy is required to suppress rapid
rhythms like atrial fibrillation
Mode switching devices recognize fast atrial rhythms and
automatically switch to a non-tracking mode
Some pacemakers Medtronic AT501 have anti-tachycardia
therapies to treat Atrial Flutter but they cannot terminate atrial
fibrillation
Defibrillators
Most often ICD patient are not paced
VT/VF detection and treatment with
pacing or shock is their primary purpose
Event memory to analyze rhythm
detected and treated.
PM and ICD Pts in ER
-Direction of Medical Director is that
Cardiology should be consulted.
-Cardiologist to initiate calling in On-Call
Pacemaker Clinic nurse to assess
device function and diagnostics.
Complete assessment in ER
Assess symptoms
ECG Look for Pacing and sensing, Atrial/Ventricular
Obtain patients device info from card or old chart if
possible
Medications and compliance
Surgical Complications
Incision issues: Infected pacemaker site
presents risk for endocarditis
Needs to be brought to Pacemaker or ICD
clinic attention
Cardiac perforation/Tamponade
– Early Post implant
– New or unusual symptoms of sharp,
stabbing chest pain.
– Worse with deep breath
– Usual cause is atrial perforation or tear
–
PACE, Vol. 25, No.5 Post Pacemaker Implant Pericarditis: Incidence and Outcomes
with Active Fixation Leads. Soori Kivakumaran, M. E. Irwin, S. S. Gulamhusein
Management in ER
Echocardiogram
– Look for blood in pericardium
– CV surgery consult
– Don’t anticoagulate
Less than appropriate reasons we
are called..
Don’t need to bother Cardiologist
Will call Cardiology after device assessed
Pt in ER with angina, not appropriate to
assess device at that time
We do follow patients with devices on a
routine basis so they don’t need to be
checked just because they have a device.
Magnets and devices
No universal response to magnet
application with cardiac devices.
Pacemaker Clinic
FMC Monday to Friday 0800-1600
Phone 944-1188
EP Clinic
ICD patient issues
Monday to Friday 0800-1600 hrs
Phone 944-1248
Nurse on Call
CHR pager
Please consult Cardiology
Medical Director: Dr. A. M. Gillis
– Electrophysiology