Welcome to the Laguna Laser Center

Download Report

Transcript Welcome to the Laguna Laser Center

Outcomes Update for an
Aspheric Accommodating IOL
John A. Hovanesian, M.D.
Clinical Instructor, UCLA Jules Stein Eye Institute
Harvard Eye Associates, Laguna Hills, California
Consultant to:
Allergan (1, 3)
Abbott Medical Optics (1)
Baxter (1)
Bausch & Lomb (1, 2)
Diopter, Inc. (1)
Essex Woodlands (1)
Glaukos (1, 2)
IOP (1, 2)
Inspire (1)
Ista (1, 2, 3)
Ivantis (1)
1 = consulting fees, 2 = research support, 3 = equity interest
Ocular Therapeutix (1, 2, 3)
Kurtis Eyewear, (1)
ReVision Optics (1)
Sirion Therapeutics (1)
Vistakon (1)
Vista Research (1)
Crystalens: How it Works
Primary Force: Ciliary Muscle
• Accommodation causes the ciliary muscles to
contract, displacing the vitreous mass &
forcing the lens forward
Hovanesian,
2008
Hovanesian 2011
Crystalens: How it Works
Secondary Force: Optic Arching
• Capsular contraction slightly distorts the optic
as it is forced anteriorly, changing the radius
of the lens
Distance
Near
Δ 2.50 D
Hovanesian,
2008
Hovanesian 2011
Crystalens AO
Accommodating IOL
• Aberration-free optic design
• Maintains depth of focus
of spherical IOL
• Improves contrast
sensitivity over
spherical IOLs
• Aspheric surface on anterior
and posterior surfaces
• Less sensitive to decentration
Hovanesian,
2008
Hovanesian 2011
Effect of SA on Image Quality
• Spherical aberration degrades retinal image quality.
• Reduces contrast sensitivity and low-contrast visual
acuity
• High-contrast visual acuity usually not affected
• Creates halos
SA = +0.45 mm
SA = 0
20/40 Letter E
Hovanesian,
2008
Hovanesian 2011
Crystalens AO Outcomes
Data Collection
• Uncorrected distance (UCDVA) and near
(UCNVA) visual acuities entered into
SurgiVision® DataLink IOL Registry
• Data from:
• Our site: all eyes with data entered
• Global data: all Crystalens AO eyes entered
into database
• Data from 1 and 3 month postoperative visits
Hovanesian,
2008
Hovanesian 2011
Mean Postoperative
Refraction
Our Results
Global Results*
Postop
Visit
N
Mean MRSE (D) ± SD
N
Mean MRSE (D) ± SD
1 Month
39
-0.36 ± 0.47
2779
-0.27 ± 0.59
3 Months
17
-0.32 ± 0.47
1182
-0.29 ± 0.61
*Global Results from SurgiVision® DataLink IOL Registry
Hovanesian,
2008
Hovanesian 2011
Our Results: Monocular
UCDVA
Hovanesian,
2008
Hovanesian 2011
Our Results: Monocular
UCNVA
Hovanesian,
2008
Hovanesian 2011
One Month Results Summary
Our Results
Global Results*
Postop
Visit
N
%20/25
or better
%20/40 or
better
N
%20/25 or
better
%20/40 or
better
UCDVA
44
57%
91%
3228
52%
85%
UCIVA
NA
NA
NA
2866
74%
93%
UCNVA
32
69%
97%
2825
39%
82%
*Global Results from SurgiVision® DataLink IOL Registry
Hovanesian,
2008
Hovanesian 2011
Conclusion
• Crystalens AO provides acceptable distance
vision and functional near vision.
• Predictable outcomes: Mean MRSE within
0.5 D of plano in both groups
Hovanesian,
2008
Hovanesian 2011
Thank you!
John A. Hovanesian, M.D.
Laguna Beach, California
[email protected]
Hovanesian, 2008