feasibility of A wide-field instrument for the NRO telescopes
Download
Report
Transcript feasibility of A wide-field instrument for the NRO telescopes
Erin Elliott, Sr. Astronomical Optics Scientist
John MacKenty, WFC3 Team Lead
Space Telescope Science Institute
1
Optical Design by Erin Elliott
Instrument shown here is a proof-of-concept
design only!
Design issues that remain are engineering
challenges, NOT fundamental limitations.
2
Instrument feasibility
The as-is configuration of the NRO
telescopes can support wide-field
instruments.
Physical geometry doesn’t preclude
fields of larger than ~2 degrees.
Adequate space for instrument packages
exists behind the primary mirror support
structure. (~ 2.4 m in dia x 1+ m).
Could extend downwards further.
On-axis wavefront performance could
potentially support an on-axis instrument
– 1.6 arc min FOV without tertiary mirror.
Wavefront error of telescope system is
reported as < 60 nm RMS. Will limit
performance at wavelengths < 600nm.
volume
potentially
available for
instruments
Layout of a telescope similar to the
delivered units.
3
Proof-of-concept wide-field imager design
Instrument consists of
two folds, two
powered mirrors, and
a spherical corrector
plate.
Covers a field of view
of 0.375 square
degrees.
Optics occupy a
volume of 1.9 x 1.0 x
0.65 m (1.24 m^3).
Possible to include
two such instruments
(Note: Primary ID and OD are not to scale in this plot.)
4
Layout details
Order of reflections: fold 1, fold
2, tertiary, pupil & spherical
corrector plate, quaternary,
image.
Provides an accessible pupil for
filters. (Currently 5.5 inches in
diameter.)
Image plane configured in three
squares, for good tiling efficiency.
(Each ray bundle shown
represents a different field point.)
5
Additional views
6
Infrared Instrument Considerations
NRO Telescope “use as is”
Design assumes room temperature telescope
Similar to HST situation
Silver mirror coating lower emissivity than HST
IR Detector
WFC3
○ 1.7 mm cutoff at 145°K dark <0.04 e-/s/pix
○ Zodi limited in broad filters (readnoise ~ 12e-)
○ Filters at -30°C
NRO Telescope
○ Zodi limited ~2 mm cutoff 100-120°K detector
○ Filters must be ~ -<50 °C
Current design accommodates
Cold enclosure for filters/corrector plate/cold stop
7
Performance and image plane layout
Instrument performance contours of RMS WFE (assuming 0 WFE for OTA).
Total field of 0.375 degrees square.
27 4k x 4k arrays with 0.11 arcsec pixels.
10 micron pixels (Hawaii-4RG10) – larger FOV possible with 15 micron pixels
Bottom field region is inaccessible because of the beam clearances required for a
reflective system.
8
Corrector plate has a spherical aspheric
term.
Plate is thin and unpowered, so doesn’t
introduce significant chromatic aberration.
Mirror sizes:
Fold 1, rectangular, 0.37 x 0.39 m
Fold 2, three-square config., ~ 0.38 x 025
Tertiary, rectangular, 0.58 x 0.38 m.
Pupil & corrector plate, circular, ~ 5.5 inches
Quaternary, rectangular, 0.4 x 0.26 m.
Image plane, three-sqr. config., 0.36 x 0.2 m
Footprint plots (not to scale), showing beam position on the
mirrors, for 12 field points at the corners of the image plane.
fold 1
fold 2
Tertiary and quaternary are conics with a
coma aspheric term. Both are convex.
tertiary
quaternary
Mirror details
image
9
A 6 x 3 Pointing Mosaic with ErinCAM
Conclusion
The as-is configuration of the NRO telescopes can
support wide-field instruments with good image
quality.
The proof-of-concept design presented demonstrates
a FOV of over 0.375 square degrees in a single
instrument.
Thermal requirements for cooling of detectors and
optical elements and thermal stability of telescope
require careful trade for long wavelength cutoff.
11