FRCR 2011 - 12 Lecture 2-
Download
Report
Transcript FRCR 2011 - 12 Lecture 2-
Nuclear Medicine:
Planar Imaging and the
Gamma Camera
Katrina Cockburn
Nuclear Medicine Physicist
Methods of Analysis
Once tracer has traced – need some
method of analysing distribution
Imaging
Gamma Camera,
PET Camera
Compartmental
Analysis
Sample Counter
Radiation Detectors
Converts incident photon into electronic
signal
Most commonly used detectors are
scintillation
Photon interacts with crystal to convert
incident photon into light photons
PMT changes light into electrical signal
Electrical signal recorded and analysed
Imaging Equipment
The Gamma Camera
Basic principle hasn’t
changed since 1956!
Scintillation Imaging
Administration of Isotope
Scintillation Imaging
Localisation and Uptake
Scintillation Imaging
Localisation and Uptake
Scintillation Imaging
Localisation and Uptake
Scintillation Imaging
Localisation and Uptake
Scintillation Imaging
Localisation and Uptake
Scintillation Imaging
Localisation and Uptake
Scintillation Imaging
Localisation and Uptake
Enhanced contrast
between Organ of
Interest and rest of
body
Scintillation Imaging
Imaging distribution
Gamma-rays emitted by
radiopharmaceutical
Collimator ‘selects’ only
those rays travelling at
right angles to face of
camera
Scintillation events in
crystal recorded
Early Scintillation Study
Components of a Modern Gamma
Camera
The components of a modern gamma
camera
Lead Shield
Electronics
PMTs
Lightguide
Crystal
Collimator
The Collimator
The collimator consists
of:
a lead plate
array of holes
It selects the direction of
the photons incident on
the crystal
It defines the geometrical field of view of the
camera
The Collimator
Detector
Detector
Patient
Patient
In the absence of collimation:
no positional relationship between source –
destination
In the presence of collimation:
all γ-rays are excluded except for those travelling
parallel to the holes axis – true image formation
Types of Collimators
Several types
of collimator:
Parallel-Hole
Converging
Diverging
Pin-Hole
Energy Ranges of Collimators
Type of
Collimator
Energy
Range
Low Energy
(LE)
0 - 200 keV
Medium
Energy (ME)
High Energy
(HE)
Typical
Nuclide
Tc-99m
Tl-201
200 - 300 keV In-111
300 – 400 keV I-131
The Scintillation Crystal
First step of image
formation
Photon detected by its
interaction in the
crystal
γ-rays converted into
scintillations
Scintillation
Can be thought of as
“partial ionisation”
Electrons excited and
gain energy
As electrons fall
back to ground
state, photons
emitted
Use of doping (eg
NaI:Tl) creates
smaller gaps
Scintillation Crystal Properties
High stopping efficiency
Stopping should be without scatter
High conversion of γ-ray energy into visible light
Wavelength of light should match response of
PMTs
Crystal should be transparent to emitted light
Crystal should be mechanically robust
Thickness of scintillator should be short
Properties of NaI(Tl) Scintillator
The crystal – NaI(Tl)
emits light at 415 nm
high attenuation
coefficient
intrinsic efficiency:
90% at 140 keV
conversion efficiency:
10-15%
energy resolution:
15-20 keV at 150 keV
Disadvantages of NaI(Tl) crystal
NaI(Tl) crystal suffers from the
following drawbacks:
Expensive (~£50,000 +)
Fragile
sensitive against mechanical stresses
sensitive against temperature changes
Hygroscopic
encapsulated in aluminium case
Lightguide and Optical Coupling
Lightguide acts as optical coupler
Quartz doped plexiglass (transparent
plastic)
The lightguide should:
be as thin as possible
match the refractive index of the scintillation crystal
Silicone grease to couple lightguide, crystal
and PMT
No air bubbles trapped in the grease
The Photomultiplier Tube
A PMT is an evacuated
glass envelope
It consists of:
a photocathode
an anode
~ 10 dynodes
The Photomultiplier Tube
Photocathode of PMT emits 1
photoelectron per ~ 5 – 10 photons
Photoelectron accelerated towards first
dynode
Dynode emits 3 – 4 secondary e- per
photoelectron
Secondary e- accelerated towards next
dynode
Multiplication factor ~ 106
Output of each PMT proportional to the
number of light photons
PMT Properties
The photocathode
should
be matched to blue light
have high quantum
efficiency
High stability voltage
supply: ~1kV
Positional and Energy Co-ordinates
PMT signals processed
spatial information – X and Y signals
energy information – Z signal
Z signal – the sum of the outputs of all PMTs
proportional to the total light output of the crystal
Light output proportional to the energy of incident
gamma
Pulse height analyser accepts or rejects the
pulse
Pulse Height Analysis
Z-signal goes to PHA
PHA checks the energy of the γ-ray
If Z-signal acceptable
γ-ray is detected
position
determined by
X and Y signals
20% window still
includes 30% of
scattered photons
Determining the Position of Events
Image Acquisition Techniques
Static
Dynamic
Gated
Tomography
-
(Bones, Lungs)
(Renography)
(Cardiac)
-
(Cardiac)
SPECT
PET
List Mode
Static Imaging
1
Camera
Computer Memory
Image Display
Camera FOV divided into regular matrix of pixels
Each pixel stores number of gamma rays
detected at corresponding location on detector
Typical Matrix Sizes: 2562, 1282, 642
Static Imaging
1
1
Camera
Computer Memory
Image Display
Camera FOV divided into regular matrix of pixels
Each pixel stores number of gamma rays
detected at corresponding location on detector
Typical Matrix Sizes: 2562, 1282, 642
Static Imaging
1
1
1
Camera
Computer Memory
Image Display
Camera FOV divided into regular matrix of pixels
Each pixel stores number of gamma rays
detected at corresponding location on detector
Typical Matrix Sizes: 2562, 1282, 642
Static Imaging
1
1
1
1
Camera
Computer Memory
Image Display
Camera FOV divided into regular matrix of pixels
Each pixel stores number of gamma rays
detected at corresponding location on detector
Typical Matrix Sizes: 2562, 1282, 642
Static Imaging
2
1
1
1
Camera
Computer Memory
Image Display
Camera FOV divided into regular matrix of pixels
Each pixel stores number of gamma rays
detected at corresponding location on detector
Typical Matrix Sizes: 2562, 1282, 642
Static Imaging
2
1
1
1
Camera
Computer Memory
1
Image Display
Camera FOV divided into regular matrix of pixels
Each pixel stores number of gamma rays
detected at corresponding location on detector
Typical Matrix Sizes: 2562, 1282, 642
Static Imaging
3
1
1
1
Camera
Computer Memory
1
Image Display
Camera FOV divided into regular matrix of pixels
Each pixel stores number of gamma rays
detected at corresponding location on detector
Typical Matrix Sizes: 2562, 1282, 642
Static Imaging
3
1
1
1
Camera
Computer Memory
1
Image Display
Camera FOV divided into regular matrix of pixels
Each pixel stores number of gamma rays
detected at corresponding location on detector
Typical Matrix Sizes: 2562, 1282, 642
Dynamic Imaging
Series of sequential
static frames
E.g. 90 frames each
of 20s duration
Image rapidly
changing distribution
of activity within the
patient
Used in Renography
Dynamic Imaging Analysis
Split Renal
Function
ROIs
Curves
showing
changing
renal activity
over time
Gated Imaging
Several frames
acquired covering the
cardiac cycle
Acquired over many
cycles