Scanner Qualification
Each site must qualify by submitting volunteer MRI exam utilizing the ACRIN
PA 4008 scanning protocol parameters- injection of Gadolinium is not
necessary
The qualifying exam will be reviewed for image quality (e.g. artifacts,
distortion, and signal-to-noise) and protocol compliance
An email will be sent to inform the site team of approval. In the event that
the qualifying exam is not approved, required corrections for requalification
will be included in the e-mail.
Please submit tonometry (.txt) qualification file at the same time for
subsequent transmission to Dr. Chirinos for approval
ACRIN CV Committee
Participant Scheduling
Participants will undergo two 1.5T cardiac MRI studies
for the trial.
MRI SCAN 1: completed within (3) weeks prior to AVR
MRI Scan 2: completed (6) months post AVR
ACRIN CV Committee
Participant Preparation
Standard practice for pre-treating severe claustrophobia or
anxiety
eGFR must be >30 mL/min/1.73m² of body surface area.
Confirm participant has abstained from smoking, food, and
alcohol for a minimum of four (4) hours
Confirm participant has abstained from caffeine-containing
beverages for a minimum of 24 hours
If a short-acting nitrate was taken within four (4) hours prior to
testing, the MRI scan will need to be rescheduled
ACRIN CV Committee
Imaging Schema
ACRIN CV Committee
Imaging Forms
The following imaging forms should be completed in
their entirety by the designated personnel for every
submitted exam:
1. Image Transmittal Worksheet (ITW)
2. MRI Assessment Form (MR)
It is recommended that each of the above forms be
made available, in hard copy format, for the Imaging
Technologists to be completed during image
acquisition when the requested information on
these forms is most readily available
ACRIN CV Committee
Image Submission
Site should submit all ACRIN PA 4008 imaging to the ACR
Imaging Core Lab in DICOM format via TRIAD.
All imaging should be submitted within 48 hours after
acquisition
An Image Transmittal Worksheet (ITW) MUST be
completed upon image submission
ACRIN CV Committee
TRIAD
TRIAD (Transfer of Images and Data) is American College of
Radiology Imaging Network’s image submission and
management software
TRIAD-OA offers a web-based software solution which allows
participating imaging sites a secure method to submit clinical
trial images
TRIAD is remotely installed on a designated PC at your site
TRIAD software anonymizes, encrypts, and performs a lossless
compression of the images before they are transferred to the
ACRIN image archive
Frequently asked questions (FAQs) regarding TRIAD software
can be found at https://triad.acr.org/faqs.htm
ACRIN CV Committee
ACR Core Laboratory QC
Upon image receipt an ACRIN Imaging Technologist
will:
• Check for missing images/sequences
• Appropriate image anonymization
• Complete anatomical coverage of the heart
• Adherence of all sequences to imaging protocol
• Absence of image artifact
In cases where image sets are judged to be
suboptimal the trial PI will be informed, and a
replacement participant should be accrued
ACRIN CV Committee
Image Data Queries
Any submitted exam which has missing data or does not
follow the ACRIN PA 4008 protocol guidelines a Protocol
Deviation form will be generated
Sites are expected to resolve data queries expeditiously
Queries not resolved within 7 business days will be sent
to the ACRIN 6701 trial team for additional follow-up
ACRIN CV Committee
Questions?