IHE Mammography Image Integration Profile
Download
Report
Transcript IHE Mammography Image Integration Profile
Integrating the Healthcare Enterprise
Mammography Image – MAMMO
Chris Lindop, GE Healthcare
Co-Chair Radiology Planning Committee
Feb 7-8, 2007
1
IHE Participant's Workshop
A content profile for Mammo
Patient Info. Reconciliation
Import Reconciliation Workflow
Charge
Posting
Scheduled Workflow
-
Teaching
Files &
Clinical
Export
Key
Image
Notes
Presentation
of Grouped
Procedures
NM
Image
Mammo
Image
Consist.
Present.
of
Images
PostProcessing
Workflow
Evidence
Docs
Reporting
Workflow
Key
Image Image
Fus- Notes
ion
Simp.
Image
& Num
Reports
Access to Radiology Information
Portable Data for Imaging
Cross-Enterprise Document Sharing for Imaging
Feb 7-8, 2007
2
Mammo Image – The Problem
Mammography image review relies on comparison with priors
increases the occurrence of simultaneous review of images from
different systems
Multi-vendor Mammography Systems
Vary with detector technologies and sizes result in varying image
matrices sizes.
Vary with contrast presentation methods result in varying image
presentation
Mammography Image Review requires a minimal set of display
features
zoom, LUT rendering, hanging protocols
Feb 7-8, 2007
3
Example of different image sizes
Image Contrast Adjustment
VOI LUT
Linear LUT
MAMMO – Value proposition
Minimum contents for the generated DICOM instances
Image Data (for processing and for presentation SOP Classes)
CAD data (CAD SR)
Basic display capabilities of Image displays to suit specific
mammography review requirements
CAD acting as Evidence Creators
Mammography Workflow
Managing the process of creating, storing and using Mammography
Image content is similar to workflow for other image content (e.g.
see Scheduled Workflow and Post-Processing Workflow Profiles).
Feb 7-8, 2007
6
Mammography Image
Actors and Transactions
Acquisition
Modality
Evidence
Creator
[RAD 8] Modality Images Stored ↓
[RAD 10] Storage Commitment ↓
↓ [RAD 18] Evidence Document Stored
↓ [RAD 10] Storage Commitment
Image
Manager
Image
Archive
↓ [RAD 14] Query Images
↓ [RAD 16] Retrieve Images
↓ [RAD 44] Query Evidence Documents
↓ [RAD 45] Retrieve Evidence Documents
Image
Display
Print
Composer
↓[RAD-23] Print Request with Presentation LUT
Print Server
Feb 7-8, 2007
7
Mammography Image
Process Flow
Evidence
Creator
(CAD System)
Acquisition
Modality
Perform
Acquis ition
Image Manager/
Image A rchive
Image D isplay
Modality Images
Stored [8]- Mamm o
Images
Storage
Commi tment [10]
Perform
CAD
Evidence Document
Stored [43]Mamm oCAD
Storage
Commi tment [10]
Query Images [14]
Retrieve Images [16] - Mamm o Images
Present Mammo
Images
to User
Query Evid ence Document [44]
Retrieve Evidence Document [45] Mamm oCAD
Present Mammo
Images with CAD
to User
Feb 7-8, 2007
8
Mammo Image Profile
Acquisition Modality
FFDM & CR are required to create MG objects
Create MG objects as specified
•
•
40 Attributes with specific requirements defined
Including new pixel padding range limit (0x0028,0x0121)
Evidence Creator
Mammo CAD systems
Support Evidence Document Stored with Storage
Commitment
Support Mammography CAD SR objects
Feb 7-8, 2007
9
Mammo Image Profile (cont’d)
Image Manager/Archive
Support both Modality Images and Evidence Document Stored
with Storage Commitment
Support Query/Retrieve of both Images and Evidence
Documents
Image Display
Support Query/Retrieve of both Images and Evidence
Documents
Support specified display capabilities including background air
suppression, LUT rendering, zoom management and CAD SR
display
Feb 7-8, 2007
10
Mammo Image Profile (cont’d)
Print Composer
Support Print Request with Presentation LUT
• including the LIN-OD presentation LUT shape
Use Requested Image Size to achieve true size printing
Support 12 bits printing
Print Server
Support Print Request with Presentation LUT
• including the LIN-OD presentation LUT shape
Use Requested Image Size to achieve true size printing
Support 12 bits printing
Feb 7-8, 2007
11
Mammo Image contents
Background air suppression
New tag in DICOM (CP692) extending the pixel padding
concept
Mammo CR generate MG instances only
specific fields (i.e. gantry ID) have been added to the
required attribute list.
Tiled Views option defined
Allows for large breast imaging (multiple images of the same
breast)
Feb 7-8, 2007
12
Image Display
Zoom
Adjust zoom accounting for varying acquisition pixel sizes
True size zoom accounting for display (monitor) pixel size
Background air suppression
When inverting/windowing the image, the detected background
stays black
Image orientation
Flipping according to Patient Orientation ((0x20,0x20))
Annotations
Set of required annotations to be displayed is defined
CAD rendering (using CAD SR)
Feb 7-8, 2007
13
Feb 7-8, 2007
14