Fibrillation and Defibrillator
Download
Report
Transcript Fibrillation and Defibrillator
Engr. Hinesh Kumar
1899 by Prevost
and Batelli, two Italian physiologists .
They discovered that electric shocks could convert
ventricular fibrillation to sinus rhythm in dogs.
The first case of a human life saved by
defibrillation was reported by Beck in 1947 .
Defibrillation was invented in
Fibrillation
is the rapid, irregular, and
unsynchronized contraction of muscle fibers.
There are two major classes of cardiac fibrillation:
Atrial Fibrillation and Ventricular Fibrillation.
Atrial
fibrillation is an irregular and
uncoordinated contraction of the cardiac muscle of
atria.
Ventricular
Fibrillation is an irregular and
uncoordinated contraction of the cardiac muscle of
ventricles.
Figure:
(a) Normal waveform.
(b) Ventricular fibrillation.
(c) Ventricular tachycardia.
Defibrillation is a process in
which an electronic device sends
an electric shock to the heart to
stop an extremely rapid,
irregular heartbeat, and restore
the normal heart rhythm.
Defibrillation should be
performed with in the first 8
minutes after cardiac arrest.
Ideally, the sooner, the better.
Defibrillator is the device used to
deliver the electrical shock and it can
be manual or automatic.
Internal Defibrillator
Electrodes directly placed to the heart
e.g., ICD (Internal Cardioversion Defibrillator)
External Defibrillator
Electrodes placed directly on the chest.
e.g., AED (Automatic External Defibrillator)
Internal
External
An
implantable
cardioverter-defibrillator
(often called an ICD) is a device that briefly
passes an electric current through the heart.
It is "implanted," or put in your body surgically.
It includes a pulse generator and one or more
leads.
The pulse generator constantly watches your
heartbeat.
AED
is a portable electronic device
that
automatically diagnoses the ventricular fibrillation in
a patient.
Automatic refers to the ability to autonomously
analyse the patient's condition.
AEDs require self-adhesive electrodes instead of hand
held paddles.
There are two general classes of waveforms:
a) Monophasic waveform
•
Energy delivered in one direction through the
patient’s heart
b) Biphasic waveform
•
Energy delivered in both direction through the
patient’s heart
A monophasic type, give a high-energy shock, up to 10 to
360 joules due to which increased cardiac injury and in
burns the chest around the shock pad sites.
A biphasic type, give two sequential lower-energy
shocks of 5 - 200 joules, with each shock moving
in an opposite polarity between the pads.
Low energy biphasic shocks may be as effective as high
energy monophasic shocks.
Biphasic waveform defibrillation used in implantable
cardioverter defibrillator (ICD) and automatic external
defibrillators.
1. Lown Waveform
2. Monopulse Waveform
3. Tapered DC Delay
4. Trapezoidal
In 1962, Dr Bernard Lown of Harward University
introduced the waveform that bears his name.
The voltage and current applied to the patient's chest
plotted against time.
The current will rise very rapidly to about 20 A under
the influence of slightly less than 3 KV.
The waveform then decays back to zero within 5 ms,
and then produces a smaller negative pulse also of
about 5 msec.
The charge delivered to the patient is stored in a
capacitor and is produced by a high-voltage dc power
supply.
Lown Defibrillator Waveform
The operator can set the charge level using the set
energy knob on the front panel.
The knob controls the dc voltage produced by the highvoltage power supply and so can set the maximum
charge on the capacitor
The energy stored in the capacitor is given by:
where,
U is the energy in joules (J)
C is the capacitance of C1, in farads (F)
V is the voltage across C1, (V)
Calculate the energy stored in a 16-µF capacitor
that is charged to a potential of 5000 V dc.
Solution
The monopulse waveform shown in figure is a modified
Lown waveform and is commonly found in certain
portable defibrillator.
This waveform differs from the others in that it uses a
lower amplitude and longer duration to achieve the
energy level.
The energy transferred is proportional to the area
under the square of the curve, so we may attain the
same energy as in other waveforms.
The trapezoidal waveform shown in figure is another
low-voltage, long duration shape.
The initial output potential is about 800 V, which
drops continuously for about 20 ms until it reaches
500 V, where it is terminated .
Types of defibrillator electrodes:a)
Spoon Shaped Electrode
Applied directly to the heart.
b)
Paddle type electrode
Applied against the chest wall
c)
Pad Type Electrode
Applied directly on chest wall
fig: Electrodes used in defibrillator (a) a spoon shaped internal
electrode that is applied directly to the heart. (b) a paddle type
electrode applied against the anterior chest wall.
Fig.- Pad electrode
White is negative, anterior chest wall.
Red is Positive , left anterior axillary line.
“Red on Ribs! White on right!”
• The paddles used in the procedure should not be
placed:• on a woman's breasts
• over an internal pacemaker patients.
• Before the paddle is used, a electrochemical gel must
be applied to the patient's skin to avoid the skin burns.
• Skin burns from the defibrillator paddles are the most
common complication of defibrillation.
• Other risks include injury to the heart muscle,
abnormal heart rhythms, and blood clots.
Attach the external and internal paddles if the monitor
reads, "No paddles”.
Check to ensure that the leads are securely attached if
the monitor reads, "No leads.“
Connect the unit to AC power if the message reads,
"Low battery.“
Verify that the Energy Select control settings are
correct if the defibrillator does not charge.
Change the electrodes and make sure that the
electrodes adapter cable is properly connected.
Close the recorder door and the paper roll if the
monitor message reads, "Check recorder”.