Transcript ppt - Cern

C. Guyot (Saclay)
Dubna 06/2000
BARREL ALIGNMENT PLATFORM POSITIONING
Goal:
• Positioning of the platforms at their nominal position on the
MDT chambers (No position measurement in a database).
Principle:
• Use a precisely adjusted template squares (precision < 10 mm) to
hold the platform at their nominal position during the glue curing
w.r.t. the endplugs (2 points) and the outer tube surface or on the
jig (BOS only) at the level of the first comb.
6 adjustable (on CMM) balls for the platform positioning
Half sphere on end plugs
Half sphere on the
tube outer surface or
on the jig (BOS)
Flat surface on endplugs
Praxial platform positioning tools:
Top view
Bottom view
Positioning in x of the template tools on the tube layer:
Ruler with
straigthness < 100 mm
Tube with calibrated length (+- 50 mm)
Platform screwed on the positioning square
The optical control tools
Goal:
• Check the platform orientation (angles) during the tube layer assembly
( Reminder: The platforms are glued on the first layer)
• Register on a database the values of the angles measured after only one
tube layer glued (i.e. right after the platform gluing) and the angles mesured at the
end of the multilayer assembly. For the module 0 assembly, register the values after each
tube layer assembly (Not possible on BIL due to the stiff back geometry).
Principle: 4 tools (towers) carrying 2 cameras and 2 LEDs
looking at each neighbour
CMOS image sensor
2 leds
lens
f~ 80 mm (except for BIS transverse: 60 mm)
Plateform angle ~ (X/Y(pixel) - X0/Y0) * pixel_size / f
Average led spot coordinate
From the calibration on the granite table
Measured platform angles
Image sensor axis (pixel direction)
Xch
Ych
y14
Tower 1
x12
y12
x14
x23
y23
x21
y21
Signal side
Tower 2
HV side
Tower 4
y43
x43
y41
y34
x41
x32
x34
y32
Tower 3
Zch
Rotation of platform 1 around Xch axis: qX = (X14-X014)/f
Rotation of platform 1 around Ych axis: qY = - (Y12-Y012)/f = - (Y14-Y014)/f
Rotation of platform 1 around Zch axis: qZ = - (X12-X012)/f
Rotation of platform 2 around Xch axis: qX = (X23-X023)/f
Rotation of platform 2 around Ych axis: qY = (Y21-Y021)/f = (Y23-Y023)/f
Rotation of platform 2 around Zch axis: qZ = (X21-X021)/f
Control tool calibration
Longitudinal lines :
Distance adjusted to +- 2mm
Precision wedges
z
Calibrates rotations around z and y axis
x
Control tool calibration
Transverse lines :
Distance adjusted to +- 2mm
Precision wedges
Square ( 90 degrees +- 1 mrad)
x
z
Calibrates rotations around x axis only
•Saclay will supply each MDT lab (barrel) with a set of 4
template squares and a set of 4 control tools (with cables to the
RASMUX).
•Each lab will have to provide:
- A length calibrated tube
- A straight ruler of length ~ chamber width - square width
• The absolute calibration of the control tools will have to done
on a granite table prior to the module 0 assembly.
This calibration requires a precision side on the granite table or
(better) a precision wedge. If the assembly table on side does not
allow such a calibration (no room or no precise side face), this
calibration will be done at Saclay (hoping that the image sensor
does not move during the transport).