Transcript Slide 1

Detection of VOC’s in Spoiling Pork using Field
Asymmetric Ion Mobility Spectroscopy
Meat Transportation
Method
The spoilage of meat during storage and transportation, due to the
growth of bacteria populations, is a problem with obvious
consequences, such as a loss of profits to a meat company due to
subsequent recalls of product, not to mention the possibility of food
borne illness, and loss of consumer confidence.
8 Million trucks a day are used to transport meat across the US and
prevention of meat spoilage is achieved through appropriate
temperature control. However there are reported instances (on a
daily basis) where the meat has arrived at the destination spoiled,
even though passing the incoming core and surface temperature
checks. The cause of this can be attributed to mechanical or
deliberate failure of the refrigeration unit, exposing the meat to
elevated temperatures, which increases the bacteria's growth rate
for the period of time that the refrigeration unit is turned off or
being repaired. This type of failure cannot be detected with current
incoming inspection procedures and unfortunately the methods for
determining the presence of bacteria can take several days to give
result. Therefore a rapid method of detection of bacteria is needed.
With this in mind the feasibility of using the Lonestar system for the
detection of the bacteria metabolites within the headspace of a
meat sample was investigated.
A Lonestar unit was used to monitor the out-gassed volatiles of pork
samples incubated at 25C every hour over a 24 hour period. A
control sample was kept refrigerated over the period and used as a
reference. For each hour adsorbent trap samples were taken and
analyzed via thermal desorption gas chromatography (TD-GC-MS)
for volatile identification, along side bacterial swabs so that bacterial
APC counts could be determined through traditional growth
methods.
The Lonestar Spectra in Figure 2 clearly shows peak height
increases as the bacterial count increases over time. The excel
plot in Figure 3 shows how the peak of pork at 25C increases
over the 24 hour period while the refrigerated pork does not
increase.
Results
Acetic
anhydride
methoxy 2propanol
1
Methanecarbothiolic acid
Tetramer
Acetaldehyde
Lonestar: Field Asymmetric Ion Mobility Spectrometer
Lonestar (Figure 1 below) is a powerful and adaptable chemical
monitor in a portable, self contained unit. Incorporating Owlstone’s
proprietary FAIMS technology, the instrument offers the flexibility to
provide both rapid alerts and detailed sample analysis. It can be
trained to respond to a broad range of chemical scenarios and can
be easily integrated with other sensors and third party systems to
provide a complete monitoring solution. As a result, Lonestar is
suitable for a broad variety of applications ranging from online/at
line process monitoring to lab based R&D.
Sulfurous
acid
T0 – Log 2.75
3 methylbutanal
T4 – Log 2.96
Figure 4 Chromatogram of pork headspace at T4
Subsequently, volatile organic compounds were identified via TD
GC-MS and the resultant chromatogram is shown in Figure 4, along
with the primary chemical compounds identified.
Taking in to consideration the negative mode response, it is thought
that the Lonestar was responding to the sulphur containing
metabolites released from the bacteria that was growing on the
pork. Namely sulphurous acid and the methanecarothiolic acid.
Conclusion
Figure 1 Lonestar
T8 – Log 4.32
Figure 2 Lonestar Spectra
T24 – APC not
done
The Lonestar FAIMS unit detected pork with a bacterial log 2.96,
well before the meat is considered spoiled. Therefore providing
proof of principle for early detection of bacterial APC count.
The method of detection is both simple to use (dock personnel) and
gives a response in under a 2 minutes, which will allow meat
companies to perform quality control measurements for the
presence of bacteria in delivered meat (2000 lb. combos) quickly
and reliably.