Transcript column
HIGH PERFORMANCE
LIQUID
CHROMATOGRAPHY
(HPLC)
When particles of small diameter (microns)
are used as a stationery phase support, the
technique is called HPLC.
H
* Column effeciency is
inversely proportional to
column-packing particle
size.
**Better resolution
**High performance
P LC
High pressure to pump
liquids through an
efficient column.
It is the most widely used form of LC in
assaying many analytes:
-amino acids,
-peptides and proteins,
-carbohydrates,
-lipids,
-nucleic acids,
-vitamins,
-hormones,
-drugs eg. antibiotics, antiepileptics,
antidepressants.
Solvent Reservoir
Solvents used in the mobile phase are
contained in solvent reservoirs.
The reservoirs may be glass bottles or
flasks into which “feed lines” to the
pump are inserted.
Solvent Clarity
The solvent must be free of particulate
matter as impurities can affect
column
pump
Detection system
The Pumping System
It is used to aspirate the mobile phase from the
solvent reservoir and forcing it through the column.
The main features of a good pumping system are:
Capable of producing high pressure
outputs
No pulses (cyclic variations in
pressure)
The dual piston reciprocating
pump
It is a pump that delivers a fixed volume of
solvent (0.01-20ml) onto the column by a
piston driven by a motor.
The piston is moved by a motorized crank
(cam) which is assymetrical, and the entry of
the solvent to the column is regulated by
check valves.
The dual piston reciprocating
pump
How is the HPLC pump operated
Modes of solvent application
Isocratic mode:
#The mobile phase remains constant
throughout the run.
# used for separation of compounds with
similar structures.
# The isocratic mobile phase can be a
single solvent (eg. methanol) or a
mixture of solvents delivered from
a single reservoir.
Gradient mode
# The mobile phase composition
is changed during the run in
either a stepwise or a continuous
fashion.
# The gradient profile can be
generated by different
techniques.
The Injector
An aliquot of the sample (0.5- 50µL) is introduced
into the column via an injector.
The most widely used type is the fixed loop injector.
It consists of a stainless-steel loop of a fixed volume,
that can be filled with the sample.
It has a valve switching system that creates two
positions:
a- a fill or loading position ,and,
b- An inject position.
The Fixed Loop Injector**see animation2
In the fill (loading position), the sample is
introduced with a syringe into the external
loop of the injector.
In the inject mode, the sample loop is rotated
into the flowing stream of the mobile phase.
This results in the flushing of the sample into
the column.
----------------------------------------------N.B. Auto samplers are also available. They are
extremely precise and can be programmed
for automated operation.
Columns
Most analytical HPLC columns in clinical
laboratories are fabricated from stainless
steel.
The following items will be referred to:
* Guard columns.
* Column dimensions.
* Column packing.
After a number of separations, a guard
column is routinely replaced.
Column Dimensions
The internal diameters (ID) of the columns
range from 0.3-5mm.
The length of the columns range from
50-250mm.
Column end fittings (with zero dead
volumes) and frits to retain the support
particles are used to connect the column to
the injector on the inlet and the detector on
the outlet end.
In addition, capillary columns (of 0.1-0.5mm
ID) and 10-50cm in length are available. They
are fabricated through coating the inner wall
of a fused- silica tube with a thin film of a
liquid phase.
Column Packing
Columns are packed with small diameter (3-10µm)
particles made of an inert material. The surfaces of
these particles are covered by the stationery phase.
As packing size is decreased, efficiency and pressure
requirements are increased.
Analytical columns are packed with a
variety of stationery phases, providing
enormous versatility in the separation
processes.
The most popular is the bonded phase packing.
Here the stationery phase is bonded chemically to
the surface of silica particles through silica ester
linkage.
The columns packed in this way can act by various
separation mechanisms, depending on the type of
the stationery phase bonded to the particles.
The C18 type stationery phase in reversed phase partition
chromatography is commercially called octadecyl silane
(ODS).
In normal phase partition chromatography the mobile phase is
relatively non-polar eg.hexane.
In reversed phase partition chromatography the mobile phase
is relatively polar eg.water\methanol.
N.B. Sephadex, sepharose and other non-rigid gels are of limited
use in HPLC as they cannot withstand very high pressures.
Detectors
Many detectors have been developed in HPLC.
The most common are:
Photometric (Visible\UV) detectors.
Fluorometric detectors.
Electrochemical detectors.
Refractory index detectors.
# A key component in these detectors is the flow cell
through which the elute from the chromatographic
column passes.
Spectrophotometric (Visible\UV)
detectors
These measure the radiant energy absorbed by
analytes as they elute from the chromatographic
column.
Most organic compounds absorb in the UV region
(190-400nm) of the spectrum. Some compounds
absorb in the visible region of the spectrum (400700).
The magnitude of light absorption and the
wavelength at which absorption of light occurs
are a function of the molecular structure and
the concentration of the eluted compounds.
Fluorometric Detectors
These detect analytes that emit light (fluorescent
compounds).
Fluorescence occurs when a molecule absorbs light at one
wavelength and becomes excited; then when it returns to its
non- excited (ground) state, it emits light of another
wavelength specific for that analyte.
The analyte detected by fluorometers must have a fluorescent
character or treated so that it becomes fluorescent (i.e.
derivatized).
Fluorometric Detectors
Derivatization can be done before the
sample enters the column OR to the eluent
eluted from the column.
Thus either a pre-column or post- column
reactor, chemically tags a compound for
subsequent detection.
Example: tagging amino acids with
fluorescamine tag.
Fluorometric Detectors
Light sources in fluorometers:
Deuterium lamps
Xenon arc lamps
Laser lamps
Electrochemical detectors
Here an electro-active analyte enters the flow cell
where it is oxidized or reduced at an electrode
surface under a constant potential.
It is useful for measuring catecholamines.
Derivatization by putting electrochemically active
tags (eg. bromine) for compounds as unsaturated
fatty acids and prostaglandins can be done.
Computer
System Control.
Data Processing.
Data Processing.
Analyte identification
The retention time, or volume, at which an
unknown solute elutes from the column is
matched to that of a reference (standard
compound).
Analyte Quantification
A calibration (standard) curve must be used for
measuring analyte concentration.
External calibration
Internal calibration
External calibration
Reference (standard) solutions containing known
quantities of same analyte to be measured are
processed in a manner identical to the samples
containing the analyte.
A calibration curve of peak height or (area) versus
concentration is constructed and used to calculate
the concentration of the analyte in the samples.
Internal calibration
Also called internal standardization.
A different compound called internal standard, is added to
each reference solution and each sample.
By plotting the ratio of the peak height or (area) of the
calibrator to the peak height or (area) of internal standard
versus the concentration of the calibrator,
a calibration curve that corrects for systematic losses is
constructed. The curve is then used to compute the
concentration of the analyte in unknown samples.
Sample preparation
Sample concentration/purification:
This is done using extraction cartridges (1-3ml
propylene syringes that contain the stationery
phase).
Sample derivatization: eg.
a- Treating amino acids with ninhydrin to produce a
color for photometric detection.
b- Tagging analytes with bromine for electrochemical
detection.
c- Tagging analytes with fluorescamine tag for
fluorometeric detection.