return electrode
Download
Report
Transcript return electrode
Electrosurgery
Units
ELECTROSURGICAL
UNITS (ESUs)
Jclemens (2009), electrosurgery [photograph]. Retrieved from https://commons.wikimedia.org/wiki/File:Electrosurgery.jpg
Electrosurgery: Overview
•
•
•
•
Electrical arc used in surgery (≈3000C)
Usually 1Mhz, 300W max.
Modes
– Cut
– Coagulation
– Blend
Probes:
– Monopolar
– Bipolar
ESU Summary
Modern ESUs have great flexibility, control, and feedback
measurements to control cutting and stop bleeding (coag).
Additional functions include fulguration and dessication
BMET Tests include power measurements by simple devices such
as EWH tester, or sophisticated tests by ESU Analyzers
All testing is on fixed loads that don’t resemble actively changing
tissue resistance, so surgeons may still complain
Failure modes are most likely “traumatized” parts such as power
cords, leads, and output power stages (or power supplies)
General Surgery
The electrosurgical unit (ESU) is generally used
in surgery.
The laser is less efficient, less powerful, more
costly, more bulky in the operating room
The skill of the surgeon is a major factor in the
selection of any surgical “knife.” (metal, ESU, or
laser)
ELECTROSURGICAL UNITS
•
The Electro Surgical Unit (ESU) cuts tissue by heat,
created from radio frequency (RF) currents in the
range of 100 kilohertz to 5 megahertz.
•
Surgical Functions:
•
•
•
•
•
Incisions (cutting deep into tissue)
Excisions (removing surface growths)
Coagulation (stopping blood flow)
Desiccation (drying tissues)
Fulguration (charring tissues)
Efficient, Powerful, Economical
•
ESU is the most efficient, powerful, and economical of
the thermal knives presently available.
•
It is most widely used in general surgery and in
cutaneous surgery.
•
It is capable of fast cutting through massive tissue and
of effective hemostasis (stability).
•
Adverse side effect is thermal tissue damage.
Bovie Operation 1
Transformer steps up the line voltage, which is then applied
across a spark-gap gas tube.
High-voltage AC peaks ionize gas in the spark-gap, making low
resistance and high current.
This drops the voltage and extinguishes the gas, which now
has high resistance again, so cycle starts over.
This “limit cycle” repeats at RF frequencies, where oscillation
frequency is set by series capacitor and primary coil.
Bovie Operation 2
When a return plate is used in surgery, the voltage is
taken off the primary coil shown in the figure.
The Oudin coil is a secondary coil that increases the
voltage by transformer action, so that fulguration can be
done without the return electrode.
For other surgical procedures (cutting) the patient
return plate is used.
Bovie Operation 3
During surgery, the RF current exits the relative sharp
electrode, dissipating between 50 and 400 W of power
into the tissue to make an incision.
The cutting electrode is about 0.1 mm thick and contacts
several millimeters of the tissue.
Voltage of several thousand volts sets up a line of small
sparks and raises the tissue temperature, such that the
tissue separates as the cells vaporize.
Bovie Operation 4
The electrode—tissue interface is
illustrated below.
Tissue Vaporizing
The cells themselves form capacitors with a
conductive electrolyte inside separated by a
nonconductive membrane from the interstitial
fluid (fluid between cells)
That membrane passes the RF currents into the
cell, causing it to vaporize.
Optimal Cutting Currents 1
If the voltage is high enough and is passed quickly enough through
the tissue, the thermal damage is almost imperceptible.
However, if one goes slowly or if the voltage is too low, thermal
tissue damage will result.
Surgeons might tell a BMET the cut isn’t “smooth” – they operate by
feel, and there are several “mixes” of AC current
This could mean the ESU RF power output stage has damage
Optimal Cutting Currents 2
To achieve hemostasis, a certain amount of
damage is desirable
• seals the wound, or “cauterizes”
The control of this factor is key to good surgical
technique using the ESU.
Modern ESU’s have complex RF modulation controls
Modern ESU RF Generators
An RF oscillator still forms the basis for a modern ESU, but
the AC waveform is modulated to give more control:
• Cut mode — Pure sine wave, for cutting with the least
•
•
•
coagulation (allows bleeding)
Coag mode — Pulsed sine wave, low-duty cycle, for
coagulation of bleeding tissue (stops bleeding)
Blend 1 mode — Modulated sine wave, for coagulating
as the tissue is cut.
Blend 2 mode — Modulated sine wave, for coagulating
as the tissue is cut.
Modulated RF Modes
• Coag mode—Pulsed sine wave, low-duty cycle, for
coagulating bleeding tissue.
• Blend 1 mode—Modulated sine wave, for coagulating
with cut, moderate duty cycle
• Blend
2 mode—Modulated sine wave, for coagulating
with cut, higher duty cycle
• Cut mode—Pure sine wave, for cutting with the least
coagulation.
Cut Mode
To cut tissue, the switch is usually set to
the cut position
• This connects the RF voltage to the amplifier,
which then delivers to the active electrode
1,000 to 8,000 volts peak-to-peak AC at from
100 kHz to about 2 megahertz. High-density
currents emerge from the active electrode to
do the cutting.
Return Current Dispersion 1
A blade electrode is moved through the tissue like
a knife to do the cutting.
Knife high-density currents disperse throughout
the conductive fluids of the body and return at a
low-current density to the patient return
electrode to complete the circuit back to the ESU.
• Return electrode has uniform contact to
avoid burns due to current concentrations
Return Current Dispersion 2
•
The return electrode is large
in area and gelled to keep
the skin resistance low and
the region cool.
Ground Isolation to Avoid Burns
The RF circuit is isolated from ground in newer ESUs
So if the patient’s body comes in contact with ground
(through a metal operating table, for example), there
should be no current flow (or burns)
In some older machines, there is no ground isolation, so
a patient touching grounded table could mean burns
But stray capacitance can still lead to some ground
current even in modern equipment
A tingle can result from touching active ESU patient
Cut Mode is Smoothest
In the cut mode, the ESU continuously delivers its
highest average power.
Thus, at every instant as the blade is moved along,
the tissue receives the same treatment. This results
in a smooth cut with no jagged edges.
Complaints that cut isn’t smooth could mean
modulation or power output problems (or operator
error in setting)
Coag Mode is Slower, Ragged
In the coag mode, the average power
delivered to the tissue is reduced from cut
mode, as AC pulses are “on” for only 15%
to 20% of the time instead of 100%
• A blunt (ball-tipped) electrode may also
be touched to the tissue
• The larger surface gives less-dense
current and doesn’t vaporize cells
Coag Mode
Automatically pulsing the AC voltage on 20% and off
80% slows the cutting process, and allows the heat to
propagate into the tissue to form the coagulum.
The depth of coagulation also depends on how long the
electrode contacts the tissue, because tissue damage is
caused by heat propagating into the tissue.
The edge of the cut will tend to be ragged, and some
browning of the tissue will be visible
Blend Modes 1 and 2
Blend modes 1 and 2 are used to cut and seal “bleeders”
simultaneously.
Blend 1: The pulsed AC ON time to OFF time ratio is
25/75 (25% duty cycle) so cuts more than Coag, but seals
blood better.
Blend 2: The pulsed AC ON time to OFF time ratio is
50/50 (50% duty cycle), so cuts half the time with less
“blood sealing” than Blend 1, but smoother cut.
Fulguration
This word means “lightning,” and this is exactly what the
fulguration spark is.
The air between the body and a sharp ESU pencil ionizes
when the electric field intensity exceeds 3000 kV/m, like
lightning
This ionized “plasma” chars the unwanted tissue
A return electrode is not needed at the high voltage, but
does improve current flow
Dessication
If the ESU needle electrode is introduced into a mass,
such as a vascular tumor, the currents will inject power
that raises the fluids to above 100 C, vaporizing and
dehydrating the lesion.
Since lipids and proteins require more than 500 C to
decompose, the surgeon has a mechanism to control
dehydration.
He or she keeps the temperature below 500 C so as to
not decompose the tissue while dehydrating it.
Minimum Testing:
Lights & alarms
Electrode monitoring
Arc on wet soap
Coagulation, Cut & Blend modes
Footswitch and handpiece operation
Surgical Techniques, Problems
The use of an ESU is very dependent on the surgeon preferences, so
one surgeon may see an ESU “fault” where another doesn’t.
The use of the ESU is a refined surgical skill, developed by practice.
ESU machines from different manufacturers have different waveform
shaping, amplitudes, duty cycles, and crest factors
A new surgeon or new ESU machine may see a machine problem that
really just requires re-training to the new machine (new settings)
The BMET must have an ESU tester to prove all modes are OK
ESU Performance Testing 1
Also, calibration of the power levels is done into a test load of fixed
resistance, even with sophisticated testers
But the actual tissue resistance depends upon its type as well as
the electrode contact area pressure against the tissue.
Some high-end ESU machines use active feedback to measure
resistance and adapt power output
All of these factors influence how much power actually gets into the
tissue.
ESU Performance Testing 2
The energy then getting into the tissue depends on the
duration of contact, which is his “fine control”
The effect of the RF current on the tissue cannot be
controlled completely from the ESU machine
It is controlled by the surgeon who has experience in the
procedures required and with the specific ESU being used
Patient Lead Classifications
Monoterminal — An ESU with one wire for patient contact.
Bi-terminal — An ESU with two wires for patient contact.
Active electrode — The electrode that delivers treatment to the surgical
field.
Patient electrode — The large surface area return electrode.
Monopolar electrode — An active electrode that uses a patient
electrode to complete the circuit.
Bipolar electrode—Two electrodes in close proximity and of
approximately the same size that are arranged so that the current tends
to be confined to a small region between the two electrodes (used for
precise coagulation, such as removing skin growths).
Lead Resistance-to-Ground Check
The resistance of both leads to ground should exceed
several megohms, as a check for ground isolation.
This will insure that any alternative path from the
patient to ground would not complete the circuit, to
prevent burn injuries at the point of patient-to-ground
contact.
Someone may set the return plate on a radiator, or it
may make contact with a grounded bed or operating
table.
Split Return Pad with Monitor
Serious burns can result from Patient Return Electrodes which
make poor body contact, or none at all.
One solution is to use two “split” return pads, and monitor the
resistance between the pads.
The ESU will show a fault and not deliver current If the
resistance doesn’t match the expected skin resistance – due to
poor contact of either pad (or shorted pads on a radiator!)
The Cost: Every safety “interlock” gives another mode of
instrument failure.
Common Problems:
Patient Burns
Broken cables
Faulty foot switch or hand piece
Blown power transistors
Adjustments (REM etc.)
Fires
Safety:
Safety is extremely important;
High power spark in O2 rich
environment
Possible high current density for patient
Improper use
Electrical safety methods:
Electrical Isolation
Return electrode resistance
REM electrode