AMPTEC 601ES Explosive Safety DMM

Download Report

Transcript AMPTEC 601ES Explosive Safety DMM

AMPTEC 601ES Explosive Safety DMM– Operator Training
AM PTEC
RESEARCH
CALIBRATED
601ES
6 0 1 E S- 5 0 3
601ES
Battery Compartment
4 each "AA" Alkaline
or Ni-MH rechargeable
EXPLOSIVE SAFETY
DIGITAL MULTIMETER
0 5 APR 2 0 0 5
0 5 APR 2 0 0 6
13
LO BAT
I9999
....
A
V
>>> OR
POWER
ON
STRAY CURRENT
FUSE ( 5 AMPERE)
STRAY CURRENT
PRESS TO ME AS URE
2.0V
AC/DC
20V
AC/DC
2 AMP
AC/DC
2.0 OHM
20 OHM
200V
AC/DC
200 OHM
600VRMS
20 KOHM
press/hold
50m V - 5mA
RESISTANCE
OFF
AC - DC VOLTAGE
VRMS
DISPLAY
BACKLIGHT
13
2.0 KOHM
FUNCTION / RANGE
milliamme ter
0 - 1 0 m A DC
P RE SS TO MEASURE
SIGNAL
SOURCE
+ HIGH
2 Wire
VHI
I
V
VLO
4 Wire
LOW
2 Wire
Term inals
CASE
GROUND
AMPTEC 601ES Explosive Safety Safety DMM –
Startup Check Out Procedure
The first item to examine is to check the calibration sticker due date for the
AMPTEC 601ES Explosive Safety DMM. If the due date is still valid and not
expired, it Is ready to begin the safety check procedure. If the calibration due
date is expired then contact your calibration laboratory or AMPTEC RESEARCH.
A tamper resistant calibration sticker is usually located around the perimeter area
to the AMPTEC 601ES DMM covering one of the calibration access screws.
AMPTEC 601ES Explosive Safety Safety DMM –
Startup Check Out Procedure
Step 1: Procedure for turning on the 601ES DMM –
Using main operating power switch.
POWER
ON
OFF
Hold down the “POWER ON” switch ~ (2) seconds
for power “ON”. The 601ES automatically turns “OFF”
after (30) minutes of idle use to conserve battery power.
Step 2: Observe the 601ES DMM display after unit is “powered
up”. There should be no low battery (“LO BAT”) indicator present
in the LCD display. ( see red circle area below)
The “LO BAT” display means it is
time to replace the unit’s “AA”
batterries. ( shown without backlight)
Since no “LO BAT” is displayed means,
it means the unit’s batterries are OK.
( display shown with backlight lit mode )
The 601ES DMM is capable of
indicating : “A” for amperes, “V”
for voltage, , “k” for Kilo, “Ω ” for
ohms. When the display
indicates “>>> OR” or Overrange
mode – it means possible “open
circuit connection” or a higher
range selection is required.
Step 3: Place the unit in “2 wire” measurement mode to begin the self
check. (make sure that the 2 wire / 4 wire switch is in the the “up”
position).
+ HIGH
2 Wire
VHI
I
V
VLO
4 Wire
LOW
Step 4: The next set of steps envolves checking (by viewing the analog
milliammeter) each range and function to see that less than 0.005 A ( or 5 mA)
is ever output from any FUNCTION or selected range, into the unit’s onboard
analog milliammeter.
2 sections of the AMPTEC 601ES DMM are used for this process.
The FUNCTION/RANGE section and the on-board analog
milliammeter.
2.0V
AC/DC
2 AMP
AC/DC
2.0 OHM
20V
AC/DC
20 OHM
200V
AC/DC
200 OHM
600VRMS
20 KOHM
RESISTANCE
AC- DC VOLTAGE
VRMS
5 millia
limit p mpere
oint
4
2
6
8
MILLAMPERES
10
2.0 KOHM
FUNCTION / RANGE
milliammeter
0 - 10 mA DC
PRESS TO MEASURE
Step 5: Select the 600 VRMS range (rotary switch in far counterclock position).
Then press and momentarily hold down the milliammeter “measure” switch.
You should see virtually no movement in the analog milliammeter. (it’s scale is 0
to 10 mA fullscale).
2.0V
AC/DC
2 AMP
AC/DC
2.0 OHM
20V
AC/DC
20 OHM
200V
AC/DC
200 OHM
600VRMS
20 KOHM
RESISTANCE
AC- DC VOLTAGE
VRMS
5 millia
limit p mpere
oint
4
2
6
8
MILLAMPERES
10
2.0 KOHM
FUNCTION / RANGE
milliammeter
0 - 10 mA DC
PRESS TO MEASURE
Step 6: Select the 200 VRMS range (rotary switch in 8 O’clock position).
Then press and hold the milliammeter “measure” switch. Again , you should
see virtually no movement in the analog milliammeter. (it’s scale is 0 to 10 mA
fullscale).
2.0V
AC/DC
20V
AC/DC
2 AMP
AC/DC
20 OHM
200V
AC/DC
200 OHM
600VRMS
20 KOHM
5 millia
limit p mpere
oint
2.0 OHM
RESISTANCE
AC- DC VOLTAGE
VRMS
The yellow line on range switch is shown for
emphasis of position purposes.
4
2
6
8
MILLAMPERES
10
2.0 KOHM
FUNCTION / RANGE
milliammeter
0 - 10 mA DC
PRESS TO MEASURE
Step 6: Select the 20 VRMS range (rotary switch in 10 O’clock position).
Then press and hold the milliammeter “measure” switch. Again, you should see
virtually no movement in the analog milliammeter. (it’s scale is 0 to 10 mA fullscale).
2.0V
AC/DC
20V
AC/DC
2 AMP
AC/DC
2.0 OHM
20 OHM
200V
AC/DC
200 OHM
600VRMS
20 KOHM
5 millia
limit p mpere
oint
4
RESISTANCE
AC- DC VOLTAGE
VRMS
The yellow line on range switch is shown
for emphasis of position purposes.
2
6
8
MILLAMPERES
10
2.0 KOHM
FUNCTION / RANGE
milliammeter
0 - 10 mA DC
PRESS TO MEASURE
Step 7: Select the 2.0 VRMS range (rotary switch in ~11 O’clock position).
Then press and hold the milliammeter “measure” switch. Again, you should see
virtually no movement in the analog milliammeter.
2.0V
AC/DC
20V
AC/DC
2 AMP
AC/DC
2.0 OHM
20 OHM
200V
AC/DC
200 OHM
600VRMS
20 KOHM
2.0 KOHM
FUNCTION / RANGE
5 millia
limit p mpere
oint
4
RESISTANCE
AC- DC VOLTAGE
VRMS
The yellow line on range switch is shown for
emphasis of position purposes.
2
6
8
MILLAMPERES
10
milliammeter
0 - 10 mA DC
PRESS TO MEASURE
Step 8: Select the 2.0 AMP AC/DC range (rotary switch in 12 O’clock position).
This step requires simultaneously pressing and holding down 2 buttons.
*Press and hold the 2.0 AMP button ( see red arrow below) while you also
press and hold the milliammeter “measure” switch. Again, you should see
virtually no movement in the analog milliammeter.
2.0V
AC/DC
2 AMP
AC/DC
20V
AC/DC
20 OHM
200V
AC/DC
200 OHM
600VRMS
20 KOHM
4
2.0 OHM
2.0 KOHM
FUNCTION / RANGE
2
RESISTANCE
AC- DC VOLTAGE
VRMS
5 millia
limit p mpere
oint
6
8
MILLAMPERES
10
milliammeter
0 - 10 mA DC
PRESS TO MEASURE
Step 9: Select the 2.0 OHM range (rotary switch in the 1 O’clock position).
Press and momentarily hold down the milliammeter “measure” switch. You
should see the analog milliammeter needle deflect to to midscale or the 5 mA
(verticle needle position is “OK” – the milliammeter scale is 0 to 10 mA fullscale).
You should not see 5 ½ milliamperes, avoid paralax viewing error of the
milliammeter by taking reading straight over the milliammeter - use one eye to
view. ( 8 mA is the failsafe current level).
5 millia
n om in m per e
al lim it
p oin t
20 OHM
200V
AC/DC
200 OHM
600VRMS
20 KOHM
0
2.0 OHM
8
10
2.0V
AC/DC
20V
AC/DC
2 AMP
AC/DC
6
MILLAMPERES
RESISTANCE
AC- DC VOLTAGE
VRMS
4
2
2.0 KOHM
FUNCTION / RANGE
milliammeter
0 - 10 mA DC
PRESS TO MEASURE
Step 10: Select the 20 OHM range (rotary switch in the 2 O’clock position).
Press and momentarily hold down the milliammeter “measure” switch. You
should see the analog milliammeter needle deflect to to midscale or the 5 mA
level. It should be the same as the 2.0 ohm current source level. A verticle
needle position is “OK” – the milliammeter scale is 0 to 10 mA fullscale. Again,
you should not see 5 ½ milliamperes, avoid paralax viewing error of the
milliammeter by taking reading straight over the milliammeter - use one eye to
view.
2.0 OHM
20 OHM
200V
AC/DC
200 OHM
600VRMS
20 KOHM
4
2
0
2 AMP
AC/DC
6
8
MILLAMPERES
10
2.0V
AC/DC
20V
AC/DC
RESISTANCE
AC- DC VOLTAGE
VRMS
5 millia
n om inam per e
l lim it p
oin t
2.0 KOHM
FUNCTION / RANGE
milliammeter
0 - 10 mA DC
PRESS TO MEASURE
Step 11: Select the 200 OHM range (rotary switch in the ~3 O’clock position).
Press and momentarily hold down the milliammeter “measure” switch. You
should see the analog milliammeter needle deflect to barely ~1 mA level. You
should not see any level even close to 5 ½ milliamperes.
2.0 OHM
20 OHM
200V
AC/DC
200 OHM
600VRMS
20 KOHM
4
2
0
2 AMP
AC/DC
6
8
MILLAMPERES
10
2.0V
AC/DC
20V
AC/DC
RESISTANCE
AC- DC VOLTAGE
VRMS
5 millia
n om in m per e
al lim it
p oin t
2.0 KOHM
FUNCTION / RANGE
milliammeter
0 - 10 mA DC
PRESS TO MEASURE
Step 12: Select the 2.0 KOHM range (rotary switch in the ~ 4 O’clock position).
Press and momentarily hold down the milliammeter “measure” switch. You
should see the analog milliammeter needle deflect to barely any current or ~<1 mA
level. You should not see any level even close to 5 ½ milliamperes.
2.0 OHM
20 OHM
200V
AC/DC
200 OHM
600VRMS
20 KOHM
4
2
0
2 AMP
AC/DC
6
8
MILLAMPERES
10
2.0V
AC/DC
20V
AC/DC
RESISTANCE
AC- DC VOLTAGE
VRMS
5 millia
n om in m per e
al lim it
p oin t
2.0 KOHM
FUNCTION / RANGE
milliammeter
0 - 10 mA DC
PRESS TO MEASURE
Step 13: Select the 20 KOHM range (rotary switch in the 6 O’clock position).
Press and momentarily hold down the milliammeter “measure” switch.
You should see the analog milliammeter needle indicate barely any current at all
or ~<1 mA level. You should not see any level even close to 5 ½ milliamperes.
2.0 OHM
20 OHM
200V
AC/DC
200 OHM
600VRMS
20 KOHM
4
2
0
2 AMP
AC/DC
6
8
MILLAMPERES
10
2.0V
AC/DC
20V
AC/DC
RESISTANCE
AC- DC VOLTAGE
VRMS
5 millia
n om in m per e
al lim it
p oin t
2.0 KOHM
FUNCTION / RANGE
milliammeter
0 - 10 mA DC
PRESS TO MEASURE
If the unit has successfully completed all the steps indicated it is ready to use.
This concludes the AMPTEC 601ES Explosive Safety DMM check out procedure.
1. Calibration/Maintenance access screws
sealed with tamper resistance calibration
sticker (see red arrow).
2. Purge Valve - Automatic Air Pressure
Equalization
3. Handles - See item #1, only lift for calibration access.
4. Power “On/Off ” switch, silicon rubber boot
covered, operates on 4 “AA” batteries (alkaline or NiMH). Automat- ically turns “OFF”
after (30) min. of in-activity. Located near
shoulder of handle to avoid accidental power
turn ON.
5. Protection Fuse (5.0 amp fast blow type)
protects the 2.0 ampere Stray Current Range
(measurement). The type of fuse used is a
Little Fuse 3AG or equivalent.
6. 20,000 count LCD Displaywith ( A) amps,
(V ) voltage,
( ) Ohms, (k ) kilo symbols plus (>>>OR )
overrange indicator if a range’s magnitude is
exceeded.
6A. The LCD Displaywill indicate "LO
BAT" when it is time to replace the batteries.
Replace (4) ea. “AA” batteries with a heavy
duty 1.5 V Alkaline i.e. Duracell Ultra, or
rechargeable Energizer 2500 mAHr 1.2V
Nickel Metal Hydride rechargeable or equivalent.
7. Display Backlight is push button activated
(dim ambient light areas). Note: Use sparingly
to conserve battery power.
8. Meter Test Lead Input Jacks - For two
wire test lead connections, you should usethe
far left input jacks both
VHI and VLO, shown as a dotted box. Use
these two terminals V HI and VLO for all voltage, current and simple ohms measurements.
The “4 Wire” switch position and terminals are
only for 4 Wire Kelvin resistance measurements.
9. Zero. 'Short' the test lead ends prior to
adjustment of the Zero knob only for resistance measurements. Readjust of the Zero
knob is only needed if connections are
changed. This is a precision 10 turn pot with
delicate stops at both ends (i.e. used with 2
Wire 2 ohm, 20ohm and 200 ohm ranges).
10. Case or Chassis Ground Jack is connected to the
meter’s internal Faraday Cage, metal top plate and the
unit’s bottom plate (an accessible chassis to earth
ground connection).
11. RMS AC/DC Voltage Measurement : 600.0V,
200.0V, 20.0 V & 2.0 VRMS AC/DC (i.e. 0.0500
VRMS measurement on the 2.0 VRMS range - 4 ½
digit resolution).
12. Stray Current Press To Measure - To measure
AC/DC current, connect input and output wires from
the UUT to the far left 2 terminals ( VHI V
/ LO
inputs). Connect to the “UUT”. Depress and Hold
the Stray Current button. A 5.0 ampere fuse protects
the meter from current > than 5 amps.
13. 0 to 10 mA DC Analog Milliammeter measures
DC current output levels with the "Press to
Measure” button. To test the AMPTEC 601ES
resistance function's test current level select the
resistance range ( rotary knob) and press the
milliammeter's "Press to Measure " button.
14. AMPTEC 601ES Safety DMM Resistance
Ranges include a 2.0 Ohm range with 0.1 milliohm
resolution, 20.0 Ohm, 200 Ohm, 2.0 KOhm and 20
KOhm ranges.
15. The Signal Source function is used to verify
DVM voltage measurement (i.e. the2.0 V range) and
stray current measurement at low levels. Select the
2.0 V Range. Depress and hold the signal source
button. The display should indicate approx. 0.0500 V
( around 50 mV ). Next, select the 2 AMP range.
Depress and hold the Signal Source button along
with the Stray Current button. The display should
indicate approx. 0.0050 amperes ( ~ 5 mA ).
16. Flip Latch - Place palm on the lid corner. Press
while flipping up the latch to open the AMPTEC
601ES DMM.