Electrochemistry
Download
Report
Transcript Electrochemistry
MLAB 2401: CLINICAL
CHEMISTRY
KERI BROPHY-MARTINEZ
Designs in Instrumentation
Electrochemistry
ELECTROCHEMISTRY
Basic principle
Electrodes are used to selectively measure
particular ions
Instruments utilizing electrodes measure the
potential difference (charge) that builds up at
an interface when two different concentrations
of the same ion are in contact with each other
The electrodes are sensitive to this potential
difference
2
ELECTROCHEMISTRY: TERMS
Oxidation
Anode
gain of electrons (electrode becomes negative)
Cathode
the electrode where oxidation occurs (where electrons
are lost)
Reduction
loss of electrons (leaves electrode positive)
the electrode where reduction occurs (where the
electron gain occurs)
Polarography
the measurement of the gain or loss of electrons in a
chemical reaction by detecting a change in the potential
3
ELECTROCHEMISTRY
Galvanic cell – two half cells used to demonstrate
the flow of electrons during the processes of
oxidation and reduction
4
oxidation
Galvanic c ell
reduction
ELECTROCHEMISTRY
Electrolytic cell differs from the galvanic cell in
that
The cathode and anode are immersed in the same
solution
There is an electromotive force used to drive the
chemical reaction
5
ELECTROCHEMISTRY
Testing Procedures involving electrochemical
principles
Potentiometry
pH meters
Ion-selective electrodes
Blood gas analysis
Chloridometers
POTENTIOMETRY
Measurement of potential or voltage between two
electrodes in a solution
Nerst equation
Predicts amount of change in millivolts due to a
temperature change
As the temperature rises, so does hydrogen ion
activity
POTENTIOMETRY
System Components
Reference electrode
Indicator or measuring electrode
Liquid junction
Readout device
POTENTIOMETRY
Reference electrodes
The electrode against which the potential charge
created in the indicator electrode is measured.
Consists of a metal and its salt in contact with a
solution containing the same anion. These are half-cell
potentials extremely stable and easy to reproduce
9
POTENTIOMETRY
Reference
Electrodes
Calomel electrode composed of
mercury/mercurous chloride is dependable but
large, bulky, and affected by temperature
Silver/silver chloride reference electrodes
are more compact and handle temperature
fluctuations better - overall better & faster
10
POTENTIOMETRY
Indicator (measuring) electrodes
The electrode, whose half-cell potential
responds to
changes in the activity or concentration of the
substance in the solution that is being measured.
The response of the electrode to the substance
(species) being measured and not to other substances
present is defined as the selectivity of the electrode.
The selectivity of an electrode for a particular species
is dependent on the type of membrane that
separates the electrode and the solution containing
the species to be measured.
11
ELECTROCHEMISTRY
Liquid junction – also known as a salt bridge
are required to complete the circuit (between the
reference and without contaminating anything.
KCl (potassium chloride) is the most common
12
POTENTIOMETRY
Ion Selective Electrodes
Also called Ion Specific Electrodes
Indicator electrode
Modified electrochemical half cells
Advantages
No reagent preparation
No standard curve preparation
Cost effective
Fast analysis times
Very sensitive and selective for an analyte
Easy to maintain
13
ION SELECTIVE ELECTRODES
Different membranes that are selectively sensitive
to the electrical effects of different electrolytes can
be placed over the tips of the measuring electrodes,
making them susceptible only to the effects of
these particular electrolytes
Differences in the measured potential of the circuit
can be calibrated with known concentrations of
electrolytes
Because ISE measures electrical potential, it is an
example of potentiometry
14
ION SELECTIVE ELECTRODES
Common
substances measured by ISE
Sodium (Na) –
membrane composed of selective glass
Potassium (K)
valinomycin ( example of liquid layer membrane)
Chloride (Cl) –
crystal membrane
Ionized Calcium (Ca)
Hydrogen ions (H)
glass membrane different composition than Na’s
15
ION SELECTIVE ELECTRODES ( ISE )
Consist of two main components to an
electrochemical half cell
Reference Electrode ( generates a known,
constant voltage )
Measuring Electrode ( generates a variable
voltage )
These components must be connected together,
but can not contaminate each other. The joining
is done using a “salt bridge”
16
POTENTIOMETER MEASUREMENT
pH Electrode
based on the measurement of a potential
(voltage) difference between two electrodes
immersed in a solution under the condition
of zero current.
Each cell has a half-cell reaction
The potential difference between the two
electrodes is usually measured using a
pH/milli-volt meter.
17
PH ELECTRODES
Combination electrode
Internal reference
Both the indicator &
reference electrodes are
housed in the same
assembly.
Ag/AgCl type bathed
in KCl
The measuring or
indicator electrode has
a “glass membrane”
18
BLOOD GAS ELECTRODES
pCO2 Electrode
Measurement of PCO2 in
routine blood gases is
done using the
Severinghaus electrode
A modified pH electrode
with a CO2 permeable
membrane covering the
glass membrane surface
A bicarbonate buffer
separates the membranes
Change in pH is
proportional to the
concentration of dissolved
CO2 in the blood
19
BLOOD GAS ELECTRODES
pO2
Electrode
Clark PO2 (polarographic) electrode.
Measurement of PO2 is done using a
Clark electrode which works on the
principles of polarography
principle of polarography- involves
measurement of gain or loss of
electrons in a chemical reaction, by
detecting a change in potential.
20
PO2
ELECTRODE
The cathode is a platinum wire and the anode is a
silver wire in AgCl
both electrodes are in contact with electrolyte
solution
Oxygen in the sample diffuses through the semipermeable
membrane and is reduced (gains electrons) at the cathode.
The flow of electrons stops when no more oxygen is
available.
Current generated in the system is proportional to the
amount of oxygen
21
ELECTROCHEMISTRY COULOMETRIC/AMPEROMETRIC
CHLORIDOMETERS
Coulometry
a titration in which the titrant is
electrochemically generated
Silver (Ag) is the most common titrant.
Amperometry
Measurement of current flow produced by an oxidationreduction reaction
22
COULOMETRIC/AMPEROMETRIC
CHLORIDOMETERS
The basic components consist of two pair of
electrodes immersed in an electrolyte diluent
(containing acetic and nitric acids)
E q u iva le n ts A g + g e n e ra t ed =
–
e q u iva le n t C l
23
COULOMETRIC/AMPEROMETRIC
CHLORIDOMETERS
Procedure
Sample added to
diluent
Electrodes releases
silver ions into the
diluent
Ions combine with
chloride to form a
precipitate
When chloride is gone,
excess silver causes a
change in current flow
Flow change signals
the end of titration
REFERENCES
Bishop, M., Fody, E., & Schoeff, l. (2010). Clinical
Chemistry: Techniques, principles, Correlations. Baltimore:
Wolters Kluwer Lippincott Williams & Wilkins.
Sunheimer, R., & Graves, L. (2010). Clinical Laboratory
Chemistry. Upper Saddle River: Pearson .
25