Waveform display devices

Download Report

Transcript Waveform display devices

Course Outlines
 Calibration and Maintenance of Medical
Equipment
 Recording devices
Strip chart recorder
Magnetic tape recorder
Digital recorder
 Ultrasound Instrument
 X-Ray Instrument
 Dialysis Machine
 CT Scan
 Magnetic Resonance Imaging
Calibration and Electrical Safety of
Medical Equipment
 Why Test & Calibration
 As components age and equipment undergoes
changes in temperature or humidity or sustains
mechanical stress, performance gradually
degrades. This is called drift.
 When this happens your test results become
unreliable and both design and performance
quality suffer.
 While drift cannot be eliminated, it can be detected
and either corrected or compensated for through
the process of calibration.
Definitions
Calibration: process of comparing an
unknown against a reference standard
within defined limits, accuracies and
Uncertainties
Verification: process of comparing an
unknown against a reference standard at
usually one data point
Test & Calibration service
What to TEST for?
• Performance Testing
• Safety Testing
When to test equipment
prior to being accepted for use
During preventative maintenance.
After repairs.
Need for Medical Equipment Testing
• Medical device incidents resulting in patient injury and
death
• Ensure that the equipment is performing to the expected
standards of accuracy, reliability, free of hysteresis and
linear (as designed).
• Safe and effective devices need to be available for
patient care
– Downtime costs money
• Regulations, accreditation requirements and standards.
Why do we do electrical safety?
 Ensure patient safety
Protect against macroshock
Protect against microshock
 Test for electrical internal breakdown / damage to
power cord, AC mains feed, etc.
 Meet codes & standards
 Association of Medical Instrumentation (AAMI),
The International Electrotechnical Commission (IEC),
National Federation of Paralegal Associations (NFPA),
etc.
 Protect against legal liability
In case of a patient incident
Waveform display devices
Permanent magnet moving coil instruments
‫أجهزة الملف المتحرك لمغناطيس مستديم‬
 Chart recorder: (Recording Oscillographs)
 used in a medical instrumentation to make permanent recording of the
waveform.
 most common type: permanent magnet moving coil (PMMC).
 The PMMC is very similar to the Galvanometer movement.
 A writing pen replaced the meter point.
 Current flowing in the moving coil creates a magnetic field that interact with the
magnetic field of the permanent magnet. This will cause deflection of the pen.
 The tip of the pen is positioned over a strip of chart paper that is pulled under the pen tip at a
constant speed. In this mechanism:
• Y-axis is the deflection to the pen and
•
X-axis is the time base established by moving the chart paper at a constant speed.
 This will produce a chart recording of the waveshape of the applied waveform
Curvilinear recording
 The pen assembly sweeps an arched manner
and so will write in a curvilinear manner.
Rectilinear recording
 The pivoted pen motor assembly is a solution of
curvilinear action.
 The pen is connected through a mechanical link
that translates the curvilinear motion of the
PMMC to the rectilinear motion at the pen tip.
Pseudo rectilinear writing system
 In this type of recorder the pen assembly is very long
compared to width of the chart paper.
 The pen tip, therefore travels in an arc the length of which
is very short compared with the radius.
 The trace will appear to be nearly linear
Writing Methods Used in Strip Chart
Recorders
 Thermal, and Direct contact
 Both of these types use a special writing stylus (rather
than a pen and a knife edge, also called writing edge)
 The mark on the paper by the contact of the stylus on
the paper along the knife edge,
 The stylus tip travels in a curvilinear path, but the
resulting trace is rectilinear because the knife edge is
straight.
 The stylus can write anywhere along its length, so by
keeping the knife edge straight under the paper, we
obtain the rectilinear recording that shows waveshape
as well as amplitude
PMMC writing system
 Several writing systems are commonly used on PMMC recorders:
1- direct contact,
2- thermal pen
3- ink pen
4- ink jet
5- optical
 Recorder that use any type of pen (ink) or stylus (Direct contact and
thermal) have a relatively low-frequency response due to the inertia of the
pen or stylus assembly (100-200 Hz) Ink jet and optical type
 Lighter weight fixtures and writing stylus system have a frequency response
of 1000-3000 Hz
Direct contact
The direct contact uses a special types of
chart paper that is chemically treated to
have a carbonized back.
When a pressure applied to that front of
the paper a black mark will appear.
Most of these instrument have a frequency
response of less than 25 Hz and so are
not commonly used in medical
instrumentation.
The thermal recorder
 The thermal recorder also uses special paper, but in this
case it is waxed or treated with paraffin so that it will turn
black when treated
 The thermal recorder is the most commonly employed in
medical instrumentation, especially in cardiovascular
instruments such as the ECG and pressure monitors
The Stylus
 The stylus in a thermal system is little more than a heated resistance wire
connected to a low voltage ac or dc power supply.
 Early models formed a U-shaped electrical resistance element
 Modern models use a wire inside a cylindrical metal stylus
 In both cases a low voltage electrical power supply energizes the element,
causing the tip to become heated.
 The black mark is made at the points where heated stylus touches the paper
Ink pen writers
 use a hollow pen and an ink supply to write on the chart paper.
 In some machines the ink pressured by an atomizer-like hand pump
 EEG system use this type of writing system.
 More automatic machines use a thick, viscous ink in a special cartridge
that is placed under pressure by a spring-driven piston
 In multichannel instruments, the ink may be distributed to all pens from
the same cartridge by connecting the line from the cartridge to the input
side of a special ink manifold and additional pressure is applied to the
ink by the solenoid-operated manifold bladder
Ink jet recorder
 Higher-frequency response than
are the types mentioned
previously.
 This type is popular on European
instrument
 Low-viscosity ink is directed to a
nozzle mounted on a PMMC
galvanometer in place of the pen
assembly.
 The ink jet produced by the nozzle
is directed at the paper.
 When the system is properly
adjusted, it will produce a
recording that is very linearly
rectilinear.
 Only a small amount of trace
fuzziness due to ink splattering is
apparent
‫االنواع التى تستجيب للتردادات المرتفعة اكثر‬
‫من االنواع السابقة مشهورة فى اوروبا‬
‫حبر منخفض اللزوجة يوجه الى فوهة فى‬
‫الجلفانومتر مكان القلم‬
‫الحبر المتدفق من الفاتحة يوجه الى الورقة‬
‫عند ضبط النظام تماما سوف ينتج خط تام‬
‫االستواء‬
‫ينتج بعض التشوش نتيجة بعثرة الحبر‬





Optical recorder
 There are two types of optical recorder:
 - PMMC type uses small inertia mirror in place
of the pen assembly or stylus.
 - The other uses a photographic paper that is
pulled across a cathode ray tube and is
called a CRT camera recorder.
 Most of these recorders use wide paper.
 On multichannel optical recorder it is possible
to examine the time relationships between
different traces more easily because the
traces can be allowed to overlap each other.
 The paper in the optical recorder is often develop by
exposure to an ultraviolet lamp as the paper comes out
from the recorder.
 Unless the paper is either wet developed following the
recording session or stored in a light-light box.
 In CRT camera, the CRT sweeps only the vertical axis.
 The time base is provided by pulling the photosensitive
paper in the front of the CRT screen.
 The frequency response of the CRT camera recorder is
better than that of any of the other types, being limited
mostly by the writing speed of the photosensitive paper.
Servo recorders and recording
potentiometers
 In potentiometeric measurements a three-terminal variable resistor
(potentiometer) is connected to produce an output voltage that is a
function of both a reference potential and position of the variable
resistor’s wiper arm.
 A galvanometer will read zero when the unknown voltage and
potentiometer voltage are equal.
Servo recorder
 A servo recorder is a self-nulling potentiometer that records the
waveshape of the applied signal on graph paper.
 The pen attached to a string that is wound around a pair of idler
pulleys and drive pulley that is on the shaft of a dc servomotor.
 The pen assembly is also linked to a potentiometer (R1) in such a
way that the position of the wiper arm on the resistance element is
proportional to the pen position.
 The potentiometer element is connected across a
reference potential Eref, so potential E is the electric
analog of pen position.
 The pen position is controlled by the dc servomotor,
which in turn driven by the output of the servo
amplifier.
 The amplifier has a differential inputs: Ein (unknown)
is connected to one input, and the position signal E
is connected to the other input
 The difference signal (Ein – E) represents the error
between the actual pen position and the position the
pen should be in for the applied voltage.
 If the error is zero , meaning that the amplifier output is also zero and
the pen is correctly positioned, then the motor remain turned off.
 If Ein# E then the amplifier creates an output signal that turns on the
motor
 The motor drive the pen and potentiometer in such a direction as to
cancel the error signal
 When the input signal and position signal are equal, then the motor
turns on the motor.
A paper drive motor
 A paper drive motor forms time base because it pulls the paper
underneath the pen at constant rate.
 Most high-quality servorecorders use a stepper motor to drive the
paper supply.
 Such motor will rotate only a few degree every time a pulse applied
to its windings
 A few models use a continuously running motor that drive the
sprocket through a speed-reducing gear box.
 The stepper motor system is capable very good accuracy because a crystal oscillator
or the ac power mains are used to drive the pulses used to advance the motor.
 Digital integrated circuit frequency dividers (counter circuit) can be used to reduce the
clock frequency to the frequency required to drive the motor at desired speed.
 The reference potentiometer used to measure the pen position may be any of the
following devices:
- Slide wire - Rectilinear, or - Rotary.
 The slide-wire system is often used because it can be built with less friction and no
mechanical linkage.
A slide-wire potentiometer
 A resistance wire and a shorting wire are stretched taut parallel to
each other and the direction of the pen travel.
 A shorting bar on the pen assembly serves as a wiper on the
resistance element and also connects the shorting wire.
 A,B, and C in Figure refer to the potentiometer terminals
 the shorting wire serves as terminal B of the potentiometer (the
wiper)
X-Y recorders
 The X-Y recorder uses two servo mechanisms
connected to the same writing assembly, but at right
angles to each other.
 The X-axis servomechanism moves the pen-bar
assembly back and forth across the paper in the
horizontal plane, while the Y-axis servomechanism
moves the pen vertically up and down along the bar.
 The paper itself does not move. It is held in place either
by clamp or, in high-quility instruments by a vacuum
pump that is used to evacuate a hollow chamber below
the paper platform.
 Holes in the platform create the negative pressure
needed to keep the paper in place.
 One advantage of the X-Y recorder is that almost any
type of paper may be used.
Problem of recorder design
 1- A dead band signal: largest signal to which the recorder will not respond
due to pen assembles mass (inertia). Sufficient preamplification of the
signal is required
 2- Overshooting and undershooting of the recorded trace.
 3- undercritically damped reader will overshoot the correct point and then
hunt back and forth across the correct point for a few cycle until it hones in
and settles properly
 4- over critically damped recorder is sluggish the pen approaches the
correct position very slowly .
Deadband
 Pen assembles are often damaged if they strike the limits-of-travel
steps at a high speed.
 A pair of Zener diodes connected back to back across the PMMC
coil are sometimes used to accomplish the same job.
 The Zenner potential of the diodes is selected so that the bodies
break over and conduct current only when a voltage greater than the
normal full-scale potential is applied to the input of the amplifier
Maintenance of the PMMC writing stylus
and pens
 Figure shows how to remove an ink blockage from an ink pen recorder that has
been allowed to stand too long without being used.
 As a general rule, such recorders should be run for about five minutes or so once a
week when not in regular service.
 Fill a 3- t0 10-cc syringe with water (or acetone for certain type of ink)
 Insert end into the ink inlet
 The pen has to removed from the machine for this operation.
 Quickly, and with a single sharp motion, drive the plunger “home” so that a high
pressure jet of water or acetone is forced into the pen. The ink clot should be forced
out the other end.
Precautions
 Always wear protective-goggles and protective clothing.
 Make sure that the pen is aimed downward into a sink
 As always when dealing with needles, be careful not to stick yourself
 The thick high-viscosity ink strains every thing it touches and a
nearly impossible to remove.
 Ink pen tips are designed to operate parallel to the paper surface
 If the pen is worn or when a new pen is installed, it is necessary to lap the
tip in order to reestablish the parallelism.
 The sign that lapping is needed will be either (or both) of the following:
 A- a blob f ink when the machine first stats a waveform, or
 B- a too-thick trace
 To lap the pen, place a piece of fine Emery cloth (sandpaper) under the tip.
 Work the pen tip back and forth 5 to 10 times to sand the tip parallel to the
paper
 The pressure of the stylus or pen is also important.
 If the pressure is not correct, then the waveform may be distorted
 .
 In medical equipment it is possible to make a normally healthy lead-1
ECG as though the patient has.
 Suitable stylus ECG gages can be purchased from ECG machine
manufactures.
 The stylus pressure adjustment is made using a screw that is usually
located on the rear of the stylus or the assembly that holds it in place.
Dot matrix analog recorders
 The original dot matrix printers used a 5 X 7 matrix of dots to form
alphanumeric data.
 The dot matrix machine used a print head to cause the correct dot element
Two different methods:
A- Thermal based machine:
 Some of the earliest machines were thermally based.
 The dots were thermally connected to heating coils and could be heated
when needed.
 Special sensitive paper was used to receive the text.
 This method is no longer widely used.
 B- an array of seven print hammers (pins).
 The pins would either extend or retract depending whether or not that
particular data was active for the character being printed.
 An advantage of the pin method is that ordinary paper can be used
 The pen impacted at inked ribbon to leave the impression
 Higher resolution models are now valuable with 9, 18, 24 pins